ZDL

Test Report

JOYTECH HEALTHCARE CO., LTD. 0021 Fingertip Pulse Oximeter 2AQVU0021 2AQVU0021 0021

PDF Viewing Options

Not Your Device? Search For Manuals or Datasheets below:


File Info : application/pdf, 46 Pages, 2.33MB

Document DEVICE REPORTGetApplicationAttachment.html?id=6130197
MRT Technology (Suzhou) Co., Ltd Phone: +86-512-66308358 Web: www.mrt-cert.com

Report No.: 2207RSU031-U2

Report Version:

V01

Issue Date:

2022-09-09

RF MEASUREMENT REPORT

FCC ID:

2AQVU0021

Applicant:

JOYTECH HEALTHCARE CO., LTD.

Address: Product:

No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China Fingertip Pulse oximeter

Model No.:

XM-103

FCC Classification: Digital Transmission System (DTS)

FCC Rule Part(s): Part 15 Subpart C (Section 15.247)

Result:

Complies

Test Date:

2022-07-27 ~ 2022-07-28

Reviewed By: Approved By:

Sunny Sun Robin Wu

The test results relate only to the samples tested. This equipment has been shown to be capable of compliance with the applicable technical standards as indicated in the measurement report and was tested in accordance with the measurement procedures specified in ANSI C63.10-2013. Test results reported herein relate only to the item(s) tested.
The test report shall not be reproduced except in full without the written approval of MRT Technology (Suzhou) Co., Ltd.
1 of 46

Report No. 2207RSU031-U2

Version Rev. 01

Revision History
Description Initial Report

Report No.: 2207RSU031-U2

Issue Date 2022-09-09

Note Valid

2 of 46

Report No.: 2207RSU031-U2

CONTENTS

Description

Page

1. General Information ................................................................................................................................5

1.1. Applicant ...................................................................................................................................... 5 1.2. Manufacturer ................................................................................................................................ 5 1.3. Testing Facility ............................................................................................................................. 5 1.4. Product Information...................................................................................................................... 6 1.5. Radio Specification under Test .................................................................................................... 6 1.6. Working Frequencies ................................................................................................................... 7

2. Test Configuration ...................................................................................................................................8

2.1. Test System Connection Diagram................................................................................................ 8 2.2. Test Software ............................................................................................................................... 8 2.3. Applied Standards........................................................................................................................ 9 2.4. Test Environment Condition ......................................................................................................... 9

3. Antenna Requirements .........................................................................................................................10

4. Measuring Instrument ........................................................................................................................... 11

5. Decision Rules and Measurement Uncertainty ..................................................................................12

5.1. Decision Rules ........................................................................................................................... 12 5.2. Measurement Uncertainty .......................................................................................................... 12

6. Test Result..............................................................................................................................................13

6.1. 6.2. 6.2.1. 6.2.2. 6.2.3. 6.2.4. 6.2.5. 6.3. 6.3.1. 6.3.2. 6.3.3. 6.3.4. 6.3.5. 6.4. 6.4.1. 6.4.2. 6.4.3.

Summary .................................................................................................................................... 13 6dB Bandwidth Measurement.................................................................................................... 14 Test Limit .................................................................................................................................... 14 Test Procedure ........................................................................................................................... 14 Test Setting ................................................................................................................................ 14 Test Setup .................................................................................................................................. 14 Test Result ................................................................................................................................. 14 Output Power Measurement ...................................................................................................... 15 Test Limit .................................................................................................................................... 15 Test Procedure ........................................................................................................................... 15 Test Setting ................................................................................................................................ 15 Test Setup .................................................................................................................................. 15 Test Result ................................................................................................................................. 15 Power Spectral Density Measurement ...................................................................................... 16 Test Limit .................................................................................................................................... 16 Test Procedure ........................................................................................................................... 16 Test Setting ................................................................................................................................ 16

3 of 46

Report No.: 2207RSU031-U2

6.4.4. 6.4.5. 6.5. 6.5.1. 6.5.2. 6.5.3. 6.5.4. 6.5.5. 6.6. 6.6.1. 6.6.2. 6.6.3. 6.6.4. 6.6.5. 6.7. 6.7.1. 6.7.2. 6.7.3. 6.7.4. 6.7.5. 6.8. 6.8.1. 6.8.2. 6.8.3.

Test Setup .................................................................................................................................. 16 Test Result ................................................................................................................................. 16 Conducted Band Edge and Out-of-Band Emissions Measurement .......................................... 17 Test Limit .................................................................................................................................... 17 Test Procedure ........................................................................................................................... 17 Test Settitng ............................................................................................................................... 17 Test Setup .................................................................................................................................. 17 Test Result ................................................................................................................................. 18 Radiated Spurious Emission Measurement............................................................................... 19 Test Limit .................................................................................................................................... 19 Test Procedure ........................................................................................................................... 19 Test Setting ................................................................................................................................ 19 Test Setup .................................................................................................................................. 21 Test Result ................................................................................................................................. 22 Radiated Restricted Band Edge Measurement ......................................................................... 23 Test Limit .................................................................................................................................... 23 Test Procedure ........................................................................................................................... 24 Test Setting ................................................................................................................................ 24 Test Setup .................................................................................................................................. 25 Test Result ................................................................................................................................. 25 AC Conducted Emissions Measurement ................................................................................... 26 Test Limit .................................................................................................................................... 26 Test Setup .................................................................................................................................. 26 Test Result ................................................................................................................................. 26

Appendix A - Test Result................................................................................................................................27

A.1 Duty Cycle Test Result ............................................................................................................... 27 A.2 6dB Bandwidth Test Result ........................................................................................................ 28 A.3 Output Power Test Result .......................................................................................................... 29 A.4 Power Spectral Density Test Result ........................................................................................... 30 A.5 Conducted Band Edge and Out-of-Band Emissions Test Result............................................... 31 A.6 Radiated Spurious Emission Test Result ................................................................................... 33 A.7 Radiated Restricted Band Edge Test Result.............................................................................. 36 A.8 AC Conducted Emissions Test Result ....................................................................................... 44

Appendix B - Test Setup Photograph ...........................................................................................................45

Appendix C - EUT Photograph ......................................................................................................................46

4 of 46

Report No.: 2207RSU031-U2
1. General Information
1.1. Applicant JOYTECH HEALTHCARE CO., LTD. No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China

1.2. Manufacturer JOYTECH HEALTHCARE CO., LTD. No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China

1.3. Testing Facility

Test Site ­ MRT Suzhou Laboratory

Laboratory Location (Suzhou - Wuzhong)

D8 Building, No.2 Tian'edang Rd., Wuzhong Economic Development Zone, Suzhou, China

Laboratory Location (Suzhou - SIP)

4b Building, Liando U Valley, No.200 Xingpu Rd., Shengpu Town, Suzhou Industrial Park, China

Laboratory Accreditations

A2LA: 3628.01

CNAS: L10551

FCC: CN1166

ISED: CN0001

VCCI:

R-20025 R-20141

G-20034 G-20134

C-20020 C-20103

T-20020 T-20104

Test Site ­ MRT Shenzhen Laboratory

Laboratory Location (Shenzhen)

1G, Building A, Junxiangda Building, Zhongshanyuan Road West, Nanshan District, Shenzhen, China

Laboratory Accreditations

A2LA: 3628.02

CNAS: L10551

FCC: CN1284

ISED: CN0105

Test Site ­ MRT Taiwan Laboratory

Laboratory Location (Taiwan)

No. 38, Fuxing 2nd Rd., Guishan Dist., Taoyuan City 333, Taiwan (R.O.C.)

Laboratory Accreditations

TAF: L3261-190725

FCC: 291082, TW3261

ISED: TW3261

5 of 46

Report No.: 2207RSU031-U2

1.4. Product Information

Product Name

Fingertip Pulse oximeter

Model No.

XM-103

20220719Sample#04 (Conducted Test) Test Device Identification No.
20220725Sample#03 (Radiated Test)

Bluetooth Version

V5.0 (BLE only)

Hardware Version

Ver Z

Software Version

V1.0

Power Supply

2 x 1.5v AAA batteries

Remark:

The information of EUT was provided by the manufacturer, and the accuracy of the information shall be the

responsibility of the manufacturer.

1.5. Radio Specification under Test

Bluetooth Frequency Channel Number Type of modulation Data Rate Antenna Type Antenna Gain

2402 ~ 2480MHz 40 GFSK 1Mbps PCB Antenna 0 dBi

6 of 46

1.6. Working Frequencies

Channel 00 03 06 09 12 15 18 21 24 27 30 33 36 39

Frequency 2402 MHz 2408 MHz 2414 MHz 2420 MHz 2426 MHz 2432 MHz 2438 MHz 2444 MHz 2450 MHz 2456 MHz 2462 MHz 2468 MHz 2474 MHz 2480 MHz

Channel 01 04 07 10 13 16 19 22 25 28 31 34 37 --

Frequency 2404 MHz 2410 MHz 2416 MHz 2422 MHz 2428 MHz 2434 MHz 2440 MHz 2446 MHz 2452 MHz 2458 MHz 2464 MHz 2470 MHz 2476 MHz
--

Report No.: 2207RSU031-U2

Channel 02 05 08 11 14 17 20 23 26 29 32 35 38 --

Frequency 2406 MHz 2412 MHz 2418 MHz 2424 MHz 2430 MHz 2436 MHz 2442 MHz 2448 MHz 2454 MHz 2460 MHz 2466 MHz 2472 MHz 2478 MHz
--

7 of 46

Report No.: 2207RSU031-U2
2. Test Configuration
2.1. Test System Connection Diagram The device was tested per the guidance ANSI C63.10-2013 was used to reference the appropriate EUT setup for radiated emissions testing. Connection Diagram ­ Radiated Emission testing

Cable Type A USB Cable
Product 1 Notebook

Manufacturer Lenovo

Cable Description Non shielded, > 10m
Model No. E495

2.2. Test Software The test utility software used during testing was "PhyPlusKit", and the version was V2.3.6d.

8 of 46

Report No.: 2207RSU031-U2
2.3. Applied Standards According to the specifications of the manufacturer, the EUT must comply with the requirements of the following standards:  FCC Part 15.247  KDB 558074 D01v05r02  ANSI C63.10-2013

2.4. Test Environment Condition Ambient Temperature Relative Humidity

15 ~ 35°C 20 ~75%RH

9 of 46

Report No.: 2207RSU031-U2
3. Antenna Requirements
Excerpt from §15.203 of the FCC Rules/Regulations: "An intentional radiator antenna shall be designed to ensure that no antenna other than that furnished by the responsible party can be used with the device. The use of a permanently attached antenna or of an antenna that uses a unique coupling to the intentional radiator shall be considered sufficient to comply with the provisions of this section."  The antenna of the device is permanently attached.  There are no provisions for connection to an external antenna. Conclusion: The unit complies with the requirement of §15.203.
10 of 46

Report No.: 2207RSU031-U2

4. Measuring Instrument

Instrument EMI Test Receiver Horn Antenna Preamplifier TRILOG Antenna Anechoic Chamber Thermohygrometer Signal Analyzer Thermohygrometer Loop Antenna Horn Antenna Horn Antenna USB Power Sensor Signal Analyzer Thermohygrometer Shielding Room Attenuator

Manufacturer R&S Schwarzbeck Agilent Schwarzbeck TDK testo Keysight testo Schwarzbeck ETS Schwarzbeck Keysight Keysight testo HUAMING MVE

Model No. ESR7 BBHA 9120D 83017A VULB 9168 WZ-AC1 608-H1 N9010B Testo 608-H1 FMZB 1519 3117 BBHA 9170 U2021XA N9010B 608-H1 WZ-SR5 MVE2213

Asset No. MRTSUE06001 MRTSUE06023 MRTSUE06076 MRTSUE06172 MRTSUE06212 MRTSUE06403 MRTSUE06607 MRTSUE11039 MRTSUE06025 MRTSUE06257 MRTSUE06597 MRTSUE06447 MRTSUE07027 MRTSUE06402 MRTSUE06442 MRTSUE11086

Cali. Interval Cali. Due Date Test Site

1 year

2022-12-29

WZ-AC1

1 year

2022-09-16

WZ-AC1

1 year

2022-11-12

WZ-AC1

1 year

2023-06-21

WZ-AC1

1 year

2023-04-21

WZ-AC1

1 year

2023-06-06

WZ-AC1

1 year

2022-12-29

WZ-AC1

1 year

2022-11-11

WZ-AC1

1 year

2022-10-28

WZ-AC1

1 year

2022-09-25

WZ-AC1

1 year

2022-12-01

WZ-AC1

1 year

2023-06-04

WZ-SR5

1 year

2022-12-05

WZ-SR5

1 year

2023-06-06

WZ-SR5

N/A

N/A

WZ-SR5

1 year

2023-06-09

WZ-SR5

Software

Version

EMI V3

V 3.0.0

Controller_MF 7802

2.03C

Agilent Power Analyzer/Agilent Power Panel V R03.09.00

Function EMI Test Software RE Antenna & Turntable Power

11 of 46

Report No.: 2207RSU031-U2
5. Decision Rules and Measurement Uncertainty
5.1. Decision Rules
The Decision Rule is based on Simple Acceptance in accordance with ISO Guide 98-4: 2012 Clause 8.2. (Measurement uncertainty is not taken into account when stating conformity with a specified requirement.)
5.2. Measurement Uncertainty
Where relevant, the following test uncertainty levels have been estimated for tests performed on the EUT as specified in CISPR 16-4-2. This uncertainty represents an expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor of k = 2. AC Conducted Emission Measurement
Measurement Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): 9kHz~150kHz: 3.74dB 150kHz~30MHz: 3.44dB Radiated Emission Measurement Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): Horizontal: 30MHz~300MHz: 5.04dB 300MHz~1GHz: 4.95dB 1GHz~40GHz: 6.40dB Vertical: 30MHz~300MHz: 5.24dB 300MHz~1GHz: 6.03dB 1GHz~40GHz: 6.40dB Spurious Emissions, Conducted Measuring Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): 0.78dB Output Power Measuring Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): 1.13dB Power Spectrum Density Measuring Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): 1.15dB Occupied Bandwidth Measuring Uncertainty for a Level of Confidence of 95% (U=2Uc(y)): 0.28%
12 of 46

Report No.: 2207RSU031-U2

6. Test Result

6.1. Summary

FCC Section(s) 15.247(a)(2)

Test Description 6dB Bandwidth

Test Condition

Verdict Pass

15.247(b)(3) 15.247(e)

Output Power Power Spectral Density

Conducted

Pass Pass

15.247(d)

Conducted Band Edge / Out-of-Band Emissions

Pass

15.205 15.209

General Field Strength (Restricted Bands and Radiated Emission)

Radiated

Pass

15.207

AC Conducted Emissions 150kHz - 30MHz

Line Conducted N/A

Remark:

1. The analyzer plots shown in this section were all taken with a correction table loaded into the analyzer.

The correction table was used to account for the losses of the cables and attenuators used as part of the

system to connect the EUT to the analyzer at all frequencies of interest.

2. All modes of operation and data rates were investigated. For radiated emission test, every axis (X, Y, Z) was also verified. The test results shown in the following sections represent the worst-case emissions.

3. "N/A" means not applicable, the details refer to relevant section.

13 of 46

Report No.: 2207RSU031-U2 6.2. 6dB Bandwidth Measurement 6.2.1. Test Limit The minimum 6dB bandwidth shall be at least 500 kHz. 6.2.2. Test Procedure ANSI C63.10-2013 - Section 11.8 6.2.3. Test Setting 1. The Spectrum's automatic bandwidth measurement capability was used to perform the 6dB bandwidth
measurement. The "X" dB bandwidth parameter was set to X = 6. The bandwidth measurement was not influenced by any intermediate power nulls in the fundamental emission. 2. Set RBW = 100 kHz 3. VBW  3 × RBW 4. Detector = Peak 5. Trace mode = Max hold 6. Sweep = Auto couple 7. Allow the trace to stabilize 6.2.4. Test Setup
6.2.5. Test Result Refer to Appendix A.2.
14 of 46

Report No.: 2207RSU031-U2
6.3. Output Power Measurement 6.3.1. Test Limit The maximum output power shall be less 1 Watt (30dBm). The conducted output power limit specified in paragraph FCC Part 15.247(b) of this section is based on the use of antennas with directional gains that do not exceed 6 dBi. If transmitting antennas of directional gain greater than 6 dBi are used, the conducted output power from the intentional radiator shall be reduced below the stated values in paragraphs FCC Part 15.247(b)(3) of this section, as appropriate, by the amount in dB that the directional gain of the antenna exceeds 6 dBi. 6.3.2. Test Procedure ANSI C63.10-2013 - Section 11.9.1.3 ANSI C63.10-2013 - Section 11.9.2.3.2 6.3.3. Test Setting Method PKPM1 (Peak Power Measurement of Signals with DTS BW  50MHz) Peak power measurements were performed only when the EUT was transmitting at its maximum power control level using a broadband power meter with a pulse sensor. The pulse sensor employs a VBW = 50MHz so this method was only used for signals whose DTS bandwidth was less than or equal to 50MHz. Average Power Measurement Average power measurements were performed only when the EUT was transmitting at its maximum power control level using a broadband power meter with a pulse sensor. The power meter implemented triggering and gating capabilities which were set up such that power measurements were recorded only during the ON time of the transmitter. 6.3.4. Test Setup

6.3.5. Test Result Refer to Appendix A.3.

15 of 46

Report No.: 2207RSU031-U2
6.4. Power Spectral Density Measurement 6.4.1. Test Limit The maximum permissible power spectral density is 8dBm in any 3 kHz band. The same method of determining the conducted output power shall be used to determine the power spectral density. 6.4.2. Test Procedure
ANSI C63.10-2013 Section 11.10.2
6.4.3. Test Setting 1. Analyzer was set to the center frequency of the DTS channel under investigation 2. Span = 1.5 times the DTS channel bandwidth 3. RBW = 3kHz 4. VBW = 10kHz 5. Detector = peak 6. Sweep time = auto couple 7. Trace mode = max hold 8. Trace was allowed to stabilize 6.4.4. Test Setup
6.4.5. Test Result Refer to Appendix A.4.
16 of 46

Report No.: 2207RSU031-U2
6.5. Conducted Band Edge and Out-of-Band Emissions Measurement 6.5.1. Test Limit The limit for out-of-band spurious emissions at the band edge is 20dB below the fundamental emission level, as determined from the in-band power measurement of the DTS channel performed in a 100 kHz bandwidth per the PSD procedure. 6.5.2. Test Procedure ANSI C63.10-2013 - Section 11.11 6.5.3. Test Settitng Reference level measurement 1. Set instrument center frequency to DTS channel center frequency 2. Set the span to  1.5 times the DTS bandwidth 3. Set the RBW = 100 kHz 4. Set the VBW  3 x RBW 5. Detector = peak 6. Sweep time = auto couple 7. Trace mode = max hold 8. Allow trace to fully stabilize Emission level measurement 1. Set the center frequency and span to encompass frequency range to be measured 2. RBW = 100kHz 3. VBW = 300kHz 4. Detector = Peak 5. Trace mode = max hold 6. Sweep time = auto couple 7. The trace was allowed to stabilize 6.5.4. Test Setup
17 of 46

6.5.5. Test Result Refer to Appendix A.5.

Report No.: 2207RSU031-U2

18 of 46

Report No.: 2207RSU031-U2

6.6. Radiated Spurious Emission Measurement

6.6.1. Test Limit

All out of band emissions appearing in a restricted band as specified in Section 15.205 of the Title 47 CFR must not exceed the limits shown in Table per Section 15.209.

Frequency [MHz]
0.009 - 0.490 0.490 - 1.705
1.705 - 30 30 - 88 88 - 216
216 - 960 Above 960

FCC Part 15 Subpart C Paragraph 15.209 Field Strength [uV/m] 2400/F (kHz) 24000/F (kHz) 30 100 150 200 500

Measured Distance [Meters] 300 30 30 3 3 3 3

6.6.2. Test Procedure ANSI C63.10-2013 - Section 11.11 & 11.12 ANSI C63.10-2013 - Section 6.3 (General Requirements) ANSI C63.10-2013 - Section 6.4 (Standard test method below 30MHz) ANSI C63.10-2013 - Section 6.5 (Standard test method above 30MHz to 1GHz) ANSI C63.10-2013 - Section 6.6 (Standard test method above 1GHz)

6.6.3. Test Setting

Table 1 - RBW as a function of frequency

Frequency

RBW

9 ~ 150 kHz

200 ~ 300 Hz

0.15 ~ 30 MHz

9 ~ 10 kHz

30 ~ 1000 MHz

100 ~ 120 kHz

> 1000MHz

1MHz

19 of 46

Report No.: 2207RSU031-U2
Quasi-Peak Measurements below 1GHz 1. Analyzer center frequency was set to the frequency of the radiated spurious emission of interest 2. Span was set greater than 1MHz 3. RBW = as specified in Table 1 4. Detector = CISPR quasi-peak 5. Sweep time = auto couple 6. Trace was allowed to stabilize Peak Measurements above 1GHz 1. Analyzer center frequency was set to the frequency of the radiated spurious emission of interest 2. RBW = 1MHz 3. VBW = 3MHz 4. Detector = peak 5. Sweep time = auto couple 6. Trace mode = max hold 7. Trace was allowed to stabilize Average Measurements above 1GHz (Method VB) 1. Analyzer center frequency was set to the frequency of the radiated spurious emission of interest 2. RBW = 1MHz 3. VBW; If the EUT is configured to transmit with duty cycle  98%, set VBW = 10 Hz. If the EUT duty cycle is < 98%, set VBW  1/T. T is the minimum transmission duration. 4. Detector = Peak 5. Sweep time = auto 6. Trace mode = max hold 7. Trace was allowed to stabilize
20 of 46

6.6.4. Test Setup Below 1GHz Test Setup:

Report No.: 2207RSU031-U2

Above 1GHz Test Setup:

21 of 46

6.6.5. Test Result Refer to Appendix A.6.

Report No.: 2207RSU031-U2

22 of 46

Report No.: 2207RSU031-U2

6.7. Radiated Restricted Band Edge Measurement

6.7.1. Test Limit

For 15.205 requirement: Radiated emissions which fall in the restricted bands, as defined in Section 15.205(a) of FCC part 15, must also comply with the radiated emission limits specified in Section 15.209(a).

Frequency (MHz)
0.090 - 0.110 1 0.495 - 0.505 2.1735 - 2.1905 4.125 - 4.128 4.17725 - 4.17775 4.20725 - 4.20775 6.215 - 6.218 6.26775 - 6.26825 6.31175 - 6.31225 8.291 - 8.294 8.362 - 8.366 8.37625 - 8.38675 8.41425 - 8.41475 12.29 - 12.293 12.51975 - 12.52025 12.57675 - 12.57725 13.36 - 13.41

Frequency (MHz)
16.42 - 16.423 16.69475 - 16.69525 16.80425 - 16.80475
25.5 - 25.67 37.5 - 38.25
73 - 74.6 74.8 - 75.2 108 - 121.94 123 - 138 149.9 - 150.05 156.52475 - 156.52525 156.7 - 156.9 162.0125 - 167.17 167.72 - 173.2 240 - 285 322 - 335.4
--

Frequency (MHz)
399.9 - 410 608 - 614 960 - 1240 1300 - 1427 1435 - 1626.5 1645.5 - 1646.5 1660 - 1710 1718.8 - 1722.2 2200 - 2300 2310 - 2390 2483.5 - 2500 2690 - 2900 3260 - 3267 3332 - 3339 3345.8 - 3358 3600 - 4400
--

Frequency (GHz)
4.5 - 5.15 5.35 - 5.46 7.25 - 7.75
8.025 - 8.5 9.0 - 9.2 9.3 - 9.5 10.6 - 12.7 13.25 - 13.4 14.47 - 14.5 15.35 - 16.2 17.7 - 21.4 22.01 - 23.12 23.6 - 24.0 31.2 - 31.8 36.43 - 36.5
(2) --

23 of 46

Report No.: 2207RSU031-U2

All out of band emissions appearing in a restricted band as specified in Section 15.205 of the Title 47CFR

must not exceed the limits shown in Table per Section 15.209. FCC Part 15 Subpart C Paragraph 15.209

Frequency [MHz]
0.009 - 0.490 0.490 - 1.705
1.705 - 30 30 - 88 88 - 216
216 - 960 Above 960

Field Strength [uV/m]
2400/F (kHz) 24000/F (kHz)
30 100 150 200 500

Measured Distance [Meters] 300 30 30 3 3 3 3

6.7.2. Test Procedure ANSI C63.10-2013 Section 6.3 & 6.6 & 11.13 6.7.3. Test Setting Peak Field Strength Measurements 1. Analyzer center frequency was set to the frequency of the radiated spurious emission of interest 2. RBW = 1MHz 3. VBW = 3MHz 4. Detector = peak 5. Sweep time = auto couple 6. Trace mode = max hold 7. Trace was allowed to stabilize

Average Field Strength Measurements

1. Analyzer center frequency was set to the frequency of the radiated spurious emission of interest

2. RBW = 1MHz

3. VBW  1/T

4. As an alternative, the instrument may be set to linear detector mode. Ensure that video filtering is applied

in linear voltage domain (rather than in a log or dB domain). Some instruments require linear display

mode in order to accomplish this. Others have a setting for Average-VBW Type, which can be set to

"Voltage" regardless of the display mode

5. Detector = Peak

24 of 46

6. Sweep time = auto 7. Trace mode = max hold 8. Allow max hold to run for at least 50 times (1/duty cycle) traces 6.7.4. Test Setup

Report No.: 2207RSU031-U2

6.7.5. Test Result Refer to Appendix A.7.
25 of 46

Report No.: 2207RSU031-U2

6.8. AC Conducted Emissions Measurement

6.8.1. Test Limit

Frequency (MHz)

FCC Part 15 Subpart C Paragraph 15.207 Limits QP
(dBuV)

AV (dBuV)

0.15 - 0.50

66 - 56

56 - 46

0.50 - 5.0

56

46

5.0 - 30

60

50

Note 1: The lower limit shall apply at the transition frequencies.

Note 2: The limit decreases linearly with the logarithm of the frequency in the range 0.15MHz to 0.5MHz.

6.8.2. Test Setup

6.8.3. Test Result Refer to Appendix A.8.

26 of 46

Appendix A - Test Result

A.1 Duty Cycle Test Result

Test Site Test Date

WZ-SR5 2022-07-28

Test Mode BLE

Report No.: 2207RSU031-U2

Test Engineer

Lynn Yang

Duty Cycle

Duty Cycle 100%

27 of 46

A.2 6dB Bandwidth Test Result

Test Site Test Date

WZ-SR5 2022-07-27

Report No.: 2207RSU031-U2

Test Engineer

Lynn Yang

Test Mode
BLE BLE BLE

Data Rate
1Mbps 1Mbps 1Mbps

Channel No.
00 19 39

Frequency (MHz) 2402 2440 2480

6dB Bandwidth (MHz) 0.7743 0.7636 0.7658

Limit (MHz)  0.5  0.5  0.5

Channel 00 (2402MHz)

BLE 6dB Bandwidth

Channel 19 (2440MHz)

Channel 39 (2480MHz)

28 of 46

A.3 Output Power Test Result

Test Site Test Date

WZ-SR5 2022-07-27

Report No.: 2207RSU031-U2

Test Engineer

Lynn Yang

Test Result of Peak Output Power

Test Mode

Data Rate

Channel No.

BLE

1Mbps

00

BLE

1Mbps

19

BLE

1Mbps

39

Frequency (MHz) 2402 2440 2480

Peak Power (dBm) -0.36 -0.47 -0.88

Limit (dBm)  30.00  30.00  30.00

Result
Pass Pass Pass

Test Result of Average Output Power (Reporting Only)

Test Mode

Data Rate

Channel No.

Frequency Average Power

(MHz)

(dBm)

BLE

1Mbps

00

2402

-0.99

BLE

1Mbps

19

2440

-1.14

BLE

1Mbps

39

2480

-1.53

Limit (dBm)  30.00  30.00  30.00

Result
Pass Pass Pass

29 of 46

A.4 Power Spectral Density Test Result

Test Site Test Date

WZ-SR5 2022-07-27

Report No.: 2207RSU031-U2

Test Engineer

Lynn Yang

Test Mode
BLE BLE BLE

Data Rate
1Mbps 1Mbps 1Mbps

Channel No.
00 19 39

Frequency (MHz) 2402 2440 2480

PSD Result

Limit

(dBm / 3kHz) (dBm / 3kHz)

-16.60

 8.00

-16.67

 8.00

-16.26

 8.00

Result
Pass Pass Pass

Channel 00 (2402MHz)

BLE PSD

Channel 19 (2440MHz)

Channel 39 (2480MHz)

30 of 46

Report No.: 2207RSU031-U2

A.5 Conducted Band Edge and Out-of-Band Emissions Test Result

Test Site

WZ-SR5

Test Engineer

Test Date

2022-07-27 ~ 2022-07-28

Lynn Yang

Test Mode
BLE BLE BLE

Data Rate / Mbps 1 1 1

Channel No.
00 19 39

Frequency (MHz) 2402 2440 2480

Limit (dBc)
20 20 20

Result
Pass Pass Pass

BLE - Out-of-Band Emissions Channel 00 (2402MHz)
100kHz PSD Reference Level

Low Band Edge

Spurious Emission 30MHz ~ 25GHz

31 of 46

Report No.: 2207RSU031-U2

Channel 19 (2440MHz)

100kHz PSD Reference Level

Spurious Emission 30MHz ~ 25GHz

Channel 39 (2480MHz) 100kHz PSD Reference Level

High Band Edge

Spurious Emission 30MHz ~ 25GHz

32 of 46

Report No.: 2207RSU031-U2

A.6 Radiated Spurious Emission Test Result

Test Site

WZ-AC1

Test Engineer

Edith Yu

Test Date Remark:

2022-07-27

Test Mode:

BLE

1. Average measurement was not performed if peak level lower than average limit.

2. Other frequency was 20dB below limit line within 1-18GHz, there is not show in the

report.

Test Frequency Channel (MHz)
Above 1GHz

Reading Level (dBV)

Factor (dB/m)

Measure Level
(dBV/m)

Limit (dBV/m)

Margin Detector Polarization (dB)

4799.5

49.8

3.1

52.9

74.0

-21.1 Peak Horizontal

4799.5

46.5

3.1

49.6

54.0

-4.4 Average Horizontal

8191.0

37.2

8.6

45.8

74.0

-28.2 Peak Horizontal

00

11982.0

36.8

12.1

48.9

74.0

-25.1 Peak Horizontal

4799.5

45.8

3.1

48.9

7579.0

37.4

7.9

45.3

74.0

-25.1 Peak Vertical

74.0

-28.7 Peak Vertical

10945.0

35.8

12.9

48.7

74.0

-25.3 Peak Vertical

4876.0

49.5

3.2

52.7

74.0

-21.3 Peak Horizontal

4876.0

43.6

3.2

46.8

54.0

-7.2 Average Horizontal

8148.5

37.1

8.8

45.9

74.0

-28.1 Peak Horizontal

19

11463.5

36.7

12.6

49.3

74.0

-24.7 Peak Horizontal

4884.5

45.3

3.3

48.6

8216.5

37.2

8.8

46.0

74.0

-25.4 Peak Vertical

74.0

-28.0 Peak Vertical

10860.0

37.0

12.8

49.8

74.0

-24.2 Peak Vertical

4961.0

47.6

3.5

51.1

74.0

-22.9 Peak Horizontal

4961.0

43.1

3.5

46.6

54.0

-7.4 Average Horizontal

7528.0

36.9

8.0

44.9

74.0

-29.1 Peak Horizontal

39

10970.5

36.7

12.7

49.4

74.0

-24.6 Peak Horizontal

4961.0

46.4

3.5

49.9

7477.0

36.1

8.1

44.2

74.0

-24.1 Peak Vertical

74.0

-29.8 Peak Vertical

11123.5

35.7

12.3

48.0

74.0

Note 1: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m)

-26.0 Peak Vertical

Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m) - Pre_Amplifier Gain (dB)

Note 2: Average measurement was not performed when peak measure level was lower than the average limit.

33 of 46

Report No.: 2207RSU031-U2

The Result of Radiated Emission below 1GHz: Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: VULB 9168_25-2000MHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2402MHz

Test Date: 2022-07-28 Engineer: Edith Yu Polarity: Horizontal Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

Factor

Type

(MHz)

Level

Level

(dB)

(dBV/m) (dB/m)

(dBV/m) (dBV)

1

35.820

24.175

6.375

-15.825

40.000

17.800

PK

2

55.705

24.475

6.634

-15.525

40.000

17.841

PK

3

149.310 29.183

11.215

-14.317

43.500

17.968

PK

4 *

239.520 33.813

17.704

-12.187

46.000

16.109

PK

5

335.550 30.218

10.732

-15.782

46.000

19.486

PK

6

928.705 33.788

3.842

-12.212

46.000

29.947

PK

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Note 4: QP measurement was not performed when peak measure level was lower than the QP limit.

Note 5: The amplitude of radiated emissions (frequency range from 9kHz to 30MHz and 18GHz to 25GHz) is

that proximity to ambient noise, which also are attenuated more than 20 dB below the permissible value.

Therefore, the data is not presented in the report.

34 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: VULB 9168_25-2000MHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2402MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-28 Engineer: Edith Yu Polarity: Vertical Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

Factor

Type

(MHz)

Level

Level

(dB)

(dBV/m) (dB/m)

(dBV/m) (dBV)

1 *

55.705

31.982

14.141

-8.018

40.000

17.841

PK

2

108.085 34.580

20.125

-8.920

43.500

14.455

PK

3

179.865 30.541

13.873

-12.959

43.500

16.668

PK

4

239.035 31.666

15.608

-14.334

46.000

16.058

PK

5

480.080 31.466

8.699

-14.534

46.000

22.767

PK

6

954.410 33.713

4.022

-12.287

46.000

29.691

PK

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Note 4: QP measurement was not performed when peak measure level was lower than the QP limit.

Note 5: The amplitude of radiated emissions (frequency range from 9kHz to 30MHz and 18GHz to 25GHz) is

that proximity to ambient noise, which also are attenuated more than 20 dB below the permissible value.

Therefore, the data is not presented in the report.

35 of 46

Report No.: 2207RSU031-U2

A.7 Radiated Restricted Band Edge Test Result

Site: WZ-AC1

Test Date: 2022-07-27

Limit: FCC_Part15.209_RE(3m)

Engineer: Edith Yu

Probe: BBHA9120D_1167_1-18GHz

Polarity: Horizontal

EUT: Fingertip Pulse oximeter

Power: By Battery

Test Mode: Transmit by BLE at 2402MHz

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1 *

2381.583 55.638

25.117

-18.362

74.000

2

2390.000 53.485

22.959

-20.515

74.000

3

2402.245 75.780

45.221

N/A

N/A

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.520 30.526 30.559

Type
PK PK PK

36 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2402MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Horizontal Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1

2388.708 36.367

5.843

-17.633

54.000

2 *

2390.000 36.413

5.887

-17.587

54.000

3

2402.008 65.988

35.429

N/A

N/A

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.524 30.526 30.559

Type
AV AV AV

37 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2402MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Vertical Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1 *

2375.407 55.362

24.829

-18.638

74.000

2

2390.000 50.962

20.436

-23.038

74.000

3

2402.198 75.767

45.208

N/A

N/A

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.534 30.526 30.559

Type
PK PK PK

38 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2402MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Vertical Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1 *

2386.522 36.464

5.944

-17.536

54.000

2

2390.000 36.284

5.758

-17.716

54.000

3

2402.008 73.445

42.886

N/A

N/A

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.520 30.526 30.559

Type
AV AV AV

39 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2480MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Horizontal Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1

2480.222 78.233

47.531

N/A

N/A

2

2483.500 52.761

22.058

-21.239

74.000

3 *

2485.733 55.652

24.947

-18.348

74.000

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.702 30.704 30.705

Type
PK PK PK

40 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2480MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Horizontal Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1

2480.112 76.752

46.051

N/A

N/A

2

2483.500 36.691

5.988

-17.309

54.000

3 *

2485.942 36.928

6.223

-17.072

54.000

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.702 30.704 30.705

Type
AV AV AV

41 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2480MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Vertical Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1

2480.211 77.939

47.237

N/A

N/A

2

2483.500 52.360

21.657

-21.640

74.000

3 *

2486.085 55.277

24.572

-18.723

74.000

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.702 30.704 30.705

Type
PK PK PK

42 of 46

Site: WZ-AC1 Limit: FCC_Part15.209_RE(3m) Probe: BBHA9120D_1167_1-18GHz EUT: Fingertip Pulse oximeter Test Mode: Transmit by BLE at 2480MHz

Report No.: 2207RSU031-U2
Test Date: 2022-07-27 Engineer: Edith Yu Polarity: Vertical Power: By Battery

No Mark Frequency Measure Reading Margin

Limit

(MHz)

Level

Level

(dB)

(dBV/m)

(dBV/m) (dBV)

1

2480.002 76.797

46.096

N/A

N/A

2

2483.500 36.691

5.988

-17.309

54.000

3 *

2484.930 36.713

6.009

-17.287

54.000

Note 1: " * ", means this data is the worst emission level.

Note 2: Measure Level (dBV/m) = Reading Level (dBV) + Factor (dB/m).

Note 3: Factor (dB/m) = Cable Loss (dB) + Antenna Factor (dB/m).

Factor (dB/m)
30.701 30.704 30.704

Type
AV AV AV

43 of 46

A.8 AC Conducted Emissions Test Result This device is powered by battery, so this requirement is not applicable.

Report No.: 2207RSU031-U2

44 of 46

Appendix B - Test Setup Photograph
Refer to "2207RSU031-UT" file.

Report No.: 2207RSU031-U2

45 of 46

Appendix C - EUT Photograph
Refer to "2207RSU031-UE" file.

Report No.: 2207RSU031-U2

The End 46 of 46



Related FCC IDs: