REPORT No.SZ24020013W01
TEST REPORT

APPLICANT

: Adec & Partner AG

PRODUCT NAME : earis MAX

MODEL NAME

: HM-01

BRAND NAME

: Humantechnik

FCC ID

: U94MAX0124R

STANDARD(S)

: 47 CFR Part 15 Subpart C

RECEIPT DATE

: 2024-02-02

TEST DATE

: 2024-02-21 to 2024-03-06

ISSUE DATE

: 2024-05-15

Edited by:

Peng Mi (Rapporteur)

Approved by: Shen Junsheng (Supervisor)

NOTE: This document is issued by Shenzhen Morlab Communications Technology Co., Ltd., the test report shall not be reproduced except in full without prior written permission of the company. The test results apply only to the particular sample(s) tested and to the specific tests carried out which is available on request for validation and information confirmed at our website.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]

Page 1 of 56

REPORT No.SZ24020013W01
DIRECTORY
1. Summary of Test Result ················································································ 4 1.1. Testing Applied Standards ·········································································· 5 1.2. Test Equipment List ··················································································· 6 1.3. Measurement Uncertainty ··········································································· 8 1.4. Testing Laboratory····················································································· 8 2. General Description ····················································································· 9 2.1. Information of Applicant and Manufacturer ··················································· 9 2.2. Information of EUT····················································································· 9 2.3. Channel List of EUT ··················································································10 2.4. Test Configuration of EUT·········································································· 11 2.5. Test Conditions ························································································ 11 2.6. Test Setup Layout Diagram ········································································ 11 3. Test Results ·······························································································14 3.1. Antenna Requirement ···············································································14 3.2. Duty Cycle of Test Signal ··········································································· 15 3.3. Maximum Peak Conducted Output Power ···················································· 16 3.4. Maximum Average Conducted Output Power ···············································17 3.5. 6 dB Bandwidth ························································································ 18 3.6. Conducted Spurious Emissions and Band Edge ··········································· 19 3.7. Power Spectral Density ············································································· 20 3.8. Conducted Emission·················································································21 3.9. Restricted Frequency Bands ······································································ 22 3.10. Radiated Emission····················································································23 Annex A Test Data and Result ··········································································25

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 2 of 56

Version 1.0

REPORT No.SZ24020013W01

Change History Date 2024-05-15

Reason for change First edition

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 3 of 56

1. Summary of Test Result

REPORT No.SZ24020013W01

No. Section Description

Test Date

Test Engineer

Result

Method Determination
/Remark

1 15.203

Antenna Requirement

N/A

N/A

PASS No deviation

2 N/A

Duty Cycle of Test Signal

Feb. 22, 2024

Zhong Yanshan

PASS No deviation

Maximum Peak

3 15.247(b) Conducted

Feb. 22, 2024

Output Power

Zhong Yanshan

PASS

No deviation

Maximum

4

15.247(b)

Average Conducted

Feb. 22, 2024

Zhong Yanshan

PASS

No deviation

Output Power

5 15.247(a) Bandwidth

Feb. 22, 2024

Zhong Yanshan

PASS

No deviation

Conducted

6

15.247(d)

Spurious Emission and

Feb. 22, 2024

Zhong Yanshan

PASS

No deviation

Band Edge

7

15.247(e)

Power Spectral Density

Feb. 22, 2024

Zhong Yanshan

PASS

No deviation

8 15.207

Conducted Emission

Feb. 29, 2024

Wang Deyong

PASS No deviation

Restricted

Feb. 29, 2024

9 15.247(d) Frequency

to

Gao Jianrou PASS No deviation

Bands

Mar. 01, 2024

10

15.209, 15.247(d)

Radiated Emission

Feb. 29, 2024 to
Mar. 04, 2024

Gao Jianrou

PASS

No deviation

Note 1: The tests were performed according to the method of measurements prescribed in ANSI

C63.10-2013 and KDB 558074 D01 v05r02.

Note 2: Additions to, deviation, or exclusions from the method shall be judged in the "method

determination" column of add, deviate or exclude from the specific method shall be explained in the

"Remark" of the above table.

Note 3: When the test result is a critical value, we will use the measurement uncertainty give the

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 4 of 56

REPORT No.SZ24020013W01
judgment result based on the 95% confidence intervals.
1.1. Testing Applied Standards
According to the specifications of the manufacturer, the EUT must comply with the requirements of the following standards:  47 CFR Part 15 Subpart C Radio Frequency Devices

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 5 of 56

1.2. Test Equipment List

REPORT No.SZ24020013W01

1.2.1 Conducted Test Equipment

Equipment Name Serial No.

EXA Signal Analyzer

MY53470836

RF Cable (30MHz-26GHz)

CB01

Coaxial Cable

CB02

SMA Connector

CN01

Type N9010A
RF01 RF02 RF03

Manufacturer Agilent

Cal. Date 2024.02.19

Morlab

N/A

Morlab

N/A

HUBER-SUHNER

N/A

Due Date 2025.02.18
N/A N/A N/A

1.2.2 Conducted Emission Test Equipment

Equipment Name Serial No.

Type

Receiver

MY56400093 N9038A

LISN
Pulse Limiter (10dB)
RF Coaxial Cable (DC-100MHz)

8127449
VTSD 9561 F-B #206
BNC

NSLK 8127 VTSD 9561-F
MRE04

Manufacturer KEYSIGHT

Cal. Date Due Date 2024.01.25 2025.01.24

Schwarzbeck 2024.02.02 2025.02.01

Schwarzbeck 2023.06.27 2024.06.26

Qualwave

N/A

N/A

1.2.3 List of Software Used Description Test System Morlab EMCR
TS+ -[JS32-CE]

Manufacturer MaiWei Morlab Tonscend

Software Version 2.0.0.0 V1.2 V2.5.0.0

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 6 of 56

REPORT No.SZ24020013W01

1.2.4 Radiated Test Equipment

Equipment Name

Serial No.

Receiver

MY54130016

Test Antenna Bi-Log

9163-519

Test Antenna Loop

1519-022

Test Antenna ­ Horn

01774

Test Antenna ­ BBHA9170

Horn

#773

Preamplifier (10MHz-6GHz)

46732

Preamplifier (2GHz-18GHz)

61171/61172

Preamplifier (18GHz-40GHz)

DS77209

Type N9038A VULB 9163
FMZB1519
BBHA 9120D
BBHA9170 S10M100L38
02 S020180L32
03 DCLNA0118-
40C-S

RF Coaxial Cable (DC-18GHz)

MRE001

PE330

RF Coaxial Cable (DC-18GHz)
RF Coaxial Cable (DC-18GHz)
RF Coaxial Cable (DC-40GHz)
RF Coaxial Cable (DC-40GHz)
RF Coaxial Cable (DC-18GHz)
Notch Filter
Anechoic Chamber

MRE002 MRE003 22290045 22290046 22120181
N/A N/A

CLU18
CLU18
QA360-40-K K-0.5
QA360-40-K KF-2
QA500-18-N N-5
WRCG-24002483.5-60SS
9m*6m*6m

Manufacturer Agilent
Schwarzbeck Schwarzbeck Schwarzbeck Schwarzbeck LUCIX CORP. LUCIX CORP.
Decentest
Pasternack
Pasternack Pasternack Qualwave Qualwave Qualwave Wainwright
CRT

Cal. Date 2023.06.21 2023.07.01 2023.06.26 2023.07.01 2023.07.01 2023.06.27 2023.06.27 2023.07.04
2023.06.27
2023.06.27 2023.06.27 2023.07.04 2023.07.04 2023.07.04
N/A 2022.05.10

Due Date 2024.06.20 2024.06.30 2024.06.25 2024.06.30 2024.06.30 2024.06.26 2024.06.26 2024.07.03
2024.06.26
2024.06.26 2024.06.26 2024.07.03 2024.07.03 2024.07.03
N/A 2025.05.09

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 7 of 56

1.3. Measurement Uncertainty

REPORT No.SZ24020013W01

Test Items Peak Output Power Power Spectral Density Bandwidth Conducted Spurious Emission Restricted Frequency Bands Radiated Emission Conducted Emission
1.4. Testing Laboratory

Uncertainty ±2.22dB ±2.22dB ±5% ±2.77dB ±5% ±2.95dB ±2.44dB

Remark Confidence levels of 95% Confidence levels of 95% Confidence levels of 95% Confidence levels of 95% Confidence levels of 95% Confidence levels of 95% Confidence levels of 95%

Laboratory Name
Laboratory Address
Telephone Facsimile FCC Designation Number FCC Test Firm Registration Number

Shenzhen Morlab Communications Technology Co., Ltd. FL.3, Building A, FeiYang Science Park, No.8 LongChang Road, Block 67, BaoAn District, ShenZhen, GuangDong Province, P. R. China +86 755 36698555 +86 755 36698525 CN1192
226174

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 8 of 56

2. General Description

REPORT No.SZ24020013W01

2.1. Information of Applicant and Manufacturer

Applicant Applicant Address Manufacturer Manufacturer Address

Adec & Partner AG Staldenbachstrasse 30 CH-8808 Pfaffikon, Switzerland Humantechnik GmbH Im Wörth 25D-79576 Weil am Rhein

2.2. Information of EUT

Product Name:

earis MAX

Sample No.:

4#

Hardware Version:

HM01MB-V01

Software Version:

HM-01-V2.9

Equipment Type:

Bluetooth LE

Bluetooth Version:

5.3

Modulation Type:

GFSK

Data Rate:

1Mbps

Operating Frequency Range: 2402MHz-2480MHz

Antenna Type:

Chip Antenna

Antenna Gain:

2.25dBi

Battery (Headphones)

Brand Name: N/A

Model No.:

652936K

Accessory Information:

Serial No.: Capacity:

N/A 720mAh

Rated Voltage: 3.7V

Charge Limit: 4.2V

Manufacturer:

ZHONGSHAN ZHONGWANGDE NEW ENERGY TECHNOLOGY CO.,LTD

Note 1: We use the dedicated software to control the EUT continuous transmission.

Note 2: For a more detailed description, please refer to Specification or User's Manual supplied by

the applicant and/or manufacturer.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 9 of 56

2.3. Channel List of EUT

REPORT No.SZ24020013W01

Channel

Frequency (MHz)

Channel

Frequency (MHz)

Channel

Frequency (MHz)

Channel

Frequency (MHz)

0

2402

10

2422

20

2442

30

2462

1

2404

11

2424

21

2444

31

2464

2

2406

12

2426

22

2446

32

2466

3

2408

13

2428

23

2448

33

2468

4

2410

14

2430

24

2450

34

2470

5

2412

15

2432

25

2452

35

2472

6

2414

16

2434

26

2454

36

2474

7

2416

17

2436

27

2456

37

2476

8

2418

18

2438

28

2458

38

2478

9

2420

19

2440

29

2460

39

2480

Note 1: The black bold channels were selected for test.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 10 of 56

2.4. Test Configuration of EUT

REPORT No.SZ24020013W01

Test mode is used to control the EUT under the maximum power level during test.
2.5. Test Conditions

Temperature (°C) Relative Humidity (%) Atmospheric Pressure (kPa)

15-35 30-60 86-106

2.6. Test Setup Layout Diagram

2.6.1.Conducted Measurement

2.6.2.Conducted Emission Measurement

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 11 of 56

2.6.3.Radiation Measurement 1) For radiated emissions from 9kHz to 30MHz

REPORT No.SZ24020013W01

2) For radiated emissions from 30MHz to1GHz

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 12 of 56

3) For radiated emissions above 1GHz

REPORT No.SZ24020013W01

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 13 of 56

3. Test Results

REPORT No.SZ24020013W01

3.1. Antenna Requirement

3.1.1.Requirement

According to FCC 15.203, an intentional radiator shall be designed to ensure that no antenna other than that furnished by the responsible party shall be used with the device. The use of a permanently attached antenna or of an antenna that uses a unique coupling to the intentional radiator shall be considered sufficient to comply with the provisions of this section.

3.1.2.Test Result
The EUT has a permanently and irreplaceable attached antenna. Please refer to the EUT internal photos.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 14 of 56

3.2. Duty Cycle of Test Signal

REPORT No.SZ24020013W01

3.2.1.Requirement
Preferably, all measurements of maximum conducted (average) output power will be performed with the EUT transmitting continuously (i.e., with a duty cycle of greater than or equal to 98%). When continuous operation cannot be realized, then the use of sweep triggering/signal gating techniques can be used to ensure that measurements are made only during transmissions at the maximum power control level. Such sweep triggering/signal gating techniques will require knowledge of the minimum transmission duration(T) over which the transmitter is on and is transmitting at its maximum power control level for the tested mode of operation. Sweep triggering/signal gating techniques can then be used if the measurement/sweep time of the analyzer can be set such that it does not exceed T at any time that data are being acquired (i.e.,no transmitter OFF-time is to be considered). When continuous transmission cannot be achieved and sweep triggering/signal gating cannot be implemented, alternative procedures are provided that can be used to measure the average power; however, they will require an additional measurement of the transmitter duty cycle (D). Within this sub clause, the duty cycle refers to the fraction of time over which the transmitter is ON and is transmitting at its maximum power control level. The duty cycle is considered to be constant if variations are less than ±2%; otherwise, the duty cycle is considered to be non constant.

3.2.2.Test Result Refer to Annex A.1 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 15 of 56

REPORT No.SZ24020013W01
3.3. Maximum Peak Conducted Output Power
3.3.1.Requirement According to FCC section 15.247(b)(3), For systems using digital modulation in the 902-928 MHz, 2400-2483.5 MHz, and 5725-5850 MHz bands: The maximum peak conducted output power of the intentional radiator shall not exceed 1 Watt. 3.3.2.Test Procedures KDB 558074 Section 8.3.1 was used in order to prove compliance.
3.3.3.Test Setup Layout Refer to chapter 2.6.1 in this report.
3.3.4.Test Result Refer to Annex A.2 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 16 of 56

REPORT No.SZ24020013W01
3.4. Maximum Average Conducted Output Power
3.4.1.Requirement According to FCC section 15.247(b)(3), for systems using digital modulation in the 902-928 MHz, 2400-2483.5 MHz, and 5725-5850 MHz bands: The maximum average conducted output power of the intentional radiator shall not exceed 1 Watt. 3.4.2.Test Procedures KDB 558074 Section 8.3.2 was used in order to prove compliance.
3.4.3.Test Setup Layout Refer to chapter 2.6.1 in this report.
3.4.4.Test Result Refer to Annex A.3 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 17 of 56

3.5. 6 dB Bandwidth

REPORT No.SZ24020013W01

3.5.1.Requirement
According to FCC section 15.247(a) (2), systems using digital modulation techniques may operate in the 902 - 928 MHz, 2400 - 2483.5 MHz, and 5725 - 5850 MHz bands. The minimum 6dB bandwidth shall be at least 500 kHz.

3.5.1.Test Procedures
The steps for the first option are as follows: a) Set analyzer center frequency to channel center frequency b) Set RBW to100kHz c) Set VBW to 300kHz d) Detector = peak. e) Trace mode = max hold f) Sweep time = auto couple g) Allow the trace to fully stabilize h) Measure the maximum width of the emission that is constrained by the frequencies associated with the two outermost amplitude points (upper and lower frequencies) that are attenuated by6 dB relative to the maximum level measured in the fundamental emission The automatic bandwidth measurement capability of an instrument may be employed using the X dB bandwidth mode with X set to 6 dB, if the functionality described in 11.8.1 (i.e., RBW = 100 kHz, VBW  3 ×RBW, and peak detector with maximum hold) is implemented by the instrumentation function. When using this capability, care shall be taken so that the bandwidth measurement is not influenced by any intermediate power nulls in the fundamental emission that might be 6 dB.

3.5.2.Test Setup Layout Refer to chapter 2.6.1 in this report.

3.5.3.Test Result Refer to Annex A.4 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 18 of 56

REPORT No.SZ24020013W01
3.6. Conducted Spurious Emissions and Band Edge
3.6.1.Requirement
According to FCC section 15.247(d), in any 100kHz bandwidth outside the frequency band in which the spread spectrum or digitally modulated intentional radiator is operating, the radio frequency power that is produced by the intentional radiator shall be at least 20dB below that in the 100kHz bandwidth within the band that contains the highest level of the desired power, based on either an RF conducted or a radiated measurement.
3.6.2.Test Procedures
KDB 558074 Section 8.5 and 8.7 was used in order to prove compliance.
3.6.3.Test Setup Layout
Refer to chapter 2.6.1 in this report.
3.6.4.Test Result Refer to Annex A.5 and A.6 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 19 of 56

3.7. Power Spectral Density

REPORT No.SZ24020013W01

3.7.1.Requirement
For digitally modulated systems, the power spectral density conducted from the intentional radiator to the antenna shall not be greater than 8dBm in any 3 kHz band during any time interval of continuous transmission. This power spectral density shall be determined in accordance with the provisions of paragraph (b) of this section. The same method of determining the conducted output power shall be used to determine the power spectral density.

3.7.2.Test Procedures
The measured power spectral density was calculated by the reading of the spectrum analyzer and calibration. Following is the test procedure for PSD test: a) Set analyzer center frequency to channel center frequency b) Set span to1.5 times DTS c) Set RBW to 3kHz d) Set VBW to 10kHz e) Detector = peak f) Sweep time = auto couple g) Trace mode = max hold h) Allow trace to fully stabilize i) Use the peak marker function to determine the maximum amplitude level within the RBW

3.7.3.Test Setup Layout Refer to chapter 2.6.1 in this report.

3.7.4.Test Result Refer to Annex A.7 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 20 of 56

3.8. Conducted Emission

REPORT No.SZ24020013W01

3.8.1.Requirement

According to FCC section 15.207, for an intentional radiator that is designed to be connected to the

public utility (AC) power line, the radio frequency voltage that is conducted back onto the AC power

line on any frequency within the band 150kHz to 30MHz shall not exceed the limits in the following

table, as measured using a 50µH/50 line impedance stabilization network (LISN).

Frequency Range (MHz)

Conducted Limit (dBµV)

Quai-peak

Average

0.15 - 0.50

66 to 56

56 to 46

0.50 - 5

56

46

5 - 30

60

50

Note:

(a) The lower limit shall apply at the band edges.

(b) The limit decreases linearly with the logarithm of the frequency in the range 0.15 - 0.50MHz.

3.8.2.Test Procedures

The Table-top EUT was placed upon a non-metallic table 0.8m above the horizontal metal reference ground plane. EUT was connected to LISN and LISN was connected to reference Ground Plane. EUT was 80cm from LISN. The set-up and test methods were according to ANSI C63.10: 2013.

3.8.3.Test Setup Layout Refer to chapter 2.6.2 in this report.

3.8.4.Test Result Refer to Annex A.8 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 21 of 56

3.9. Restricted Frequency Bands

REPORT No.SZ24020013W01

3.9.1.Requirement
According to FCC section 15.247(d), in any 100kHz bandwidth outside the frequency band in which the spread spectrum or digitally modulated intentional radiator is operating, the radio frequency power that is produced by the intentional radiator shall be at least 20dB below that in the 100kHz bandwidth within the band that contains the highest level of the desired power, In addition, radiated emissions which fall in the restricted bands, as defined in 15.205(a), must also comply with the radiated emission limits specified in 15.209(a).

3.9.2.Test Procedures
The EUT is located in a 3m Semi-Anechoic Chamber; the antenna factors, cable loss and so on of the site as factors are calculated to correct the reading. For the Test Antenna: Test Antenna is 3m away from the EUT. Test Antenna height is varied from 1m to 4m above the ground to determine the maximum value of the field strength. Span = wide enough to fully capture the emission being measured RBW = 1 MHz for f  1GHz, 100 kHz for f < 1GHz VBW = 3 MHz Sweep = auto Detector function = peak/average Trace = max hold Allow the trace to stabilize

3.9.3.Test Setup Layout

Refer to chapter 2.6.3 in this report.

3.9.4.Test Result Refer to Annex A.8 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 22 of 56

3.10. Radiated Emission

REPORT No.SZ24020013W01

3.10.1.Requirement

According to FCC section 15.247(d), radiated emission outside the frequency band attenuation below the general limits specified in FCC section 15.209(a) is not required. In addition, radiated emissions which fall in the restricted bands, as defined in FCC section 15.205(a), must also comply with the radiated emission limits specified in FCC section 15.209(a).

According to FCC section 15.209 (a), except as provided elsewhere in this subpart, the emissions

from an intentional radiator shall not exceed the field strength levels specified in the following table:

Frequency (MHz)

Field Strength (µV/m)

Measurement Distance (m)

0.009 - 0.490

2400/F(kHz)

300

0.490 - 1.705

24000/F(kHz)

30

1.705 - 30.0

30

30

30 - 88

100

3

88 - 216

150

3

216 - 960

200

3

Above 960

500

3

Note1: For above 1000MHz, the emission limit in this paragraph is based on measurement

instrumentation employing an average detector, measurement using instrumentation with a peak

detector function, corresponding to 20dB above the maximum permitted average limit.

Note2:For above 1000MHz, limit field strength of harmonics: 54dBuV/m@3m (AV) and

74dBuV/m@3m (PK).In addition, radiated emissions which fall in the restricted bands, as defined

in Section 15.205(a), also should comply with the radiated emission limits specified in Section

15.209(a)(above table).

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 23 of 56

REPORT No.SZ24020013W01
3.10.2.Test Procedures
The EUT is placed on a non-conducting table 80 cm above the ground plane for measurement below 1GHz; 1.5 m above the ground plane for measurement above 1GHz.The antenna to EUT distance is 3meters. The EUT is configured in accordance with ANSI C63.10. The EUT is set to transmit in a continuous mode.
For measurements below 30MHz, the emission limits shown in the above table are based on measurements employing a CISPR quasi-peak detector except for the frequency bands 9kHz-90 kHz, 110kHz-490 kHz. Radiated emission limits in these two bands are based on measurements employing an average detector.
For measurements below 1GHz the resolution bandwidth is set to 100kHz for peak detection measurements or 120kHz for quasi-peak detection measurements. Peak detection is used unless otherwise noted as quasi-peak.
For measurements above 1GHz the resolution bandwidth is set to 1MHz, the video band width is set to 3MHz for peak measurements and as applicable for average measurements.
The EUT is rotated through 360 degrees to maximize emissions received. The antenna is scanned from 1 to 4 meters above the ground plane to further maximize the emission. Measurements are made with the antenna polarized in both the vertical and the horizontal positions. For measurements above 1 GHz, keeping the measurement antenna aimed at the source of emissions at each frequency of significant emissions, with polarization oriented for maximum response.
3.10.3.Test Setup Layout
Refer to chapter 2.6.3 in this report.
3.10.4.Test Result
Refer to Annex A.9 in this report.

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 24 of 56

REPORT No.SZ24020013W01
Annex A Test Data and Result

A.1. Duty Cycle of Test Signal

Condition NVNT NVNT NVNT

Mode BLE 1M BLE 1M BLE 1M

Frequency (MHz) 2402 2440 2480

Antenna Ant1 Ant1 Ant1

Duty Cycle (%) 61.29 61.29 60.32

Correction Factor (dB) 2.13 2.13 2.2

1/T (kHz) 2.63 2.63 2.63

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 25 of 56

REPORT No.SZ24020013W01
Test Graphs Duty Cycle NVNT BLE 1M 2402MHz Ant1
Duty Cycle NVNT BLE 1M 2440MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 26 of 56

REPORT No.SZ24020013W01
Duty Cycle NVNT BLE 1M 2480MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 27 of 56

REPORT No.SZ24020013W01

A.2. Maximum Peak Conducted Output Power

Condition
NVNT NVNT NVNT

Mode
BLE 1M BLE 1M BLE 1M

Frequency (MHz)

Antenna

Conducted Power (dBm)

Duty Factor
(dB)

Total Conducted
Power (dBm)

Total Conducted
Power (W)

Limit Conducted
(dBm)

Verdict

2402

Ant1

7.31

0

7.31

0.00538

30

Pass

2440

Ant1

7.09

0

7.09

0.00512

30

Pass

2480

Ant1

6.74

0

6.74

0.00472

30

Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 28 of 56

REPORT No.SZ24020013W01
Test Graphs Peak Power NVNT BLE 1M 2402MHz Ant1
Peak Power NVNT BLE 1M 2440MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 29 of 56

REPORT No.SZ24020013W01
Peak Power NVNT BLE 1M 2480MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 30 of 56

REPORT No.SZ24020013W01

A.3. Maximum Average Conducted Output Power

Condition
NVNT NVNT NVNT

Mode
BLE 1M BLE 1M BLE 1M

Frequency (MHz)

Antenna

Conducted Power (dBm)

Duty Factor
(dB)

Total Conducted
Power (dBm)

Total Conducted
Power (W)

Limit Conducted
(dBm)

Verdict

2402

Ant1

4.85

2.13

6.98

0.00499

30

Pass

2440

Ant1

4.58

2.13

6.71

0.00469

30

Pass

2480

Ant1

4.2

2.2

6.4

0.00437

30

Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 31 of 56

REPORT No.SZ24020013W01
Test Graphs Average Power NVNT BLE 1M 2402MHz Ant1
Average Power NVNT BLE 1M 2440MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 32 of 56

REPORT No.SZ24020013W01
Average Power NVNT BLE 1M 2480MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 33 of 56

REPORT No.SZ24020013W01

A.4. 6 dB Bandwidth

Condition Mode

NVNT NVNT NVNT

BLE 1M BLE 1M BLE 1M

Frequency (MHz) 2402 2440 2480

Antenna
Ant1 Ant1 Ant1

-6 dB Bandwidth (MHz) 0.704 0.703 0.71

Limit -6 dB Bandwidth (MHz)
0.5 0.5 0.5

Verdict
Pass Pass Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 34 of 56

REPORT No.SZ24020013W01
Test Graphs -6dB Bandwidth NVNT BLE 1M 2402MHz Ant1
-6dB Bandwidth NVNT BLE 1M 2440MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 35 of 56

REPORT No.SZ24020013W01
-6dB Bandwidth NVNT BLE 1M 2480MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 36 of 56

REPORT No.SZ24020013W01

A.5. Conducted Spurious Emissions

Condition NVNT NVNT NVNT

Mode BLE 1M BLE 1M BLE 1M

Frequency (MHz) 2402 2440 2480

Antenna Ant1 Ant1 Ant1

Max Value (dBc) -35.42 -35.63 -34.74

Limit (dBc) -20 -20 -20

Verdict Pass Pass Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 37 of 56

REPORT No.SZ24020013W01
Test Graphs Tx. Spurious NVNT BLE 1M 2402MHz Ant1 Ref
Tx. Spurious NVNT BLE 1M 2402MHz Ant1 Emission

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 38 of 56

REPORT No.SZ24020013W01
Tx. Spurious NVNT BLE 1M 2440MHz Ant1 Ref
Tx. Spurious NVNT BLE 1M 2440MHz Ant1 Emission

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 39 of 56

REPORT No.SZ24020013W01
Tx. Spurious NVNT BLE 1M 2480MHz Ant1 Ref
Tx. Spurious NVNT BLE 1M 2480MHz Ant1 Emission

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 40 of 56

REPORT No.SZ24020013W01

A.6. Band Edge

Condition NVNT NVNT

Mode BLE 1M BLE 1M

Frequency (MHz) 2402 2480

Antenna Ant1 Ant1

Max Value (dBc) -52.22 -51.78

Limit (dBc) -20 -20

Verdict Pass Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 41 of 56

REPORT No.SZ24020013W01
Test Graphs Band Edge NVNT BLE 1M 2402MHz Ant1 Ref
Band Edge NVNT BLE 1M 2402MHz Ant1 Emission

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 42 of 56

REPORT No.SZ24020013W01
Band Edge NVNT BLE 1M 2480MHz Ant1 Ref
Band Edge NVNT BLE 1M 2480MHz Ant1 Emission

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 43 of 56

REPORT No.SZ24020013W01

A.7. Power Spectral Density

Condition
NVNT NVNT NVNT

Mode
BLE 1M BLE 1M BLE 1M

Frequency (MHz)
2402 2440 2480

Antenna
Ant1 Ant1 Ant1

Conducted PSD
(dBm/3kHz) -8.09 -8.31 -8.66

Duty Factor
(dB) 0 0 0

Total PSD (dBm/3kHz)
-8.09 -8.31 -8.66

Limit (dBm/3kHz)
8 8 8

Verdict
Pass Pass Pass

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 44 of 56

REPORT No.SZ24020013W01
Test Graphs PSD NVNT BLE 1M 2402MHz Ant1
PSD NVNT BLE 1M 2440MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 45 of 56

REPORT No.SZ24020013W01
PSD NVNT BLE 1M 2480MHz Ant1

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 46 of 56

REPORT No.SZ24020013W01
A.8. Conducted Emission
The maximum conducted interference is searched using Peak (PK), if the emission levels more than the AV and QP limits, and that have narrow margins from the AV and QP limits will be re-measured with AV and QP detectors. Tests for both L phase and N phase lines of the power mains connected to the EUT are performed. Set RBW=9kHz, VBW=30kHz. Refer to recorded points and plots below. Note: Both of the test voltage AC 120V/60Hz and AC 230V/50Hz were considered and tested respectively, only the results of the worst case AC 120V/60Hz were recorded in this report. A. Test Setup: Test Mode: EUT+PC+PC Adapter + BT TX Test voltage: AC 120V/60Hz The measurement results are obtained as below: E [dBV] =UR + LCable loss [dB] + AFactor UR: Receiver Reading AFactor: Voltage division factor of LISN

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 47 of 56

B. Test Plot:

REPORT No.SZ24020013W01

(L Phase)

No.

Fre. (MHz)

Emission Level (dBµV) Quai-peak Average

1 0.1500

49.93

34.48

2 0.1635

47.92

33.35

3 0.1905

43.06

29.66

4 0.5955

29.59

24.20

5 18.4327 29.59

28.33

6 24.5762 35.14

33.78

Limit (dBµV)

Quai-peak 66.00 65.28 64.01 56.00 60.00 60.00

Average 56.00 55.28 54.01 46.00 50.00 50.00

Power-line Verdict

PASS

PASS

Line

PASS PASS

PASS

PASS

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 48 of 56

REPORT No.SZ24020013W01

(N Phase)

No.

Fre. (MHz)

Emission Level (dBµV) Quai-peak Average

1 0.1590

48.83

35.08

2 0.1770

45.16

31.26

3 3.7050

26.10

16.27

4 13.8982 26.69

20.76

5 18.4317 29.20

28.34

6 24.5754 34.67

33.42

Limit (dBµV)

Quai-peak 65.52 64.62 56.00 60.00 60.00 60.00

Average 55.52 54.62 46.00 50.00 50.00 50.00

Power-line Verdict

Neutral

PASS PASS PASS PASS PASS PASS

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 49 of 56

REPORT No.SZ24020013W01

A.9. Restricted Frequency Bands

The lowest and highest channels are tested to verify the Restricted Frequency Bands. The measurement results are obtained as below: E [dBV/m] =UR + AT + AFactor [dB]; AT =LCable loss [dB]-Gpreamp [dB] AT: Total correction Factor except Antenna UR: Receiver Reading Gpreamp: Preamplifier Gain AFactor: Antenna Factor at 3m
Note: Restricted Frequency Bands were performed when antenna was at vertical and horizontal polarity, and only the worse test condition (vertical) was recorded in this test report.

Receiver

Max.

Channel

Frequency (MHz)

Detector PK/ AV

Reading UR
(dBµV)

AT (dB)

AFactor (dB@3m)

Emission E
(dBµV/m)

Limit (dBµV/m)

Verdict

0

2386.42

PK

23.88 6.74 27.20

57.82

74

PASS

0

2390.00

AV

11.64 6.74 27.20

45.58

54

PASS

39

2484.53

PK

23.60 6.74 27.20

57.54

74

PASS

39

2486.16

AV

13.62 6.74 27.20

47.56

54

PASS

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 50 of 56

REPORT No.SZ24020013W01
(PEAK, Channel 0)

(AVERAGE, Channel 0)

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 51 of 56

REPORT No.SZ24020013W01

(PEAK, Channel 39)

(AVERAGE, Channel 39)

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 52 of 56

REPORT No.SZ24020013W01

A.10. Radiated Emission
According to ANSI C63.10, because of peak detection will yield amplitudes equal to or greater than amplitudes measured with the quasi-peak (or average) detector, the measurement data from a spectrum analyzer peak detector will represent the worst-case results, if the peak measured value complies with the quasi-peak (or average) limit, it is unnecessary to perform an quasi-peak measurement (or average).
The measurement results are obtained as below: E [dBV/m] =UR + AT + AFactor [dB]; AT =LCable loss [dB]-Gpreamp [dB] AT: Total correction Factor except Antenna UR: Receiver Reading Gpreamp: Preamplifier Gain AFactor: Antenna Factor at 3m During the test, the total correction Factor AT and AFactor were built in test software. Note1: All radiated emission tests were performed in X, Y, Z axis direction. And only the worst axis test condition was recorded in this test report. Note2: For the frequency, which started from 9kHz to 30MHz, was pre-scanned and the result which was 20dB lower than the limit was not recorded. Note3: For the frequency, which started from 18GHz to 10th harmonic of the highest frequency, was pre-scanned and the result which was 20dB lower than the limit was not recorded.

Field strength of fundamental:

Frequency Reading_Peak Antenna

Path Loss Final_Peak

Antenna

(MHz)

(dBV/m)

Factor (dB)

(dB)

(dBV/m)

Polarity

2402.25

79.68

27.20

6.74

113.62

Vertical

The field strength (the lowest) of fundamenta is more than 20dB higher than the unwanted

emissions, in accordance with FCC part 15.215(b).

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 53 of 56

Plot for Channel 0

REPORT No.SZ24020013W01

Fre. (MHz)
111.480 2401.600 4802.200 7204.600 12012.480 13493.960

PK (dBV/m)
32.27
49.35 54.40
48.97
48.15 47.06

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

47.89

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Horizontal, 30MHz to 18GHz)

Antenna
Horizontal Horizontal Horizontal Horizontal Horizontal Horizontal

Verdict
PASS PASS PASS PASS PASS PASS

Fre. (MHz)
101.780 2401.600 4802.200 5476.720 7204.600 13952.880

PK (dBV/m)
32.04 51.11
60.40
49.55 48.57
47.30

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

49.85

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Vertical, 30MHz to 18GHz)

Antenna
Vertical Vertical Vertical Vertical Vertical Vertical

Verdict
PASS PASS PASS PASS PASS PASS

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 54 of 56

Plot for Channel 19

REPORT No.SZ24020013W01

Fre. (MHz)
101.780 2406.400 4879.200 5969.520 7321.640 14245.480

PK (dBV/m)
32.17
48.37 57.90
49.53
50.23 47.39

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

49.61

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Horizontal, 30MHz to 18GHz)

Antenna
Horizontal Horizontal Horizontal Horizontal Horizontal Horizontal

Verdict
PASS PASS PASS PASS PASS PASS

Fre. (MHz)
104.690 2409.067 3989.080 5227.240 9376.000 13413.880

PK (dBV/m)
32.77 48.57
47.67
48.90 47.03
46.92

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Vertical, 30MHz to 18GHz)

Antenna
Vertical Vertical Vertical Vertical Vertical Vertical

Verdict
PASS PASS PASS PASS PASS PASS

MORLAB

Shenzhen Morlab Communications Technology Co., Ltd. FL.1-3, Building A, FeiYang Science Park, No.8 LongChang Road, Block67, BaoAn District, ShenZhen, GuangDong Province, P. R. China

Tel: 86-755-36698555 Http://www.morlab.cn

Fax: 86-755-36698525 E-mail: [email protected]
Page 55 of 56

Plot for Channel 39

REPORT No.SZ24020013W01

Fre. (MHz)
108.570 2479.467 3129.760 4959.280 7438.680 13481.640

PK (dBV/m)
32.87
50.26
44.72 57.96
49.28
46.53

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

49.38

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Horizontal, 30MHz to 18GHz)

Antenna
Horizontal Horizontal Horizontal Horizontal Horizontal Horizontal

Verdict
PASS PASS PASS PASS PASS PASS

Fre. (MHz)
107.600 2479.467 3148.240 4959.280 9040.280 14045.280

PK (dBV/m)
32.76
48.32
44.05 51.68
46.31
46.64

QP

AV

Limit-PK Limit-QP Limit-AV

(dBV/m) (dBV/m) (dBV/m) (dBV/m) (dBV/m)

N/A

N/A

N/A

43.50

N/A

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

44.45

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

N/A

N/A

74.00

N/A

54.00

(Antenna Vertical, 30MHz to 18GHz)

Antenna
Vertical Vertical Vertical Vertical Vertical Vertical

Verdict
PASS PASS PASS PASS PASS PASS