EMC Test Report

David Schramm

SAR Exemption

Bright Uro Inc GUS1000 Sensor 2BHMUGUS1000 2BHMUGUS1000 gus1000

PDF Viewing Options

Not Your Device? Search For Manuals or Datasheets below:


File Info : application/pdf, 5 Pages, 539.11KB

Document DEVICE REPORTGetApplicationAttachment.html?id=8199446
Test Report Number: 5079617EMC09 Rev.1 Bright Uro / GUS-M-1000/GUS-F-1000 Page: 1 of 5
RF Exposure Report

Project Number: 5079617 Report Number: 5079617EMC09

Offer Number: SUW-202212003799 Revision Level: 1

Client: Bright Uro
Equipment Under Test: Wireless, Catheter-Free, Urodynamic Monitoring System Model Name: Atusa Male Sensor
Model Number: GUS-M-1000 Referenced Model: Atusa Female Sensor: GUS-F-1000
FCC ID: 2BHMUGUS1000 Module Model: nRF52811

Applicable Standards: 47 CFR §§ 2.1093 (Portable) FCC KDB 447498 D01 General RF Exposure Guidance v06

Report issued on: 30 March 2024 Report issued on: 07 April 2025
Result: Exempt from SAR evaluation

FOR THE SCOPE OF ACCREDITATION UNDER CERTIFICATE NUMBER: 3212.01 This report must not be used by the client to claim product certification, approval, or endorsement by A2LA, NIST, or any agency of the Federal
Government.

Prepared by:

Daniel Alvarez, RF/EMC Sr. Staff Engineer

Reviewed by:

Stephen Whalen, SAR/EMC Manager

Remarks: This document is issued by the Company subject to its General Conditions of Service printed overleaf, available on request or accessible at http://www.sgs.com//en/Terms-and-Conditions.aspx. And for electronic format documents, subject to Terms and Conditions for Electronic Documents at http://www.sgs.com/en/Terms-and-Conditions/terms-e-document.aspx.

Attention is drawn to the limitation of liability, indemnification and jurisdiction issues defined therein. Any holder of this document is advised that information contained hereon reflects the Company's findings at the time of its intervention only and within the limits of Client's instructions, if any. The Company's sole responsibility is to its Client and this document does not exonerate parties to a transaction from exercising all their rights and obligations under the transaction documents. This document cannot be reproduced except in full, without prior written approval of the Company. Any unauthorized alteration, forgery or falsification of the content or appearance of this document is unlawful and offenders may be prosecuted to the fullest extent of the law. Unless otherwise stated the results shown in this test report refer only to the sample(s) tested and such sample(s) are retained for a maximum of 30 days only.

Test Report Number: 5079617EMC09 Rev.1 Bright Uro / GUS-M-1000/GUS-F-1000 Page: 2 of 5
TABLE OF CONTENTS
1 GENERAL INFORMATION............................................................................................................................................. 3 1.1 CLIENT INFORMATION ................................................................................................................................................... 3 1.2 TEST LABORATORY ....................................................................................................................................................... 3 1.3 GENERAL INFORMATION OF EUT .................................................................................................................................. 3 1.4 SEPARATION DISTANCE................................................................................................................................................. 3
2 SAR EXCLUSION CALCULATIONS ............................................................................................................................. 4 3 REVISION HISTORY ........................................................................................................................................................ 5

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

1 General Information

Test Report Number: 5079617EMC09 Rev.1 Bright Uro / GUS-M-1000/GUS-F-1000 Page: 3 of 5

1.1 Client Information
Name: Bright Uro, Inc. Address: 3 Goddard City, State, Zip, Country: Irvine, CA 92618 USA

1.2 Test Laboratory
Name: SGS North America, Inc. Address: 620 Old Peachtree Road NW, Suite 100 City, State, Zip, Country: Suwanee, GA 30024, USA

Accrediting Body: Type of lab:
Certificate Number: FCC Designation Number:

A2LA Testing Laboratory 3212.01 US1126

1.3 General Information of EUT

Type of Product: Model Name:
Model Number: Referenced Model:
Serial Number: Module Models:

Wireless, Catheter-Free, Urodynamic Monitoring System Atusa Male Sensor GUS-M-1000 Atusa Female Sensor: GUS-F-1000 (For reference only - Not Tested) DV0013 (Conducted Sample) nRF52811

Frequency Range: Data Modes: Antenna:
Max Conducted Output Power:

2402 ­ 2480 MHz Bluetooth Low Energy (BLE1M ­ GFSK) Internal PCB Trace (0.5 dBi gain) * Bluetooth LE: -0.39dBm

Sample Received Date: 13 November 2023, 15 March 2024 Dates of testing: 14 November 2023 to 15 March 2024

*Data was not measured by SGS laboratory and therefore not responsible for accuracy. Data obtained via customer, specification sheet, previous regulatory filing or other.

1.4 Separation Distance
The closest exposure distance occurs when the sensor is inserted into the body. For this reason, the closest separation distance of 5mm was used to calculate the SAR exemption.

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC09 Rev.1 Bright Uro / GUS-M-1000/GUS-F-1000 Page: 4 of 5
2 SAR Exclusion Calculations
The highest output power in conjunction with the Upper and Lower frequency boundaries have been used to demonstrate compliance for Bluetooth transmission mode. Simultaneous transmission is not possible.

The EUT is considered a body application.

Bluetooth LE (FCC ID: 2BHMUGUS1000)

447498 D01 General RF Exposure Guidance v06
SAR test exclusion calculations Section 4.3: General SAR test exclusion guidance / Section 4.3.1: Standalone SAR test exclusion considerations

Max Power: Duty Cycle: Min separation distance: Frequency, f:

Input
-1.47 100.0%
5 2402

Select Units dBm
mm MHz

Unit correction

0.71285303

1

<== Source based time average duty cycle

5

2402

100

Value reference Number v1 v2 v3

Values used for Calculation

1.000

mW

5

mm

1.550

Reference number definition
[max. power of channel, including tune-up tolerance, mW] 'Rounded to nearest mW [min. test separation distance, mm] 'Rounded to nearest mm [f(GHz)]

a) For 100 MHz to 6 GHz and test separation distances  50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the following: [(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] · [f(GHz)]  3.0 for 1-g SAR, and  7.5 for 10-g extremity SAR,

Exclusion Calculation(1g):

0.3100

number <== [v2 / v3] must be less than 3

Exclusion Calculation(10g):

0.3100

number <== [v2 / v3] must be less than 7.5

Conclusions (Body):

The EUT max power is BELOW the threshold. SAR Testing is NOT required for Body applications

Conclusions (Extremity): The EUT max power is BELOW the threshold. SAR Testing is NOT required for Extremity applications

447498 D01 General RF Exposure Guidance v06
SAR test exclusion calculations Section 4.3: General SAR test exclusion guidance / Section 4.3.1: Standalone SAR test exclusion considerations

Max Power: Duty Cycle: Min separation distance: Frequency, f:

Input
-0.39 100.0%
5 2480

Select Units dBm
mm MHz

Unit correction

0.914113241

1

<== Source based time average duty cycle

5

2480

100

Value reference Number v1 v2 v3

Values used

for Calculation

1.000

mW

5

mm

1.575

Reference number definition
[max. power of channel, including tune-up tolerance, mW] 'Rounded to nearest mW [min. test separation distance, mm] 'Rounded to nearest mm [f(GHz)]

a) For 100 MHz to 6 GHz and test separation distances  50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the following: [(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] · [f(GHz)]  3.0 for 1-g SAR, and  7.5 for 10-g extremity SAR,

Exclusion Calculation(1g):

0.3150

number <== [v2 / v3] must be less than 3

Exclusion Calculation(10g):

0.3150

number <== [v2 / v3] must be less than 7.5

Conclusions (Body):

The EUT max power is BELOW the threshold. SAR Testing is NOT required for Body applications

Conclusions (Extremity): The EUT max power is BELOW the threshold. SAR Testing is NOT required for Extremity applications

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC09 Rev.1 Bright Uro / GUS-M-1000/GUS-F-1000 Page: 5 of 5

3 Revision History

Revision Level
0 1

Description of changes
Initial Release Sections 1.4 and 2 were updated

Revision Date
30 March 2024 07 April 2025

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)



Related FCC IDs:

Search Any Device: