4 Setup

4 Setup

4.1 System Setup

Select

in the upper-left corner of the screen to enter the system configuration screen.

4.1.1 Setting the System Language

Perform the following procedure: 1. Select a language from the "Language" pull-down list. 2. Select [Apply], and restart the system as prompted.

4.1.2 Equipment Configurations

Select [Equipment Configurations].

Item Institution Device information
WorkMode

Description
The hospital name to be displayed and printed.
· Equipment type: Product model supported by the software. · Install mode: Forward installation and reverse installation.
Select "Demo" to enable simulated exposure: the system performs simulated exposure (generates a demo image without outputting the Xray) when the body part graph is double-clicked on the exam management screen.

Operator's Manual

4 - 1

4 Setup

Item Input Method Editor
Workflow Layout Send Image

Description
Set the default input method.
You can also modify the input method of the operating system as required:
· English (US keyboard): set the default input method of DROC to American keyboard. If the input method of the operating system has been changed, DROC will switch the input method of the operating system to American keyboard when startup.
· Keep OS keyboard settings: set the input method of DROC to be consistent with that of the operating system. If the current operating system is a non-American keyboard input method, character input errors may occur, please use this option with caution. If the character input is abnormal, end the DROC from the Task Manager, change the input method of the operating system to American keyboard, and then restart the DROC.
Set the default workflow layout.
When the system is configured with UltraSync, the "Display UltraSync button" is optional.
Set the default image destination.
Check whether to perform data masking when sending images to UltraSync.
NOTE:
When using the sending E-mail function for the first time, you need to set up the email account on the Thunderbird client of the operating system screen.

Configuring Thunderbird
Before configuring Thunderbird, ensure IMAP is enable on your E-mail account. For details, see the instructions of the mailbox. Perform the following procedure: 1. Return to the operating system screen, and run the Thunderbird client. 2. Select the existing mail account button. 3. Enter the account name, email address and password, and then select the continue button.
Thunderbird will automatically configure the mailbox connection. 4. Select the done button to automatically switch to the email login screen. 5. Enter the mailbox account and password, and then select the allow button to complete mailbox
configuration.

4.1.3 User Management
Select [User Management]. On the user management page, you can add and delete users, change the user password, and assign permissions to common users.

Operator's Manual

4 - 2

TIP:
The following operations are limited to administrator users.

4 Setup

Enabling Access Control
The system administrator can preset the access controls, that is, whether an operator has the right to access data in the system. Access control only can be set by the system administrator. If "Enable access control" is selected, you must be authorized before accessing the data. If unselected, you can access all the data without authorization.
Adding a User
Select [Add] in the lower right corner of the screen, and the "Adding New User" window is displayed. You can set the user name and password. · The user name must contain numbers or letters. · The password must contain numbers or letters.
Changing the Password
Select the desired user from the user list and select [Change Password]. You can change the password.
Deleting a User
Select the user to be deleted from the user list and select [Delete] in the lower part of the screen. A confirmation box is displayed. Select [OK] to delete the user, or select [Cancel] to cancel the deletion.
Assigning Permissions
You can select operation permissions for common users from the permission assignment list. After configuration, select [Save] to apply the settings.
4.1.4 Option Configuration and Authorization
Select [Option&Authorization]. This page displays the license status of the equipment and the license validity period of system functions or optional functions.
4.1.5 Detector Calibration
Select [Detector Cal.] to enter the FPD calibration screen. The screen displays calibration parameters in real time during FPD calibration, such as the current X-ray dosage, remaining time and calibration progress.

Operator's Manual

4 - 3

4 Setup
1 2

1.

Last calibration time

2.

Exposure prompt area

NOTE:
· Users need to calibrate the detector once a year to ensure the stability of the image quality received by the detector, so as to achieve the purpose of optimizing the image quality.
· Recalibrate the FPD if it has not been in use for more than 3 months. · Ensure that the FPD has been powered on for more than 4 hours before FPD calibration. · Only the administrator and authorized users have the right to calibrate the FPD. · During the calibration, ensure that X-ray fully covers the FPD. · Ensure that no foreign substances exist between point sources during calibration.
Perform the following procedure:
1. Manually maximize the light field of the collimator (completely covering the detector). Remove the obstacles between the light field and front baffle plate of the detector assembly.
2. After positioning is completed, select [Start Calibration] to calibrate the FPD and a prompt box will pop up.
3. Select [OK] to calibrate the FPD.
4. When the exposure prompt area is in red and the prompt bar prompts for exposure, press the handswitch for exposure.
5. Repeat the previous step until calibration is complete.
A pop-up window appears, and "Successful" is displayed in the Details area. Calibration is completed. 6. Select [OK] and then select [Exit Calibration].

Operator's Manual

4 - 4

4.1.6 Exam Information Display Configurations
Select [ExamInfoDisplay Configurations].

4 Setup

Item

Description

Image info display settings

Set whether to print animal information entries and the print positions on films.

GUI info display settings The list on the left displays the animal information that can be selected. The list on the right displays the content that will be displayed in the animal information list of the registration management screen.

4.1.7 Register Management Settings

Item DefaultSpecies Animal Breed Set Default Body Size Set Emergency body view set Default body view set Animal ID
Accession No.
Age
Operator

Description
Set the default animal type.
Set the animal species.
Set the weight range of the animal.
Set the animal emergency body position.
Set the default exam position of the animal.
Set whether the system automatically generates the default ID (Animal ID) during local login. Two generation modes are supported:
· Manual: The system does not adopt the automatic generation rule. The ID is manually entered.
· Auto: The system automatically generates the ID. You can set the prefix of this ID (such as "id").
Set whether the system automatically generates the default visit number during local login. Two generation modes are supported:
· Manual: The system does not adopt the automatic generation rule. The accession No is manually entered.
· Auto: The system automatically generates the visit number. You can set the prefix of this visit number (such as "ac").
Used to configure the method of worklist animal age calculation.
· By Actual Date and Time: The age is calculated by subtracting the actual dates. If the subtraction result is less than one year, the actual month is used as the age. If the subtraction result is less than one month, the actual number of days is used as the age. If the subtraction result is more than one year, the actual year is used as the age.
· By Difference of Years: The age is calculated by subtracting years.
Sets operation doctor record rules.

Operator's Manual

4 - 5

4.1.8 Exam Configurations

4 Setup

Item

Description

Exposure Control Set

BodyThickness Range

Set the default body thickness range and body thickness step.

AcqImageTime

Set the maximum exposure time of the FPD.

Auto play iShotHelper Set whether to auto play post assist video.

When setting up auto play, you can set up the play interval.

AfterExp ns auto switch to Set the time interval for automatically switching to

next bodyview

the next position after exposure.

Record exposure parameters

Set whether to save the exposure parameter into the DCM file.

ImageProcessSet SetBodyIcon Font Size

Set the font size of body part tag: A greater the value means a larger font size of the added body part tag during image browsing.

Set Comment Font Size

Set the font size of comment: A greater the value means a larger font size of the note on the image browsing screen.

Diag Ref image library manager

Set the diagnostic reference graph library.

Gray Window Factor

· Window Width Adjust Factor: A larger value means a higher speed for window width adjustment and a greater value change.
· Window Center Adjust Factor: A larger value means a higher speed for window level adjustment and a greater value change.

Modality Settings

Set the device type that is written when the DICOM file is saved.

Preview Image After Exposure

If this option is selected, the unprocessed preview image is displayed after exposure.

Display animal

Set whether the animal information is displayed by

information on the image default.

Enable Automatic bodyview Labeling

Set whether to automatically add body part tags based on body parts.

Save Pure Image After Exposure

Set whether to save the Pure image when the animal is saved.

Save Raw Image After Exposure

Set whether to save the Raw image when the animal is saved.

Operator's Manual

4 - 6

4 Setup

Item

Description

ImageProcessSet Image Auto Rotation/ Flipping

Used to automatically rotate/flip the image according to the selected body position.

Enable the image rotation Used to calibrate the image direction after the FPD is

calibration mode

installed.

For detailed information, contact the Customer Service Department or your local distributor.

PlugInCable

Auto Send Failed Task

If this option is selected, the system will automatically send failed tasks after the data cable is connected.

Auto Send UnSent

If this option is selected, the system will automatically send images not sent within the specified time period after the data cable is connected.

Auto Send Period: The options are today, past three days, and past one week.

CloseExamSet

Auto Send To PACS While When this function is enabled, the new acquired

Closing Exam

images will be automatically sent to the PACS

workstation when the "Exam" screen is closed.

Auto Send To Mico+ While Closing Exam

When this function is enabled, the new acquired images will be automatically sent to the UltraSync when the "Exam" screen is closed.

4.1.9 DICOM Image Printing Configuration

Item

Description

Display animal

Set whether to display animal information on films.

information on the image

Display the scale

Set whether to display a proportional scale on films.

Print Font Size

Set the font size of animal information printed on film: the greater the value, the larger the font.

Close Thumbnail If Print After print is successful, the system automatically clears the thumbnails. Success

Print Layout Config

Enable or disable the function of typesetting by body position.

· Delete: Delete the configured body position selected. The default option cannot be deleted.
· Modify: Modify the film size, print layout and film direction of the configured body position.
· Add: Configure the film size, print layout and film direction of the new body position.

Operator's Manual

4 - 7

4 Setup

Item When Plug In Cable

Description
Automatic print image configuration when the network cable is plugged in:
· Auto Print Failed Task: When this option is selected and the network cable is plugged in, the system automatically reprints the failed tasks.
· Auto Print Unprinted Images: When this option is selected and the network cable is plugged in, the system automatically prints unprinted images within the set period. Auto Print Period: The options are today, past three days, and past one week.

4.1.10 Help Manual
View the detailed operating instructions.
4.1.11 Log and Configuration Maintenance

Item Log&Maintain Log Export
User Config Management

Description

Delete Log?

Set whether to delete logs automatically.

Log Keeping Days Set the number of days during which logs are retained, if automatic log deletion is configured.

Select Disk to Export

Select destination for export, and select [Export]:
Select the time period for export, and select [Start] and wait. When the export is completed, the system prompts the corresponding information under "select the disk to export".

Select [Remove Device], and select the USB flash disk to be removed on the removal screen for safe removal.

Add Backup

Select to enter the backup dialog box, and enter the name of the system configuration to be backed up.

After you select [OK], the added backup is displayed in the backup list.

Delete Backup

Select to delete the backup from the backup list.

Restore Config

Select to restore system configuration from backup.

Restore Default Config

Select to restore the default system configuration, which is not displayed in the backup list and cannot be deleted.

4.1.12 Statistic
You can use the statistics function to view the workload under each account, and count the total number of persons and exposures based on time segments, examined parts, and other criteria.
You can export the statistical results as a .CSV file through [Export Statistics].

Operator's Manual

4 - 8

4 Setup
Use [Export Detailed Exposure Record] to export the records (including the operator, exam information, exposure parameters, and the deleted image exposure data) to a CSV file by the operator.
4.1.13 Parameter Manager
On the parameter manager screen, you can set view parameters, exposure parameters, and postprocessing parameters, and restore parameters.
Select [Logout] on the menu bar to exit from the "Parameter Manager" screen and enter the Windows operating screen. Run "DrocApp.exe" shortcut icon on the desktop to start the DROC.
View Definition
Expand all/ Hide all: To display all exam bodyviews, select [Expand all]; to hide all bodyviews, select [Hide all].
Configuring the Post-processing Parameters
On this page, you can configure the image post-processing parameter, such as set the default style, save as a new post-processing parameter, set the default parameter, and modify the post-processing parameter value.
· On the Post-processing Parameter panel, select a Root node, panel, animal type, bodypart, bodyview, and then right-click, an operation menu is displayed. To set the selected style as a default style, select [Set as default[. Then, the default styles of all views under the Root node, panel, animal type, bodypart, and bodyview are changed
· After you select a post-processing parameter, right-click the parameter, and an operation menu is displayed. ­ Save as a new post-processing parameter: To save the post-processing parameter as a new parameter, select [Save as...]. ­ Set the default post-processing parameter: To set the selected post-processing parameter as the default parameter, select [Set as default]. ­ Edit the post-processing parameter: After you select a post-processing parameter, details of the parameter are displayed on the right side. After modification, select [Save the Change].
Configuring the Exposure Parameters
On this page, you can configure the exposure parameter.
Edit an exposure parameter: After you select an exposure parameter, details of the parameter are displayed on the right side. After modification, select [Save the change].
Configuring the View Parameters
On this page, you can configure the view parameter, such as modify the view parameter value, set the default view parameter.
Select a view and right-click on it. An operation menu is displayed.
· Set the default parameter: To set a selected view parameter as a default parameter, select [Set as default] from the menu.

Operator's Manual

4 - 9

4 Setup
· Modify the view parameter value: After you select a view parameter, details of the parameter are displayed on the right side. Modify the values of the parameter and then select [Save the change] to save the modified value.
Restore Factory Parameters
Select [Restore factory Param] on the menu bar to select the parameters to be restored as required.
4.1.14 Prepay Installment
Display the prepay installment information. For detailed information, contact the Customer Service Department or your local distributor.
4.1.15 Report Configurations
Configure logo icon of the hospital, edit report template, select report printer, edit knowledge base and configure the report settings.

Item Hospital Logo
Template
Printer Knowledge base

Description Provide the function of managing LOGO images of the hospital.
· Clear: Clear the existing hospital logo image. · Change: Change the image information of the hospital logo. The default report template can be selected. · Select [Edit] to edit the report template. · Select [Restore] to restore the templates to the original factory
default. Set the default printer to print report. Provide the functions of knowledge base selection, import, export, and loading factory. · Select [Import] to import the local knowledge base. · Select [Export] to export the knowledge base to the local device. · Select [Restore] to restore the knowledge base to the factory default.

Adding a Report Template
Perform the following procedure: 1. Select [Edit] to edit the report template. 2. Select [New] to generate a blank template.

Operator's Manual

4 - 10

4 Setup

Item Add text Add the text with the print label Add text box
Alignment Control

Description
Follow the steps below:
1. Select from the lower toolbar, click and drag the cursor to add the size of text box.
2. Double-click the text box to input text and click outside to end the editing. You can view and edit the control properties in the left column.
Follow the steps below:
1. Right-click the text and select print label. 2. Select the text icon from the pop-up dialog box.
Print label of the control is displayed in the left column. When you edit the text with print label in the report editor, the print label will be recognized and the corresponding animal data will be loaded into the text, but the animal data cannot be edited.
The manual input of the report can be added to the edit box.
Follow the steps below:
1. Select from the lower toolbar, press and hold the left mouse button to add and adjust the size and position of text box.
2. Double-click the box to input text and click outside to end the editing. You can view and edit the control properties in the left column.
3. Right click and select the print label. 4. In the displayed print label dialog box, select the print label of the edit
box. Print label of the control is displayed in the left column.
NOTE: The print label must be added to the edit box, otherwise the edited content in the diagnosis report will not be saved.
Follow the steps below:
1. Hold down the <Ctrl> button, and then click to select the controls to be aligned.
2. Select the alignment button at the bottom of the window. The specified alignment operation is performed.

Editing System Template
Perform the following procedure: 1. Select [Edit] to edit the report template. 2. Select [Open] and browse to select the template you want to edit.
Modify the template as needed. For details, see Adding a Report Template above.

Operator's Manual

4 - 11

4 Setup
4.1.16 Lock Equipment
You can use the device locking function to lock the system. To unlock the screen, log in again.
4.2 System Date and Time Setup
NOTE:
Ensure that the current exam date and time are the same as the system date and time.
Click the Time/Date of the system status icons area to set the date format. Switch to Windows interface to set the system date and time if necessary.
4.3 Network Setup
Select to enter the Network Setup screen. The network setup screen displays the network connection of the equipment and the IP address of the network adapter. · Network card name: Displays all network card names in the system. · Connection status: Displays the connection status of the currently selected network card. · IP address: You can configure the system to automatically obtain an IP address, or set an IP
address manually.
4.4 Disk Space Configuration
Select (Disk space setup) to enter disk space setup screen.

Item Image Delete Method

Description
· Don't auto delete image: Delete the image manually as needed. · Auto delete image by time: When the maximum storage time is
reached, the system will automatically delete the images that have been saved for a long time. · Auto delete image by capacity: When the remaining storage space reaches the set value, the system will automatically delete the images according to the storage order.
NOTE:
The deleted images will be saved to the recycle bin. You can restore the deleted image from the recycle bin before the recycle bin is cleared.

Operator's Manual

4 - 12

Item Recycle

4 Setup
Description · Do not automatically clean up the recycle bin: Clean the recycle bin
manually as required. · Automatically clean up the recycle Bin: When the maximum storage
days are reached, the device automatically deletes the animal data of the recycle bin that exceeds the storage days.
NOTE: The cleaned data in the recycle bin cannot be recovered. Back up animal data before enabling the auto clean recycle bin function.

4.5 DICOM Setup

Select (DICOM Setup) to enter the DICOM setup screen.

4.5.1

DICOM Local Preset
Select "Local" tab, you can configure the local DICOM entity name and port number.

4.5.2 DICOM Worklist Preset
On this page, you can set DICOM Worklist parameters and test connection.

Worklist Configuration

Item Enable Worklist AE Title Port
IP Address
Verification
Worklist Search Conditions

Description If it is selected, the Worklist function will be enabled. Application Entity title. It is consistent with that of the Worklist server.
DICOM communication port. The port should be consistent with that of the Worklist server port. IP address of the Worklist node. The value consists of four segments of integers ranging from 0 to 255. Verify whether each worklist node is connected.

Item Time Range Refresh interval(s) Identity Modality

Description Set the date range. Set the worklist communication interval. Set the identity of a task list. Set the device type for obtaining information.

Operator's Manual

4 - 13

Item Age
Body View Map

4 Setup
Description Set the calculation method of age in the animal information obtained from worklist.

Item Enable bodyviewMap Set the mapping label Worklist Code Adding a body part mapping
Deleting a body part mapping

Description
If it is selected, body part mapping is enabled.
Set the DICOM field for the worklist server to store body part information.
Enter the information (such as the standard thorax position) sent by the worklist.
Follow the steps below:
1. In the worklist check item, enter the information sent by the worklist. 2. In the body part list, select the body part corresponding to the DROC
application. 3. Select [Add] > [Save] to add the body part mapping to the mapping
list.
Select a body part in the body part list, and select [Delete] > [Save] to delete.

4.5.3 DICOM Print Preset
On this page, you can set DICOM Print parameters and test connection.

Operator's Manual

4 - 14

Item Print Server List

Enable Print Add

Delete Set as default

Cancel Default

Node Attribute

Verification AE Title Port

IP address

Brightness

4 Setup
Description If it is selected, the print service is enabled. You can add and configure multiple print nodes. Follow the steps below: 1. Select [Add], and enter the following information
about the added node in sequence: AE Title, port number, IP address and print brightness, 2. Select [Save]. The added node is displayed in the print node list. The attribute of the added node is displayed in the node attribute box. You can delete the currently selected print node. You can configure a default print node based on a default node. You can cancel the currently selected default print node. Verify the connectivity of each DICOM print node. Name of a DICOM node entity Port number of the service provided by the DICOM node. IP address of the DICOM node. The value consists of four segments of integers ranging from 0 to 255. You can select the print effect after brightness compensation adjustment.

Operator's Manual

4 - 15

4 Setup

Item
Print Configuration

Min density Max density Border density Empty Density Destination Medium Type Inter Type

Print Border Polarity Priority Config info Owner Tag

Film Tag Set Film Size

Description Enter the minimum density of the film. Enter the maximum density of the film. Set the density between film images. Set the density of the surrounding area.
Specify where the file is exposed.
Specify print medium. Select how the printer magnifies an image to fit the film. Set whether to print image border. Set the default image polarity. Specify print job priority. Enter configuration information in the field. Displayed on the control software screen of the operator console. The value can be modified based on the requirements of the hospital You can change the value based on the requirements of the hospital. You can enter letters and numbers. sets the default film size. Select [Set Film Size] to select a film size.

4.5.4 DICOM Storage Preset
On this page, you can set DICOM Storage parameters and test connection.

Operator's Manual

4 - 16

4 Setup

Item
Storage Server List

Enable Storage Add

Delete

Set Default

Cancel Default

Node Attribute

Verification AE Title Port

IP Address

Enable Lossless Compression

Description If it is selected, the storage service is enabled. You can add and configure multiple storage nodes.
Follow the steps below:
1. Select [Add], and enter the following information about the added node in sequence: AE Title, port number, and IP address.
2. Select [Save].
The added node is displayed in the storage node list. The attribute of the added node is displayed in the node attribute box. Select to delete the selected node in the storage node list. You can configure a default storage node based on a default node. You can cancel the currently selected default storage node. Verify the connectivity of each DICOM storage node. Name of a DICOM node entity. Port number of the service provided by the DICOM node. IP address of the DICOM node. The value consists of four segments of integers ranging from 0 to 255. If it is selected, the lossless compression is enabled.

4.5.5 MPPS Preset
On this page, you can set DICOM MPPS parameters and test connection.

Item Enable MPPS AE Title IP Address
Port Verification

Description If it is selected, the MPPS service is enabled. Name of a DICOM node entity. IP address of the DICOM node. The value consists of four segments of integers ranging from 0 to 255. Port number of the service provided by the DICOM node. Verify the connectivity of the node.

Operator's Manual

4 - 17

4.5.6 DICOM SC Preset
On this page, you can set DICOM SC parameters and test connection.

4 Setup

Item

Description

Enable Storage Commit If it is selected, the Storage Commit service is enabled.

AE Title

Name of a DICOM node entity.

IP Address

IP address of the DICOM node. The value consists of four segments of integers ranging from 0 to 255.

Port

Port number of the service provided by the DICOM node.

Verification

Verify the connectivity of the node.

4.5.7 DICOM Q/R Preset
On this page, you can set DICOM Q/R parameters and test connection.

Item Enable Q/R AE Title IP Address
Modality Port Verification

Description If it is selected, the Query/Retrieve service is enabled. Name of a DICOM node entity. IP address of the DICOM node. The value consists of four segments of integers ranging from 0 to 255. Set the device type, DX and DR are optional. Port number of the service provided by the DICOM node. Verify the connectivity of the node.

4.6 Screen Brightness Management
Select to enter the display brightness adjustment screen. Drag the slider to adjust the screen brightness. · [DefaultBright]: to restore the default brightness. · [SetDefaultBright]: to set the current brightness as the default.

4.7 System Information

Select

button on the upper right corner of the control software.

The version windows is displayed, and the information includes the DROC, FPD, the high voltage generator and the collimator, and so on.

Operator's Manual

4 - 18

5 Starting an Exam

5 Starting an Exam

5.1 Registration Management
You can start an animal exam in the following situations: · Local registration
­ New animal information: to start a new animal exam, animal information must first be entered.
­ Return visit registration: to start a new exam for animal who is already registered, the recorded information can be obtained through history exams.
­ Emergency: Animal ID information is generated quickly. · Worklist registration: register animals in the worklist.
5.1.1 Local registration
After startup, the system automatically enters the registration management screen. When the system is in another interface, select [Registration Management] to return to the registration management screen.
Manual Registration
Perform the following procedure:
1. Enter the animal information in the animal information registration area.
2. Select the body parts of the current animal to be radiographed from the examined body part list.
a. Select an exam part: Select the exam part area on the animal diagram, and load the body view library under the corresponding exam part.
b. Select body view: Select to select a body view in the body view library, and then select [ ] to add it to the body view list on the right side. Select [ ] to delete the selected body view from the selected body view list. In the selected body view list, use [ ]/[ ] to adjust the position order.
3. Select [Save] to register a new animal.
The completed registration animal information is displayed in the animal information list.
Emergency Registration
When a large number of animals need to be examined or treated, which requires quick input of animal information, you can use the emergency call registration function to register an animal. Select the

Operator's Manual

5 - 1

5 Starting an Exam
[DogEmr] or [CatEmr] button to enter exam, the system automatically generates an animal ID and directly enters the "Exam" screen.
NOTE:
After emergency call examination is completed, the animal information can be modified on the "History" screen.
Return Visit Registration
On the history screen, double-click an animal information in the animal list to enter the exam screen. select [Additional exam] or [New exam] to perform the return visit registration. · Additional exam: Continue to use the previous animal registration information. · New exam: Register a new animal based on the existing animal information.
Worklist Registration
Select to download Worklist animal information. The system automatically refreshes to obtain the worklist animal information.
TIP:
For details about automatic refresh configuration, see DICOM Worklist Preset section in Setup chapter.
5.1.2 Registration Information Management
You can manage registered animal information, including viewing, modifying, and deleting animal information.
Viewing Animal Information
In the animal information list, select the registered animal for examination, and view the selected animal information in the animal information registration area.
Modifying Animal Information TIP:
The animal information after radiography can be modified on the history screen.
You can modify the information about animals before radiography (examination). Perform the following procedure: 1. In the animal information list, select the animal record to be modified and the animal information
displays in the animal information registration area. 2. Modify the animal information as required in the animal information registration area.

Operator's Manual

5 - 2

5 Starting an Exam
3. Select [Save] to save the changes.
Deleting Animal Information
In the animal information list, select a record and select [Delete] to delete the selected animal record.
5.1.3 Animal Information Query
You can query animal records based on filter criteria. The query process automatically starts after search criteria are entered. · When the criteria is set to "All", the search bar is displayed on the right. Fuzzy search is
supported. Input query conditions in the text box. The content can contain "*" or be blank for fuzzy query. · When the condition is set to "Register Date", the time range is displayed, and the date range is selected by the pop-up box.
5.2 Exposure
Perform the following procedure: 1. Register the animal information, and then select [Start Exam] to enter the exam screen. 2. Selecting a radiographed body view.
The radiographed body views selected during registration are displayed in the thumbnail area. Select the radiographed body view to be exposed in the thumbnail area, and the corresponding radiographed body view parameters are loaded to the exposure parameter area. NOTE:
Select [ViewEdit] to reset the exposure body views. Body views that have been filmed cannot be deleted. 3. Adjust the range of the light field according to the selected position. 4. (Optional) Use the parameter adjustment controls to adjust the exposure parameters if necessary. 5. Perform the exposure. After image acquisition ends during the exposure process, the acquired image thumbnails are displayed in the thumbnail area and the exam will be displayed on the "History" screen. NOTE: If needed, repeat the steps 3 to 5 above to perform exposure again for the same body view.

Operator's Manual

5 - 3

6 Image Browsing and Processing

6 Image Browsing and Processing

On the "Exam" screen, you can browse the images acquired through exposure, process and diagnose the images, and determine whether to accept the images.
6.1 Image Display
The image display area displays the currently acquired images or history images. A proportional scale is displayed on each image to reflect the size proportion between the display image and the actual image. The scale marks are adjusted in real time as the image is zoomed in and out, to reflect proportional changes. The animal information area displays the current animal information. The image displays the exposure parameters and radiation dose.
6.2 Thumbnail
The image thumbnail of the selected exam appears in the thumbnail area on the screen · On the "Exam" screen: the current acquired images are displayed in the thumbnail area. · On the "History" screen: the images of the selected exam are displayed in the thumbnail area.
6.3 Image Operation
Common processing tools are provided for performing image display operations. The following figure shows the corresponding icons.

Icon Name Image roaming Cropping Magnifier

Description Pan the image. Crops images. Magnifies a portion of an image.

Operator's Manual

6 - 1

6 Image Browsing and Processing

Icon Name Local window

Description Enables automatic window width and position in local areas.

90º left rotation

Rotates left by 90º.

90º right rotation

Rotates right by 90º.

Up and down mirroring Reverses image display up and down.

Left and right mirroring Reverses image display left and right.

Rotation

Rotate the image at any angle.

Positive/Negative image Flips an image between positive and negative display.

Animal Information Display

Displays animal details on an image, including the basic animal information and exposure parameters.

Multi-cell Display Setup Set the image layout.

/

Right mouse button

Adjusts the window width and position, and position identifier.

/

Save

Save the adjustment result to the image.

/

Re-Proc

Load the original image and reprocess the image with the default post-processing parameters of the current body position.

Image Adjustment
Slide the screen in the image area to adjust the window width and position. Press and hold the right mouse key to adjust:
· Drag upward to reduce the window position value. · Drag downward to increase the window position value. · Drag to the left to reduce the window width. · Drag to the right to increase the window width.
The current window width and position of an image are displayed in the lower-right corner of the image window.
When the local window exists, the internal window adjustment is applied to the local window image, and the external window adjustment is applied to the entire image. When the local window does not exist, the window adjustment is applied to the entire image in the image window.
Cropping
Perform the following procedure:
1. Use the default crop box in the image or select to set the crop size.
2. Adjust the crop box:

Operator's Manual

6 - 2

6 Image Browsing and Processing

­ Adjust the size: select and drag the 8 small boxes of the crop box area to adjust the crop box size.
­ Adjust the position: move the mouse pointer to any of the four sides of the crop box, and after the mouse pointer turns into a hand shape, drag the crop box to move.
NOTE:
The maximum size of the crop box cannot exceed the image area.
3. Select [Save] to accept image cutting result.
Select [Re-Proc] to restore the image to its original status.

Magnifier
Select to enter the magnifier state. Select and hold the image area to display the magnifier, which is used to zoom in on a local area of the image. Release to hide the magnifier.

Rotate

Perform the following procedure:
1. Select .
2. Select and hold the image area to rotate clockwise or counterclockwise to rotate the image at any angle.

Local Window
Select to display the local window. The window width and position for image display in the local window are optimized for local images and different from those of the entire window.

Positive/Negative Image Display
When browsing images, select to display an image in positive or negative mode. To adapt to film interpretation habits, the system displays exposed images in negative mode by default.

Animal Information Display
Select to display or hide the animal exam information.

Rotation and Mirroring
You can rotate and mirror images during image browsing by using / / / , including 90° left/right rotation and up/down mirroring. During rotation and mirroring, the marks and tags of the image change their positions accordingly, but they do not flip over or rotate.

Image Scaling
Scroll the mouse wheel in the image display to zoom in or out on the image.

Image Panning
In the image display area, click and drag to pan the image.

Operator's Manual

6 - 3

6 Image Browsing and Processing

Paging
When images are displayed on multiple pages, select or to turn to the previous or next page to view the images on each page.

The page number is displayed between the paging buttons. The first digit indicates the current page number, and the second digit indicates the total pages.

Multi-cell Display Setup
Select the multi-cell display button

, and the supported display formats are displayed for selection.

6.4 Annotation
WARNING
You must ensure that the entered comments are correct. Incorrect comments may lead to misdiagnosis.

Comments can be added to an image to bring attention, annotate or communicate information observed during the examination.
The following table lists the comment buttons and functions.

Icon Name

Description

Left Position / Identification

Right Position / Identification

More Position Select "DV", "VD" and input the custom position identification. Identification

Text Annotation

Enter text annotation status.

Clear

Clear the selected text annotation.

Clear All

Clear all text annotations.

Arrow

Enter arrow annotation status.

6.4.1 Position Identification
The position identifiers in the right part of the screen include L (left), R (right), DV (Dorsoventral), VD (Ventrodorsal), and custom tags.
Perform the following procedure:

Operator's Manual

6 - 4

6 Image Browsing and Processing
· To add a position identifier: select a position identifier button to add the position identifier to the image. Select to add "DV", "VD" or custom tags.
· To move the position identifier: select the position identifier on an image and drag and drop the identifier to change its position on the image. When an image is saved, the added position identifier is rasterized to the image.
· To cancel the position identifier: a position identifier button in selected state is in highlighted color. When the button is selected again, the position identifier is not added. Select [Re-Proc] to remove all position identifiers on the image.
· Select [Save] to save the position identifier to the image.
6.4.2 Text Annotation
Adding Text Annotations
Perform the following procedure: 1. Select to enter the text annotation status.
You can also press <Ctrl> and <M> buttons on the keyboard to enter the text annotation status. Character cursor appears on the image. 2. Move the cursor to the position to be added, and click to confirm the target position. 3. Use the keyboard to enter a character comment. Press <Enter> to move the cursor to the new line. 4. Select again to exit the text annotation status. Or, press Ctrl+M button again on the keyboard to exit the text annotation status.
Editing Text Annotations
Perform the following procedure: 1. Select to enter the text annotation status. 2. Move the cursor to the text annotation to be edited, and click to select. 3. Use the keyboard to edit the text annotation. 4. Select again to exit the text annotation status.
Deleting Text Annotations
Perform the following procedure: 1. Select to enter the text annotation status. 2. Deleting annotations:
­ To delete one text annotation: move the cursor to the text annotation and double-click, and there is a frame around the selected text annotation. Select or press <Delete> button on the keyboard.

Operator's Manual

6 - 5

6 Image Browsing and Processing
­ To delete all text annotations: select . 3. Select again to exit the text annotation status.
6.4.3 Arrow Annotation
Adding Arrow Annotations
Perform the following procedure: 1. Select to enter the arrow annotation status. 2. Select and drag on the image to move the arrow to the target position.
­ To adjust the shape and angle: select and drag to adjust to the desired orientation. ­ To adjust the position: select and drag to move the arrow to the target position. 3. Select again to exit the arrow annotation status.
Editing Arrow Annotations
Perform the following procedure: 1. Select to enter the arrow annotation status. 2. Select [...] in the measurement area, shape and angle adjustment marks appear on both ends of
the arrow and position adjustment mark appear in the middle. ­ To adjust the shape and angle: select and drag to adjust to the desired orientation. ­ To adjust the position: select and drag to move the arrow to the target position. 3. Select again to exit the arrow annotation status.
Deleting Arrow Annotations
Perform the following procedure: 1. Select to enter the arrow annotation status. 2. Delete arrow annotations:
­ To delete one arrow: select [Clear Last] to delete the recently-added arrow. Repeat this operation to delete the added arrows chronologically.
­ To delete all arrows: select [Clear All]. 3. Select again to exit the arrow annotation status.
6.5 Measurements
WARNING
Be sure to measure areas of interest from the most optimal image plane to avoid misdiagnosis from inaccurate measurement values.

Operator's Manual

6 - 6

6 Image Browsing and Processing
CAUTION
· Select the proper animal image and measurement tools. Only qualified professionals can decide the appropriate measurements and results.
· Confine measurement calipers to the actual Region of Interest (ROI). Measurements that extend beyond the ROI will be incorrect.
· Using the [Clear All] will clear the measurement caliper and all data in the result window.
NOTE:
If the image has been modified, save the modified image first and then perform the measurement.
There are two kinds of measurement tools: Routine and Application.
6.5.1 Basic Operations and Buttons
The following descriptions for buttons and buttons are used during performing measurements.

Buttons Left mouse button
Mouse

Basic Operations Click the item in the measurement menu. During the measurement, press the left mouse button to confirm the current operation and proceed to the next operation.
NOTE: Unless otherwise specified, "Click" refers to "press the left mouse button."
Move the cursor.

6.5.2 Measurement Menu

TIP:
A measurement tool can be activated by selecting the item. It is described as "Select... in the measurement menu" in the following procedures.
Shortcut buttons of measurement items are as shown in Figure 6-1.
Figure 6-1 Shortcut Buttons of Measurement Items 3 2

1

Operator's Manual

6 - 7

6 Image Browsing and Processing

1.

Shortcut buttons of measurement items.

·

: Distance

·

: Angle

·

: Geom verify

· TPLO: Tibia Plateau Leveling Osteotomy

· i-VHS: Vertebral Heart Score (with measurement results)

· i-RLAD: Radiographic Left Atrial Dimension (with measurement results)

· i-VLAS: Vertebral Left Atrial Size (with measurement results)

· HD: Hip Horizontal Angle

Other application measurement items:

· VHS: Vertebral Heart Score · RLAD: Radiographic Left Atrial Dimension · VLAS: Vertebral Left Atrial Size · M-VLAS: Modified-Vertebral Left Atrial Size (with measurement results)

2.

Measurement Menu

3.

Display/Hide measurement results

Select [...] button to enter the measurement menu as shown in Figure 6-2. Figure 6-2 Measurement Menu
1 2

3

4

1.

Select to exit the measurement status.

Operator's Manual

6 - 8

6 Image Browsing and Processing

2.

Select to switch between the routine measurement menu and application measurement

menu.

3.

Select the measurement item to start the measurement.

4.

· Select [Modify] to enter the measurement editing status.

· Select [Clear Last] to delete the latest measurement.

· Select [Clear All] to clear all measurements.

· Select [Solidify measurements] to add the calipers to the image.

6.5.3 Measurement Caliper
A measurement caliper is a graphic consisting of several points and a straight line or curve drawn on the image.
Fixed/Active End
The ends of calipers can be active or fixed. The active end is called a Cursor.
Caliper End Number
Different caliper end numbers are displayed in the caliper end and results window to distinguish different measurements.
6.5.4 Results Window
The measurement window displays the conducted measurement's result and the engaging measurement in real time.
Results Display
The latest results display in the results window in time sequence. · When viewing the results: If the results window is full, the oldest value will be replaced according
to the "first in, first out" rule. A maximum of 8 results can display in the results window. · To identify the measurement results, numbers are used in the numerical results window.
Moving the Results Window
To move the results window: 1. Select and hold the results window to drag it to the desired position. 2. Release the results window to fix it.
6.5.5 Routine Measurements
Distance
Measures the distance between two points on the image. Perform the following procedure: 1. Select [Distance] in the measurement menu.

Operator's Manual

6 - 9

6 Image Browsing and Processing

2. Determine the starting point of the position to be measured on the image, and then click.

3. Move the cursor to the end point and click to set the end point and the result displays in the results window.

­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

Ellipse
Measures the area and circumference of an ellipse region on the image

Perform the following procedure:

1. Select [Ellipse] in the measurement menu.

2. Determine the starting point of the first axis of the ellipse on the image, and then click.

3. Move the cursor to the desired point and click to set the end point of the first axis of the ellipse.

The second axis appears on the screen.
4. Moving the cursor will increase or decrease the ellipse from the fixed axis. Move the cursor to trace the area of interest as closely as possible, and click on the image to anchor the ellipse region.

­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

Trace

Measures the area and circumference of a closed region on the image.

Perform the following procedure:

1. Select [Trace] in the measurement menu.

2. Determine the starting point on the image, and then click.

3. Move the cursor along the target to trace the outline of the target.

4. Click and the trace line will be closed with a straight line connecting the start and end points.

The trace will also be closed when the cursor is very near to the starting point. ­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

Angle

Measures the angle of two crossing planes on the image and the range is: 0° to 180°.

Perform the following procedure:

1. Select [Angle] in the measurement menu.

2. Set two line segments as described in "Distance" section above.

The angle appears in the results window after setting the line segments.

Operator's Manual

6 - 10

6 Image Browsing and Processing

­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

Rectangular
Measures the pixel average and standard deviation on the rectangular. Perform the following procedure: 1. Select [Rectangle] in the measurement menu. 2. Determine the starting point on the image, and then click. 3. Move the cursor to the target point and click.
The pixel average and standard deviation appear in the results window. ­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

Circle

Measures the area and circumference of a circle region on the image. Perform the following procedure: 1. Select [Circle-2PT] in the measurement menu. 2. Click on the image, the cursor appears on the screen. 3. Move the cursor to the target area and click to confirm the circle position. 4. Click on the image to anchor the circle region.
­ Select and drag on the measurement caliper to adjust the circle size. ­ Select and drag on the measurement caliper to adjust its position.

Length Ratio
Measure the length ratio of two lines. Perform the following procedure: 1. Select [Length Ratio] in the measurement menu. 2. Set two line segments as described in "Distance" section above.
The ratio value appears in the results window after setting the line segments. ­ Select and drag on the measurement caliper to adjust the position of the end point. ­ Select and drag on the measurement caliper to adjust its position.

6.5.6 Application Measurements

Geom Verify
The image will be magnified in the process of image generation. The function of geometric correction is to reduce the measurement error by correction.

Operator's Manual

6 - 11

6 Image Browsing and Processing
Perform the following procedure:
1. Select [Geom Verify] in the measurement menu.
2. Draw a calibration line to display the length of the line.
The measurement correction line has the same function as the "Distance" measurement. After the calibration line is drawn, a dialog box pops up to prompt the user to input the actual length of the calibration line. NOTE:
Select [Close] in the input box to clear the correction line and exit the geometric correction function. 3. Select [OK] to correct the current caliper.
After the calibration is completed, the scale changes accordingly, and the icon (Geom verify) is displayed beside the scale, indicating that the calibration has been performed. The calibrated effect is valid for all measurement items. For example, the display length of the measurement calibration line becomes the input length of the user. Select [Geom Verify] again to clear the last geometric correction result and restart the function.
TPLO Measurement
Perform the following procedure:

pl1

TPA ps2

ps1 1

pl2
1. Select [TPLO] in the measurement menu. The TPLO tool appears in the image.
2. Mark the two points: ps1 and ps2 at the edge of the tibial plateau. 3. Mark the two points: The pl1 of the tibial plateau and the pl2 of the accessory bone joint. At the
intersection with line ps1ps2, the vertical line of line pl1pl2 is displayed by a dotted line. 4. Mark the center and radius of the cut tibial head with two points. Modify the center position and
radius. Measurement result:

Operator's Manual

6 - 12

6 Image Browsing and Processing
TPA = Angle between line ps1ps2 and the vertical line of line pl1pl2 Saw Radius: Radius of the circle
i-VHS Measurement
Perform the following procedure:
pv2 5V
pv1

pl1 ps2

ps1 1S

pl2
1. Select [i-VHS] in the measurement menu.
There are 6 adjustable nodes (pl1, pl2, ps1, ps2, pv1, pv2) in the image, and 5V is displayed in the pv1pv2 line. 2. Measure the long axis of the heart (L):
The first point, pl1, is placed on the ventral border of the left main stem bronchial bifurcation, and the second point, pl2, is placed on the more distant point of the cardiac apex, the size that reflects the overall dimensions of the left atrium and left ventricle. 3. Measure the short axis of the heart (S):
Draw a perpendicular to the long axis at the widest cardiac width in the central third of the heart, and place points ps1 and ps2 at the two intersections of the perpendicular and the outline of the heart respectively. 4. Measure the length of 5 vertebrae:
Place the point pv1 at the front end of the fourth vertebra and the point pv2 at the end of the eighth vertebra. Measurement result: VHS (v) = pl1p12 (long axis) + ps1ps2 (short axis)

Operator's Manual

6 - 13

6 Image Browsing and Processing
NOTE: The line pl1pl2 and the line ps1ps2 are vertical to each other.
i-RLAD Measurement
Perform the following procedure:
pv2
5V
pv1

pl1

pa1

ps2

L A

ps1 1S pl2

1. Select [i-RLAD] in the measurement menu.
The i-RLAD tool appears in the image, providing 7 adjustable nodes: pl1, pl2, ps1, ps2, pa1, pv1, pv2.
2. Measure the long axis of the heart (L):
The first point, pl1, is placed on the ventral border of the left main stem bronchial bifurcation, and the second point, pl2, is placed on the more distant point of the cardiac apex, the size that reflects the overall dimensions of the left atrium and left ventricle.
3. Measure the short axis of the heart (S):
Point ps2 is placed at the dorsal junction of dorsal edge of the left atrium and dorsal border of the caudal vena cava; draw a long axis vertical line from point ps2, and place point ps1 at the intersection with the ventral side of the heart.
4. Measure the left atrial boundary:
Place point pa1 at the border of the left atrium.
5. Measure the length of 5 vertebrae:
Place the point pv1 at the front end of the fourth vertebra and the point pv2 at the end of the eighth vertebra.

Operator's Manual

6 - 14

6 Image Browsing and Processing
Measurement result: RLAD (v) = line A (pa1) NOTE:
· The line pl1pl2 and the line ps1ps2 are vertical to each other. · One end of line A is fixed at the intersection of line pl1pl2 and line ps1ps2 and at an angle
of 45° with the short axis and the long axis. Adjust pa1 to confirm the line A.
i-VLAS Measurement
Perform the following procedure:
pv2 5V
pv1 pa1
1

pa2

1. Select [i-VLAS] in the measurement menu.

i-VLAS tools appear in the image, providing 4 adjustable nodes: pa1, pa2, pv1 and pv2. 2. Measure Left Atrium (A):

Place the first point pa1 on the ventral end of the left and right branches of the main trachea, and the second point pa2 on the left atrium at the junction of the posterior vena cava and the dorsal side of the posterior vena cava.
3. Measure the length of 5 vertebrae:

Place the point pv1 at the front end of the fourth vertebra and the point pv2 at the end of the eighth vertebra.
Measurement result:
VLAS (v) = pa1pa2

HD Measurement
Perform the following procedure:

p1 a

c1

1

p2 b
c2

Operator's Manual

6 - 15

6 Image Browsing and Processing
1. Select [HD] in the measurement menu. HD tool appears in the image.
2. Adjust the center and radius, and place the center of the two circles on the center of the two femurs.
3. Drag p1 generates line c1p1 in real time, and place point p1 on the anterior border of the acetabulum.
4. Drag p2 generates line c2p2 in real time, and place point P2 on the anterior border of the acetabulum. Measurement result: Angle between the line c1p1 and the line c1c2; angle between the line c2p2 and the line c1c2. A smaller angle indicates greater joint laxity.
M-VLAS Measurement
Perform the following procedure:
pv2 5V

pv1

pa1

1

pb1

pa2
1. Select [M-VLAS] in the measurement menu. M-VLAS tools appear in the image, providing 5 adjustable nodes: pa1, pa2, pb1, pv1 and pv2.
2. Measure Left Atrium (A): Place the first point pa1 on the ventral end of the left and right branches of the main trachea, and the second point pa2 on the left atrium at the junction of the posterior vena cava and the dorsal side of the posterior vena cava.
3. Measure the distance from the most distal boundary of the left atrium to pa1pa2: Place the third point pb1 at the most distal end of the left atrium and intersect vertically with the pa1pa2 through the line pb1.
4. Measure the length of 5 vertebrae: Place the point pv1 at the front end of the fourth vertebra and the point pv2 at the end of the eighth vertebra.

Operator's Manual

6 - 16

Measurement result: M-VLAS (v) = VLAS + line (pb1) = pa1pa2 + line (pb1)
VHS Measurement
Perform the following procedure:

6 Image Browsing and Processing pv2

pv1 L S
pl1 L
ps2

ps1 1S
pl2
1. Select [VHS] in the measurement menu. There are 6 adjustable nodes (pl1, pl2, ps1, ps2, pv1, pv2) in the image.
2. Measure the long axis of the heart (L): The first point, pl1, is placed on the ventral border of the left main stem bronchial bifurcation, and the second point, pl2, is placed on the more distant point of the cardiac apex, the size that reflects the overall dimensions of the left atrium and left ventricle.
3. Measure the short axis of the heart (S): Draw a perpendicular to the long axis at the widest cardiac width in the central third of the heart, and place points ps1 and ps2 at the two intersections of the perpendicular and the outline of the heart respectively.
4. Measure VHS: Place point pv1 at the most anterior end of the fourth vertebra and adjust the pv2 to ensure that the VHS measurement line is parallel to the cone.
NOTE:
· The line pl1pl2 and the line ps1ps2 are vertical to each other.

Operator's Manual

6 - 17

6 Image Browsing and Processing
· The VHS measurement line consists of two parallel segments, aligned at the pv1 end, and their lengths change with pl1pl2 and ps1ps2 respectively, and are equal to them.
RLAD Measurement
Perform the following procedure:
pv2

pv1 L
S A

pl1

pa1

ps2

L A

ps1 1S pl2
1. Select [RLAD] in the measurement menu.
The RLAD tool appears in the image, providing 7 adjustable nodes: pl1, pl2, ps1, ps2, pa1, pv1, pv2. 2. Measure the long axis of the heart (L):
The first point, pl1, is placed on the ventral border of the left main stem bronchial bifurcation, and the second point, pl2, is placed on the more distant point of the cardiac apex, the size that reflects the overall dimensions of the left atrium and left ventricle. 3. Measure the short axis of the heart (S):
Point ps2 is placed at the dorsal junction of dorsal edge of the left atrium and dorsal border of the caudal vena cava; draw a long axis vertical line from point ps2, and place point ps1 at the intersection with the ventral side of the heart. 4. Measure the left atrial boundary:
Place point pa1 at the border of the left atrium. 5. Measure RLAD:

Operator's Manual

6 - 18

6 Image Browsing and Processing
Place point pv1 at the most anterior end of the fourth vertebra and adjust the pv2 so that the RLAD measurement line is parallel to the cone. Reference range: greater than 1.8 indicates left atrium dilatation (increase) NOTE:
· The line pl1pl2 and the line ps1ps2 are vertical to each other. · One end of line A is fixed at the intersection of the line pl1pl2 and the line ps1ps2 and at an
angle of 45° with the short axis and the long axis. Adjust point pa1 to confirm the line A. · The RLAD measurement line consists of three parallel lines, aligned at the pv1 end. The
length of the three lines changes with the pl1pl2, ps1ps2, and pa1 to the intersection of pl1pl2 and ps1ps2, and are equal to them.
VLAS Measurement
Perform the following procedure:
pv2
pv1 pa1
1
pa2
1. Select [VLAS] in the measurement menu. VLAS tools appear in the image, providing 4 adjustable nodes: pa1, pa2, pv1 and pv2.
2. Measure Left Atrium (A): Place the first point pa1 on the ventral end of the left and right branches of the main trachea, and the second point pa2 on the left atrium at the junction of the posterior vena cava and the dorsal side of the posterior vena cava.
3. Measurement VLAS: Place point pv1 at the most anterior end of the fourth vertebra and adjust the pv2 so that the VLAS measurement line is parallel to the cone. Reference range: normal range is 1.4 -2.2, and 2.3 indicates left atrium dilatation (increase). NOTE: The length of VLAS measurement line varies with the length of pa1pa2 and is equal to it.

Operator's Manual

6 - 19

6 Image Browsing and Processing
6.6 Image Postprocessing
On the Exam screen, select [Adv Proc], the parameter adjustment subpane is displayed, as shown in Figure 6-3. You can adjust the image postprocessing parameters. Figure 6-3 Image Postprocessing Menu
1 2 3
4

5 6
7

1.

Select to exit the image post-processing menu.

2.

· [Save]: to save the set parameters. The saved parameters will be the default ones for the

next exam.

· [Load]: to restore the default parameters and reprocess the image.

· [Save as]: to save the current parameter as a set of custom parameters. After selecting the button, the input parameter naming dialog box is displayed.

3.

Select the drop-down box to select the enhanced processing solution.

4.

Select to switch among different parameter groups. The corresponding parameter

adjustment panel is displayed.

The parameters can be adjusted by dragging the slider horizontally or selecting the Increase and Reduce buttons

5.

Set whether to implement grid shadow suppression on the image.

6.

Select to load the original image and reprocess the image with the default post-processing

parameters of the current body position.

7.

Load the default post-processing parameters of the selected body part.

Operator's Manual

6 - 20

Simplified Parameter Mode

6 Image Browsing and Processing

Parameter Brightness Contrast Sharpness Denoise

Description
The smaller the value, the dimmer the image. The greater the value, the brighter the image.
The smaller the value, the weaker the image contrast and layering effect. The greater the value, the higher the image contrast and layering effect.
The smaller the value, the weaker the image sharpness and less details. The greater the value, the higher the image sharpness and more details.
The smaller the value, the stronger the image noise. The greater the value, the weaker the image noise.

Optimization Curve Adjustment (Look-up Table)
You can change the global image effect by adjusting the optimization curve, to achieve the desired image quality. The optimization curve is determined by the control point position, slope of the curve passing the control point, and the optimization parameters for the curve endpoint.

2

Y

1

3

X axis Y axis 1 2 3

Input pixel Output pixel Tail end parameter 1 Tail end parameter 2 Control point

The specific adjustment parameters are as follows:

Operator's Manual

X
6 - 21

6 Image Browsing and Processing

Parameter X-axis (brightness) of the optimization curve control point
Y-axis of the curve control point
Slope (contrast) of the curve control point Curve endpoint optimization parameter (Tail)

Description
You can change the value of Brightness to adjust the image brightness. Increase the parameter value, and the entire image (including bone tissue and soft tissue) becomes brighter. Reduce the parameter value, and the entire image becomes dimmer.
Change the y-axis to change the display effect of the soft issue on the image. When Y is changed to a smaller value, the soft tissue on the image becomes brighter. When Y is changed to a greater value, the soft tissue on the image becomes dimmer.
Change the value of Contrast to change the global contrast of an image.
This parameter is used to change the shape of the curve tail, in order to change the display effect and customization of the soft tissue close to the background. Compared with tail parameter 1 of the curve, tail parameter 2 displays more soft tissue close to the background.

Dynamic Compression
You can reduce the dynamic range of an image to display more specific information.

Parameter Compression
Beta
Correction Factor

Description
You can reduce dynamic compression to reduce the change range of overall brightness trend of the image and reduce the significant black-andwhite difference.
This parameter is used to change the display range of the high-density tissue area (which corresponds to the white part of the image), in order to improve the display details.
This parameter is used to control the correction degree of the gray value. The following figure shows how the corrected value changes with the gray value.
b+

0 a
a: Gray value; b: Correction value

Operator's Manual

6 - 22

6 Image Browsing and Processing
Detail Enhancement
Detail enhancement only changes the local detail contrast of an image. It does not affect the global display effect.

Parameter Size 1
Size 2 Size 3
Size 4 Size 5 Size 6 and Size 7 Enhancement factor

Description
This parameter is used to change the thickness of the corresponding side to the contrast of a detail (such as trabecular bone) of the pixel size of a source image. The smaller the parameter value, the higher the enhancement level, and vice versa. When the parameter is set to 1.0, the details of the specific layer are not enhanced. The detailed graph of the layer contains many noises. Therefore, the noises are increased when the details are enhanced.
This parameter is used to change the thickness of the corresponding side to the contrast of a detail of the pixel size of two source images. The detailed layer graph corresponding to the parameter also contains many noises. The noise size is greater than that of Size 1 and therefore is of greater granularity.
This parameter is used to change the thickness of the corresponding side to the contrast of a detail of the pixel size of four source images. The noises corresponding to the parameter are less than those of Size 1 and Size 2. However, the noise size of Size 3 is greater than those of Size 1 and Size 2 and therefore is of greater granularity. The noises of great granularity are increased when the detail contrast of the layer graph is enhanced.
This parameter is used to change the thickness of the corresponding side to the contrast of a detail of the pixel size of eight source images. The layer graph corresponding to the parameter contains few noises. Therefore, no obvious noises are caused when the layer graph is enhanced.
This parameter is used to change the thickness of the corresponding side to the contrast of a detail of the pixel size of 16 source images. The layer graph corresponding to the parameter contains no noises.
These two parameters are used to change the thickness of the corresponding side to the contrast of a detail of the pixel size of 32 and 64 source images, respectively. When the values of Size 5, Size 6, and Size 7 are reduced, the contrast between large tissues is increased whereas the noise display is less obvious.
The value ranges from 0 to 3.5. It is an overall adjustment value of image enhancement. The smaller the value, the lower the sharpness and the softer the image is. The greater the value, the higher the sharpness and contrast are, and the clearer the details.

Noise Suppression
Noise suppression is used to suppress the image noise and smooth the image.

Operator's Manual

6 - 23

6 Image Browsing and Processing

Parameter Noise Suppression Sigma Noise Threshold
Fine Grain Middle-sized Grain Large-sized Grain

Description
· 0: No noise reduction · 1: Minor noise reduction · 2: Moderate noise reduction · 3: Major noise reduction
Controls the impact on noise reduction that is caused by the pixels near the current pixel. When the distance between the surrounding pixels and the current pixel is the same, a greater parameter value indicates less impact on noise reduction that is caused by the surrounding pixels, and vice versa.
Distinguishes between noises and details. When the contrast between a pixel in a detailed layer graph and another pixel in the surrounding area is less than the noise criterion value, the pixel in the surrounding area has great impact on the noise reduction of the current pixel. Otherwise, the pixel in the surrounding area has minor impact on the noise reduction of the current pixel. In this case, the details between the two pixels are retained.
Suppresses fine-grained noises. Noise suppression reaches the maximum level when the parameter value is 0, and reaches the minimum level (noise is not suppressed) when the parameter value is 1.0. When the parameter value changes gradually from 0 to 1, the level of noise suppression decreases.
Suppresses medium-grained noises. Noise suppression reaches the maximum level when the parameter value is 0, and reaches the minimum level (noise is not suppressed) when the parameter value is 1.0. When the parameter value changes gradually from 0 to 1, the level of noise suppression decreases.
Suppresses coarse-grained noises. Noise suppression reaches the maximum level when the parameter value is 0, and reaches the minimum level (noise is not suppressed) when the parameter value is 1.0. When the parameter value changes gradually from 0 to 1, the level of noise suppression decreases.

Auto WL Ratio
The related parameters of the auto window factor can affect the result of the auto calculated window level. The auto window left scale factor adjustment will increase the left end position of the window, the auto window right scale factor adjustment will decrease the right end position of the window, and the auto window adjustment will increase the overall position of the window.

6.7 Image Stitching
The console software provides the image stitching function to stitch the images of body parts such as the spine and legs of the animal in different areas to form a complete single image.

Operator's Manual

6 - 24

6 Image Browsing and Processing

6.7.1 Starting Stitching Management

Perform the following procedure:

1. Select the images to be stitched (the selected images are marked) on the "Exam" screen.

2. Select [Stitch] to start stitching management.

The stitched image is automatically saved and displayed in the thumbnail area. The stitching position can be adjusted manually through image panning and stitching point selection. Select [EditStitch] to switch to the main screen of image stitching. The main screen of image stitching mainly consists of the following functional areas:

B

C

A

Stitch Delete
D

A Thumbnail management area: displays the thumbnails of the stitched images. B Area 1 of image stitching: displays the upper image corresponding to the current stitching seam. C Stitching result display area: displays the overall stitching effect. D Area 2 of image stitching: displays the lower image corresponding to the current stitching seam.
You can perform, adjust, and confirm stitching using the following controls.
Moves the selected image upward. If the selected image is on top, no operation is performed. Moves the selected image downward. If the selected image is at the bottom, no operation is performed.

Operator's Manual

6 - 25

6 Image Browsing and Processing

Delete image Pick Mark Body thickness Distance Fusion degree Restitch
Save

Deletes the selected image.
Selects the stitching point between images in area B and area D.
Indicates the maximum distance between the front and rear surface of the body.
Indicates the distance between the center of the body and the surface of the FPD.
Indicates the degree of simultaneous display of two neighboring images in the overlap area.
If the automatic stitching results do not meet your expectation, you can select [Restitch] to enter the manual stitching state.
You can also discard the manual stitching results and restore to the initial state of automatic stitching.
Accepts the current stitching results, saves the stitched image, and saves the image to the database.

Image Panning
Image panning is the process of manually moving two images to be stitched relative to each other to adjust the stitching position.
Perform the following procedure: 1. Observe the stitching rendering in area C to identify the area to be adjusted.
You can also observe the thumbnails in area A to determine the approximate position of the stitching seam. Areas B and D display the upper and lower images corresponding to the selected stitching seam and the stitching relationship. 2. Click in the area D, a blue box is displayed.
3. Click and drag to pan the image in area D to adjust the stitching position of the images in area B and area D, and observe the adjustment effect in area C.
Release the mouse after adjustment. 4. If you want to continue to adjust the stitching relationship between other neighboring images,
repeat steps 1.~3.
5. After adjusting the stitching of all images, select [Save] to save the images.
Stitching Point Selection
Coincident point selection is the method of adjusting the stitching position by superposing the stitching points of two images.
Perform the following procedure: 1. Observe the stitching rendering in area C to identify the area to be adjusted.
2. Click in the area D, a blue box is displayed.
3. Select [Pick Mark] on the main screen of image stitching. 4. Click the images in area B and area D to select the stitching points.

Operator's Manual

6 - 26

6 Image Browsing and Processing
The selected stitching points are marked by a red cross. 5. Select [Restitch].
The system combines the images in area B and area D by superposing the stitching points. You can observe the adjustment effect in area C. If the stitching effect does not meet your expectation, select [Pick Mark] to reselect stitching points. 6. If you want to continue to adjust the stitching relationship between other neighboring images, repeat steps 1.~5. 7. After adjusting the stitching of all images, select [Save] to save the images
6.8 Image Confirmation
The images generated during exposure are automatically saved. After images are tagged or postprocessed (adjusted), select [Save] to save the images. Select [Save As] to save an image as a new image. The originally saved image remains unchanged.
NOTE:
Select [Cancel] to return to the status when the image is saved.

Operator's Manual

6 - 27

7 Animal Data Management

7 Animal Data Management

An exam record consists of all information and data of one exam. An exam record consists of the following information:
· Animal basic information and exam data · Image files · Report
The "History" screen provides the following functions:
· Query and display animal information · Edit animal information · Delete animal information · Send and print animal images · Back up and restore animal data
NOTE:
· DO NOT use the internal hard drive for long-term image storage. Daily backup is recommended. External storage media is recommended for archiving images.
· The system animal database space is limited, please back up or clear animal data in time. · Selecting the compressed format to export the image may cause image distortion. · The manufacturer is not responsible for lost data if you DO NOT follow suggested backup
procedures.

7.1 Searching an Animal
Perform the following procedure: 1. Select the data source.
Select [Data Source] to select the data source of animal data, the system animal database is default. 2. Set search conditions in the "Filter" drop-down list. 3. Enter the key word. The matching animal information is displayed in the animal list. When you select an animal exam in the animal list, the images of this animal will be displayed at the bottom of the screen.

Operator's Manual

7 - 1

7.2 Animal Data View and Management
NOTE:
Only administrators or authorized operators are allowed to delete data.

7 Animal Data Management

Select the desired animal information in the list.

Item Backup
Restore Lock Edit Delete Delete Image

Description Backup the selected animal data to the system-supported media. You can backup the animal data to DCM files by selecting "Copy Origin Dcm File". You can also set whether to remove the animal data from local hard disk after Backup. Import the animal data from an external media. Lock the selected animal exam. The locked animal examination cannot be deleted. View the details of the animal and modify the animal information. To delete an exam: select the desired exam and select [Delete] below the animal information list area. To delete an image: 1. Select an animal exam.
The image(s) acquired by the exam is displayed in the thumbnail area. 2. Select [DeleteImage] on the right of the thumbnail area, and select the image(s)
to be deleted. 3. Select [OK].

7.3 Image Compare

On the Exam screen, select [Diag Ref].
The title bar of the left image displays the current animal ID, animal name, and animal type.
Common processing tools are provided for performing image display operations. The following figure shows the corresponding icons.

Icon

Name

Description

Restore images Load the original image.

Invert

Flips an image between positive and negative display.

Gray Window/ Switch between the window width/position adjustment and the image

Roam

roaming.

Operator's Manual

7 - 2

7 Animal Data Management

Icon

Name

Paging /

Description
Select or to turn to the previous or next page to view the films on each page.
The page number is displayed between the paging buttons. The first digit indicates the current page number, and the second digit indicates the total pages.

History Image
· Select (Review) tab. The history image library of the previous exam with the same animal ID is loaded by default. The first image with the same position on the left is displayed. The right title bar displays the currently referenced animal ID, animal name, and animal type.
· Select (Review) again to display all thumbnails of the current exam. Select a thumbnail to load the image to the right image display area.
· A search bar is provided on the right side of the title bar. The search method equals to fuzzy search when "All" is selected in the history image. The search result is an exam that meets the requirement. Select an exam to load the image of the exam.

Standard Image
Select (Standard) tab, and the standard film of the left image is displayed by default. The right title bar displays the position of the current standard slide.
Select (Standard) tab again to display the thumbnails of all optional standard sheets. Select a thumbnail to load the image to the right image display area.

Pathological Image
Select (Case) tab. The image on the left of the tab is loaded by default, and the first one on the right of the tab is displayed in the image display area. The right title bar displays the symptoms corresponding to the current pathological slice.
Select (Case) again to display the thumbnails of all the images. Select a thumbnail to load the image to the right image display area.
Application Measurement
Select (Measured), and the first application measurement image is displayed by default. The title bar on the right side displays the name of the application measurement.
Select (Measured) again to expand the thumbnails of all application measurements. Select a thumbnail to load the image to the right image display area.

7.4 Sending Animal data
Perform the following procedure: 1. Select one or more animal exams from the animal information list. 2. Select from the destination, and set related settings.

Operator's Manual

7 - 3

7 Animal Data Management

7.5 u-Link (applicable for CE region only)
u-Link is used to connect the system with software applications which support the u-Link protocol.
7.6 Recycle Bin
The recycle bin is used to store deleted animal data, exam data and images. The system supports recovery of these data from the recycle bin. Select [Recycle Bin] to enter the "Recycle Bin" screen. Select items to be recovered in the list. · [Restore]: to restore the selected item back to history exam. · [Delete]: to delete the item permanently, and the item can never be restored again.
7.7 Image Print
On the history exam screen, select the desired exam to be printer, and select [Print] to enter the image print management screen. The print management function allows you to arrange the layout of and adjust the received images.
7.7.1 Thumbnail Area Operations
Adding an Image to a Film
Do one of the following to add an image: · Select a thumbnail to add the image to the first empty film cell. When no empty film exists, an
empty film is automatically created and the image is added to the film. · Select a thumbnail, and drag and drop it to the target film cell. Then, the image is added to the cell.
Scrolling to Display Thumbnails
When thumbnails are displayed on more than one page, you can drag the thumbnail area for position adjustment. You can also select the up and down arrows for position adjustment.
Clearing Thumbnails
Select [Clear] to clear the image(s) in the thumbnail area.

Operator's Manual

7 - 4

7 Animal Data Management
7.7.2 Image Adjustment
You can adjust the image in the image preview area before printing. The adjustment functions are the same as those on the image browsing screen, including:

Icon Name Image roaming

Description In the image display area, click and drag to pan the image.

Rotation

Rotate the image at any angle.

Image dragging

After selecting this button, click in the image preview area to exchange images between different cells and films by dragging and dropping.

Default Default

Restores an image to the state prior to adjustment.

Animal information display

Displays animal details on an image, including the basic animal information and exposure parameters.

Positive/Negative image Flips an image between positive and negative display.

90º left rotation

Rotates left by 90º.

90º right rotation

Rotates right by 90º.

Up and down mirroring Reverses image display up and down.

Left and right mirroring Reverses image display left and right.

7.7.3 Print Typesetting
The following table lists the functions and operations of typesetting tools:

Icon 14*17

Name Print format
Film size Film orientation

Description
Select different icons to select different print formats.
When you switch a print format, the left-side page is adjusted accordingly. The following figure shows the examples of various print formats.
You can select the desired film size. When the current film size is changed, the image added to the film changes its zooming status.
Select different icons to select film print orientations.
After a film print orientation is selected, the layout of the left-side page is rearranged accordingly. When the current orientation is changed, the image added to the film changes its zooming status.

Operator's Manual

7 - 5

7 Animal Data Management

Icon

Name

Description

Multi-film preview

Select the film preview window format.

/

New Film Select to add a new and empty film to the image preview area on the left

as the last page.

By default, the new film adopts the currently selected format.

/

Del Img

In the film preview window, select a film cell of the image to be deleted.

NOTE:

Select the icon on top right of the film grid to delete the image in the current film cell.

/

Del Film

Select to delete the current film. The total number of film pages reduces

by 1.

When the film preview window displays only one film, the displayed film is the current film. When the film preview window displays multiple films, the current film is highlighted.

/

Del All

Select to delete all films. Then, the total number of film pages becomes

0.

Paging /

When films are displayed on multiple pages of the printing screen, select or to turn to the previous or next page to view the films on each
page.

The page number is displayed between the paging buttons. The first digit indicates the current page number, and the second digit indicates the total pages.

7.7.4 Setting the Print Parameter
Setting a Print Node
By default, the Print Server List drop-down list displays the default node in the Setup chapter.
When no default node is configured, the first node is displayed by default.
When multiple nodes are configured, select one from the drop-down list.
Setting the Number of Printed Copies
· Select [+] to increase the number of printed copies. · Select [­] to reduce the number of printed copies. The number of copies is displayed in the middle.
Setting Image Fit Format
· Adaptive printing When you create a film on the film management screen, [Auto Fit] is selected by default. Adaptive printing indicates that the added image fills the film entirely. The image automatically adapts itself to the film cell size to fill the cell entirely when you change the film size, film orientation, and film typesetting, or rotate and flip the image on the film.
· Equal-proportion printing

Operator's Manual

7 - 6

7 Animal Data Management
Equal-proportion printing indicates that an image is printed to a film in a specific proportion of the actual image size. For example, if the 1.0 proportion is selected, the image printed to the film is of the actual size. Select [Actual Size], and select an option from the proportionality factor dropdown list. · Apply to all Switching between adaptive printing and equal-proportion printing applies only to the current film. After you select [Apply to All], adaptive printing or equal-proportion printing selected for the current film is applied to all the other films.
7.7.5 Print
Printing the Current Film
Select [Print Current] to print the current film to the selected printer node.
Printing All Films
Perform the following procedure: 1. Select [Print All] to print all films to the selected printer node.
You are prompted for operation confirmation before actual printing. 2. Select [OK] to start printing; select [Cancel] to cancel the printing.
7.8 Animal Task Management
Select in the system status icon area of the screen to bring up the task management screen. You can perform the following operations: · Select [Delete] to delete the task. · Select [Retry] to retry the failed task.
7.9 DICOM Task Management
Select in the system status icon area of the screen to bring up the DICOM task management screen. Perform the following procedure: 1. Select task type: Storage, Print and All. 2. Set search conditions:
­ Select task status. ­ Select the time range. 3. Select [Query] to view the tasks. 4. Perform the following operations:

Operator's Manual

7 - 7

­ Select [Delete] to delete the selected task. ­ Select [Redo] to retry the failed task.

7 Animal Data Management

Operator's Manual

7 - 8

8 Report Management

8 Report Management

Select [Report] in the Exam process icons area to enter the report management screen.

8.1 Select a Report

Perform the following procedure:

1. On the Report management screen, select list:

(Report list button) to expand the history report

­ "No report" tab: check the list of unsaved reports. ­ "Reported" tab: check the list of saved reports.

2. Search a report:

Select the query conditions and input the query keyword in the text box to fuzzy search for the animals meeting the conditions.
Select to restore the default query conditions and view the animals that are examined on the current day.

3. In the queried report list, select the report to be edited to enter the report editing screen.

4. Select

(Report list button) to close the history report list.

8.2 Edit a Report
NOTE:
To edit animal information, select [Edit] button on the "History" screen.

8.2.1 Adding Images
When an animal exam report is selected, the image of the animal exam is automatically loaded in the image thumbnail area. Select the image thumbnail to preview the image. Add images to the report in any of the following ways: · Click to add the image to the report. The selected image displays "".

Operator's Manual

8 - 1

8 Report Management
If you need to add different images in the report diagnosis area, select the images in the thumbnail area in turn. · In the "Diagnostic image" box, right-click and select [Add] to add the image in the thumbnail area to the box area. The thumbnail selection box of the added image is displayed as "". Right-click the image in the "Diagnostic image" box, and the menu pops up: · Select [Move to the left] or [Move to the right] to adjust the image order. · Select [Delete] to delete the added images from the report.
8.2.2 Adding Diagnosis Description
Adding from the Library
Perform the following procedure: 1. Select to expand the knowledge base in the library area of the report management screen. 2. Expand the nodes at all levels and double-click to add the image presentation and impression to
the report body directly. Double-click other nodes in the knowledge base. Add a new node to the text area of the report. To replace the content of the report body with the content of the knowledge base, right-click the knowledge base screen and select [Replace report] to replace the content of the report body with the content of the knowledge base.
Adding from the Report Body
Perform the following procedure: 1. Select the "Image presentation" or "Impression" area in the body of the report to locate the cursor. 2. Use the keyboard to edit and enter the description of the condition.
Right-click the mouse and a context menu pops up. Common context menus and shortcuts such as [Select All] and [Copy] are supported.
8.2.3 Setting Positive/Negative
In the lower part of the text area, you can directly define the case as positive or negative. The default value is "Unknown". By selecting the drop-down arrow button and selecting the corresponding value, the value will be saved when the report is saved.
8.3 Save a Report
When the report is finished and the current content needs to be saved, select [Save] in the operation area to save the current report text and the positive/negative status. The [Save] button turns gray. The [Save] button is activated when the report is edited or the positive/negative result is switched.

Operator's Manual

8 - 2

8 Report Management
8.4 Preview a Report
Select [Preview] in the operating area to display the review screen of the generated report.

8.5 Print a Report
Select [Print] to print the report, or select [Preview] > [Print] to print the report. A confirmation dialog box of printing report pops up. Select [Yes] to print the report to the connected printer. After the report is printed, is marked on the upper-right corner of the report.

8.6 Deleting a Report

Perform the following procedure:

1. On the report management screen, select list:

(Report list button) to expand the history report

­ "No reported" tab: check the list of unsaved reports. ­ "Reported" tab: check the list of saved reports.

2. Search a report:

Select the query conditions and input the query keyword in the text box to fuzzy search for the animals meeting the conditions.
Select to restore the default query conditions and view the animals that are examined on the current day.

3. In the queried report list, select the report to be deleted and select .

8.7 Knowledge Base Management
TIP:
Administrators can edit the public knowledge base template, while common users can only edit the template under the individual account.

Select to expand the knowledge base in the library area of the report management screen.
The factory default value has a public knowledge base node. Users can add or modify the content of the knowledge base.
The template of the knowledge base is a tree structure, which can be divided into public template directory and personal template directory.
Select an item in the template, and the "Image Presentation" and "Impression" will be displayed in the content display area.

Operator's Manual

8 - 3

8 Report Management
Double-click the template directory node to show or hide the directory. You are allowed to import or export the knowledge base. For details, see the Report Configurations section in the Setup chapter.
8.7.1 Adding a Knowledge Base
Perform the following procedure: 1. Right-click the main node and select [Add a knowledge base directory] from the context menu. 2. Add database directory node:
a. Select a node as required, and then right-click the node to display the shortcut menu. b. Select [Add a knowledge base directory] from the shortcut menu, and add a sub-knowledge
base catalog node under this node. c. Select the new knowledge base directory node, and then select [Edit] to edit the name in the
Name text box. 3. Add knowledge base template:
a. Select a node and right-click to display the shortcut menu. b. Select [Add a knowledge base template] from the shortcut menu. c. Select [Edit] to edit in the "Name", "Image Presentation" and "Impression" text boxes.
8.7.2 Modifying a Template
Perform the following procedure: 1. Double-click the template directory node to expand the knowledge base entry. 2. Choose a template node to edit, and then select [Edit].
The template name and content are editable, and the cursor is automatically located at the template name. 3. Edit contents in the corresponding text box. After editing: ­ [Save]: To overwrite the modified settings. ­ [Save as]: Save the modified template as a knowledge base. The original template remains
unchanged. Save the saved template in the same directory. ­ [Cancel]: Cancel the modification.
8.7.3 Saving Report Contents to a Template
Perform the following procedure: 1. Double-click the template directory node to expand the knowledge base entry. 2. Select the knowledge base template in the knowledge base directory node.

Operator's Manual

8 - 4

8 Report Management
3. In the report editing area, right-click in the "Image presentation" or "Impression" area to display the shortcut menu.
4. Select [Save to the knowledge base] to replace the "Image Presentation' and "Impression" in the selected knowledge base template.
5. Select [Save] to save the replacement.
8.7.4 Adjusting Contents
Perform the following procedure: 1. Double-click the template directory node to expand the knowledge base entry. 2. Select the knowledge base template in the knowledge base directory node. 3. Select a node and right-click to display the shortcut menu.
­ Select [Move up] or [Move down] to adjust the order of the directory nodes or templates. ­ Select [Delete] to delete the directory node or template.

Operator's Manual

8 - 5

9 System Maintenance

9 System Maintenance

CAUTION
Failure to perform periodic inspection and maintenance may cause the working conditions to deteriorate without the user's knowledge and thus result in equipment faults, damage, and even injury.
A periodic maintenance and repair program must be established to ensure continuous safe performance of the system. It is the user's responsibility to check the equipment status and perform preventive inspection of the system.
There are two levels of maintenance. One level is maintenance by the user or operator, and the other level is maintenance by professional X-ray maintenance personnel.
9.1 Operator's Tasks
WARNING
· After the system starts, do not perform cleaning or maintenance of any component of the equipment. Be sure to shut down the system and disconnect the main power supply before cleaning or maintenance. Disconnect the cables between all the operated components and the system during cleaning, maintenance, and inspection.
· Ensure that no water or other liquids enter the system, to avoid electrical system short circuit or component corrosion.
· Because some substances of cleaning agents are harmful to the human body, the concentration of such substances in air cannot exceed the upper limit specified by relevant rules and regulations. Strictly observe the use instructions provided by the manufacturers of such cleaning agents.
· After cleaning and disinfection, fully ventilate the room and then power on the device. Residual flammable gases in the room may cause fire and explosion when the system is powered on.
· Do not use any other cleaning agents or solvents because they may darken the painting or smudge printed text.
· Do not directly clean the surface of the equipment with a cleaning spray, because the spray mist may seep into the equipment, damage electronic components, and form a combustible mixture of air and water vapors, posing a safety hazard.

Operator's Manual

9 - 1

9 System Maintenance
· Do not remove any shelf cover. Do not remove or handle the internal components of the equipment. THESE ACTIONS COULD CAUSE SERIOUS BODILY INJURY AND/OR SYSTEM DAMAGE.
Perform the following procedure:
1. Power off the system and cut off the power supply to the equipment.
2. Check whether the cables between the main components of the system are connected properly.
3. Clean regularly.
You are advised to clean the device and all contaminated parts twice per day, and the duration of cleaning the same part each time does not exceed 3 minutes, especially for the parts in contact with animals, including the flat panel detector, detector tray of the radiography stand, and handrail. Soak a soft cloth in ethanol (75%) or isopropanol (70%) solution. Then use it to clean the enclosure and surface of the device. After cleaning, use a soft cotton cloth to dry the display/touch screen immediately, and wipe off the contrast agent stains (if any) as soon as possible.
9.2 Service Tasks
NOTE:
· Perform FPD calibration and maintenance periodically to keep the performance at a constant level. Refer to the Setup chapter of this manual for operation instructions of detector calibration.
· If you find abnormal sound, operation failure, equipment damage, uncontrolled patient support movement, or uncontrolled X-ray operation during use, please contact the Customer Service Department or your local distributor.
Only service engineers trained on medical X-ray equipment can perform maintenance and repair of this system. The equipment manufacturer recommends that initial system maintenance be performed after one month and before the third month following system installation and operation. In normal cases, maintenance should be performed every 12 months based on the system operation status.
If many animals are examined every day (for example, 125 animals per day) after system installation, the maintenance cycle should be shortened. For example, perform full maintenance every 6 months.

Operator's Manual

9 - 2

10 System Specifications

10 System Specifications

10.1 Operating Environment
· CPU: Intel Core 2/3.0GHz or above · RAM: 4GB · Hard Disk: 500GB (The partition should contain disk D) · Software Environment
­ Windows 10 64bit (Enterprise LTSC) ­ Windows 11 64bit (Home, Professional, Enterprise, Education) · Monitor Resolution:1920×1080; Zoom Factor: 100% · Network Condition: Gigabit Ethernet
10.2 Technical Specifications
10.2.1 Environmental Conditions

Item Operating conditions
Storage and transportation conditions

Relative humidity Atmospheric pressure Ambient temperature Relative humidity Atmospheric pressure Ambient temperature

Value 20 - 75 62 kPa to 106 kPa 10°C to 38°C 20 to 90 62 kPa to 106 kPa -20 °C to 55°C

Operator's Manual

10 - 1

10.2.2 Power supply

10 System Specifications

Item Router
Power supply box

Voltage Frequency Current Voltage Frequency Input power

Value AC single-phase, 100-240V 50/60Hz±1Hz 1.0A max AC single-phase, 100-240V 50/60Hz±1Hz 150VA

10.2.3 Flat panel detector

NOTE:
When the AMM4343RP and AMM3543RP FPD communicate in wireless mode, the wireless detector transmits data at a frequency of 2.4 GHz or 5 GHz; according to different frequency bands, the radio output power limits of AMM3543RP is as follows: no higher than 20 dBm in the frequency range of 2400 MHz to 2483.5 MHz, no higher than 23 dBm in the frequency range of 5150 MHz to 5350 MHz, and no higher than 33 dBm in the frequency range of 5725 MHz to 5850 MHz. AMM3543RP/ AMM4343RP communication protocols compliant with IEEE 802.11n or IEEE 802.11ac.

AMM4343RP

Item Input voltage Power frequency Scintillator fluorescent material Detection unit array material Valid FPD dimensions Collection matrix Pixel size A/D conversion (number of valid data bits) MTF

Value
AC single-phase 100-240V
50/60Hz±1Hz
CsI
a-Si
43 cm x 43 cm
3072 x 3072 pixels 140 m 16 bits
Under the condition of RQA5 and when spatial frequencies are 1.0 lp /mm and 2.0 lp /mm, the MTF typical values are 0.55 and 0.3. The deviation does not exceed -0.04 and there is no upper limit.

Operator's Manual

10 - 2

Item DQE
Spatial resolution Imaging time Weight Dimensions Lithium battery
kV range Protection class Weight-bearing Drop AED
AMM3543RP

10 System Specifications
Value Under the dose condition of RQA5 and 10 Gy and when the spatial frequencies are 0 lp/mm, 0.5 lp/mm, 1.0 lp/mm, 1.5 lp/mm, 2.0 lp/mm, 2.5 lp/ mm, 3.0 lp/mm and 3.5 lp/mm, the DQE typical values are 0.7, 0.53, 0.43, 0.36, 0.32, 0.27, 0.21, and 0.14 respectively. The deviation does not exceed -0.06 and there is no upper limit. 3.6 LP/mm 7.6s 4 kg (deviation  +15%) 460 mm x 460 mm x 15.5 mm 3800 mAh 500 charge-discharge cycles 40 kVp to 150 kVp IP54 200 kg on the surface of the whole plate and 100 kg within a diameter of 40 mm from the center  1 m Supported

Item Input voltage Power frequency Scintillator fluorescent material Detection unit array material Valid FPD dimensions Collection matrix Pixel size A/D conversion (number of valid data bits) MTF

Value AC single-phase 100-240V 50/60Hz±1Hz CsI a-Si 35 cm x 43 cm 2560 x 3072 pixels 140 m 16 bits Under the condition of RQA5 and when spatial frequencies are 1.0lp/mm and 2.0lp/mm, the MTF typical values are 0.55 and 0.3. The deviation does not exceed -0.04 and there is no upper limit.

Operator's Manual

10 - 3

Item DQE
Spatial resolution Imaging time Weight Dimensions Lithium battery
kV range Protection class Weight-bearing Drop AED

10 System Specifications
Value Under the dose condition of RQA5 and 10 Gy and when the spatial frequencies are 0 lp/mm, 0.5 lp/mm, 1.0 lp/mm, 1.5 lp/mm, 2.0 lp/mm, 2.5 lp/ mm, 3.0 lp/mm and 3.5 lp/mm, the DQE typical values are 0.7, 0.53, 0.43, 0.36, 0.32, 0.27, 0.21, and 0.14 respectively. The deviation does not exceed -0.06 and there is no upper limit. 3.6 LP/mm 6.7s 3.3kg (deviation  +15%) 460mm x 384 mm x 15.5 mm 3800 mAh 500 charge-discharge cycles 40 kVp to 150 kVp IP54 200 kg on the surface of the whole plate and 100 kg within a diameter of 40 mm from the center  1 m Supported

Operator's Manual

10 - 4

11 Guidance and Manufacturer's Declaration

11 Guidance and Manufacturer's
Declaration
11.1 EMC
The system complies with the EMC standard IEC 60601-1-2:2014+A1:2020 as well as Safety restrictions on electromagnetic radiation exposure from CE.
Intended Environments: professional healthcare facility environment (except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging).
WARNING
· The use of unapproved accessories may diminish system performance. · Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. · 3. Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions exposure beyond intended use or decreased electromagnetic immunity of this equipment and result in improper operation. · 4. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the system, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. · 5. Use of portable or mobile communications devices can degrade the performance of the equipment. · 6. The system is not intended for use in residential environments and can possibly not provide adequate protection to radio reception in such environments. · 7. Use the system away from heat penetration, diathermy, electrocautery, magnetic resonance imaging, RFID and security equipment (such as electromagnetic anti-theft system and metal detector). If some concealed RF transmitters that are not known to the user are exposed near the device and are disturbed by the device (for example, scanning mode changes or image disturbances affecting diagnosis), the user should immediately take mitigation measures, such as redirecting, repositioning or shielding away from the RF transmitter.

Operator's Manual

11 - 1

11 Guidance and Manufacturer's Declaration
NOTE:
· The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
· recommendations for actions that are known to affect the EMISSIONS and IMMUNITY of equipment throughout the EXPECTED SERVICE LIFE: ­ recommendations for maintenance or service intervals; ­ service procedures to maintain effectiveness of shields and grounds; ­ precautions to take if the use location is near (e.g. less than 1,5 km from) AM, FM or TV broadcast antennas.
· The system needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below.
· Operation of system, in the case that the patient physiological signal is lower than the minimum amplitude or value specified in the product specifications, may cause inaccurate results.
· Portable and mobile RF communications equipment can affects system. See tables 2, 3, and 4 below.
If the system is operated within the electromagnetic environment listed in Table 2, Table 3, Table 4 and Table 5, the system will remain safe and will provide the following basic performances: · Imaging; · Wireless communication · Date information

Table 11-1

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATIONELECTROMAGNETIC EMISSIONS

The system is intended for use in the electromagnetic environment specified below. The customer of the user of the system should assure that it is used in such an environment.

EMISSIONS TEST

COMPLIANCE

ELECTROMAGNETIC ENVIROMENT-GUIDANCE

RF emissions CISPR 11

Group1

The system uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

Operator's Manual

11 - 2

11 Guidance and Manufacturer's Declaration

Table 11-1

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATIONELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The customer of the user of the system should assure that it is used in such an environment.

EMISSIONS TEST

COMPLIANCE

RF emissions CISPR 11
Harmonic Emissions IEC 61000-3-2
Voltage Fluctuations / Flicker Emissions IEC 61000-3-3

Class A Not applicable
Not applicable

ELECTROMAGNETIC ENVIROMENT-GUIDANCE
The system is suitable for use in all establishments other than domestic and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes.

Table 11-2

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATIONELECTROMAGNETIC IMMUNITY

The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

IMMUNITY TEST IEC 60601 TEST LEVEL

COMPLIANCE LEVEL

ELECTROMAGNET IC ENVIRONMENTGUIDANCE

Electrostatic Discharge(ESD) IEC 61000-4-2

±8 kV contact; ±2kV, ±8 kV contact; ±2kV, ±4kV, ±8kV, ±15 kV air ±4kV, ±8kV, ±15kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast Transient ±2 kV for power supply ±2 kV for power supply The power supply

/ burst IEC 61000-4-4 lines;

lines;

quality should be that of

±1 kV for input / output lines

±1 kV for input / output lines

a typical commercial or hospital environment.

Surge IEC 61000-4-5

±1 kV line (s) to line (s);
±2 kV line (s) to earth

±1 kV line (s) to line (s);
±2 kV line (s) to earth

The power supply quality should be that of a typical commercial or hospital environment.

Operator's Manual

11 - 3

11 Guidance and Manufacturer's Declaration

Table 11-2

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATIONELECTROMAGNETIC IMMUNITY

The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

IMMUNITY TEST IEC 60601 TEST LEVEL

COMPLIANCE LEVEL

ELECTROMAGNET IC ENVIRONMENTGUIDANCE

Voltage dips, Short interruptions and voltage variation on power supply input voltage IEC 61000-411

0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°

0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°

Mains power quality should be that of a typical commercial or hospital environment. If the user of the system

0% UT; 1 cycle

0% UT; 1 cycle

requires continued

70% UT for 25/30 cycle at 0°

70% UT for 25/30 cycle at 0°

operation during power mains interruptions, it is recommended that the

0% UT; 250/300 cycle 0% UT; 250/300 cycle system be powered

from an uninterruptible

power supply or a

battery.

Power frequency (50/60 30 A/m HZ) magnetic field IEC 61000-4-8

30 A/m

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE: UT is the A.C. power voltage prior to application of the test level.

Operator's Manual

11 - 4

11 Guidance and Manufacturer's Declaration

Table 11-3

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATION-- ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

IMMUNITY TEST
Conduced RF IEC 61000-4-6

IEC 60601-1-2 TEST LEVEL
3 Vrms
0,15 MHz - 80 MHz
6 Vrms in ISM
bands between
0,15 MHz and 80 MHz

Radiated RF 3 V/m IEC 61000-4-3 80MHz - 2.7GHz

COMPLIANCELE ELECTROMAGNETIC

VEL

ENVIRONMENT-GUIDANCE

3 Vrms

Portable and mobile RF

0,15 MHz - 80 MHz

communications equipment should be used no closer to any part of the

6 Vrms in ISM

system, including cables, than the

bands between

recommended separation distance calculated from the equation

0,15 MHz and 80 applicable to the frequency of the

MHz

transmitter.

Recommended separation distance

d = 1.2 × P

3 V/m

d = 1.2 × P 80 MHz to 800 MHz

80MHz - 2.7GHz d = 2.3 × P 800 MHz to 2.7GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.

Interference may occur in the vicinity of equipment marked with the following symbol:

Operator's Manual

11 - 5

11 Guidance and Manufacturer's Declaration

Table 11-3

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATION-- ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

IMMUNITY IEC 60601-1-2

TEST

TEST LEVEL

COMPLIANCELE ELECTROMAGNETIC

VEL

ENVIRONMENT-GUIDANCE

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

· The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
· Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which system is used exceeds the applicable RF compliance level above, system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating system.
· Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 3V/m.

Table 11-4

GUIDANCE AND MINDRAY ANIMAL MEDICAL DECLARATION-- ELECTROMAGNETIC IMMUNITY

The system is intended for use in the electromagnetic environment specified below. The customer or the user of system should assure that it is used in such an environment.

IMMUNITY TEST IEC 60601 TEST LEVEL

COMPLIANCE LEVEL

ELECTROMAGNET IC ENVIROMENT GUIDANCE

Proximity magnetic 65 A/m

65 A/m

/

fields

134,2 kHz

134,2 kHz

IEC 61000-4-39

Pulse modulation

Pulse modulation

2,1 kHz

2,1 kHz

7,5 A/m 13,56 MHz Pulse modulation 50 kHz

7,5 A/m 13,56 MHz Pulse modulation 50 kHz

Operator's Manual

11 - 6

11 Guidance and Manufacturer's Declaration

Table 11-5 Test specifications and minimum distances

Recommended separation distances between portable and mobile RF communications equipment and The system
The system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the the system as recommended below, according to the maximum output power of the communications equipment. Portable and mobile radio communications equipment (e.g. two-way radio, cellular/ cordless telephones and similar equipment) should be used no closer to any part of this system, including cables, than determined according to the following method:

Test frequency (MHz)

Band(MHz) Service

Modulation Maximum Distance power (W) (m)

Immunity test level (V/m)

385

380 - 390 TETRA 400 Pulse

1.8

0.3

27

modulation

18Hz

450

430 -470

GMRS 460 FM ±5 kHz 2

0.3

28

FRS 460

deviation 1 kHz sine

710

704 - 787 LTE Band

Pulse

0.2

0.3

9

745

13,17

modulation

780

217 Hz

810

800 - 960 GSM 800/900, Pulse

2

870

tetra 800,

modulation

930

iDEN 820,

18 Hz

0.3

28

CDMA 850,

1720 1845 1970

1700 -1990

LTE Band 5 GSM 1800, CDMA 1900, GSM 1900,

Pulse

2

modulation

217 Hz

0.3

28

2450

2400 -2570

DECT,

LTE Band 1, 3, 4, 25, UMTS

Bluetooth, WLAN, 802.11 b/g/n,

Pulse

2

modulation

217 Hz

0.3

28

RFID 2450, LTE Band 7

Operator's Manual

11 - 7

11 Guidance and Manufacturer's Declaration

Table 11-5 Test specifications and minimum distances

Recommended separation distances between portable and mobile RF communications equipment and The system
The system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the the system as recommended below, according to the maximum output power of the communications equipment. Portable and mobile radio communications equipment (e.g. two-way radio, cellular/ cordless telephones and similar equipment) should be used no closer to any part of this system, including cables, than determined according to the following method:

Test frequency (MHz)
5240
5500
5785

Band(MHz) Service
5100 -5800 WLAN, 802.11 a/n

Modulation Maximum Distance power (W) (m)

Pulse

0.2

0.3

modulation

217 Hz

Immunity test level (V/m)
9

Table 11-6

RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATION DEVICE AND SYSTEM

The system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum output power of the communications equipment.

Rated Maximum Output power of Transmitter (W)
0.01

Separation Distance According to Frequency of Transmitter (m)

150kHz -80MHz
d  1.2 P

80MHz-800MHz
d  1.2 P

800MHz-2.7GHz
d  2.3 P

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

Operator's Manual

11 - 8

11 Guidance and Manufacturer's Declaration

Table 11-6

RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATION DEVICE AND SYSTEM

The system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum output power of the communications equipment.

Rated Maximum Output power of Transmitter (W)

Separation Distance According to Frequency of Transmitter (m)

150kHz -80MHz
d  1.2 P

80MHz-800MHz
d  1.2 P

800MHz-2.7GHz
d  2.3 P

For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

If system image distortion occurs, it may be necessary to position system further from sources of conducted RF noise or to install external power source filter to minimize RF noise to an acceptable level.

Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

11.2 Radio Management Compliance
NOTE:
This equipment is not entitled to protection against harmful interference and may not cause interference in duly authorized systems.

Table 11-7 RF Parameter Specifications Wireless Device Working frequency (for North America region)
Working frequency (for CE region)
Modulation

2.4 GHz/5.1 GHz/5.8 GHz Wi-Fi 2412MHz - 2462MHz 5150MHz - 5250MHz 5745MHz - 5825MHz
2412MHz - 2472MHz 5150MHz - 5350 MHz 5470MHz - 5725MHz 5745MHz - 5825MHz DSSS, OFDM

Operator's Manual

11 - 9

Table 11-7 RF Parameter Specifications Wireless Device Transmit power
11.3 Cable Information

11 Guidance and Manufacturer's Declaration
2.4 GHz/5.1 GHz/5.8 GHz Wi-Fi  20 dBm 23dBm 23dBm 33dBm

No. Manufacturer Name
1. MINDRAY Magnetic absorption cable
2. MINDRAY AC power cord

Cable Length (m) Shielding

10

Yes

2.5

No

Remarks /
/

11.4 FCC Warning
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
· Reorient or relocate the receiving antenna. · Increase the separation between the equipment and receiver. · Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected. · Consult the dealer or an experienced radio/TV technician for help.
Note: The Grantee is not responsible for any changes or modifications not expressly approved by the party responsible for compliance. Such modifications could void the user's authority to operate the equipment.
The device has been evaluated to meet general RF exposure requirement. The SAR limit of USA (FCC) is 1.6 W/kg averaged over one gram of tissue. Device types Veterinary Digital Detector and Imaging System with VetiPad M1 and VetiPad M1 Plus (FCC ID: 2A8WB-AP6398S) has also been tested against this SAR limit. The highest reported SAR values for body-worn are 1.404/kg (VetiPad

Operator's Manual

11 - 10

11 Guidance and Manufacturer's Declaration
M1) and 1.191W/kg (VetiPad M1 Plus). This device was tested for typical body-worn operations with the back of the device kept 0mm from the body. The use of accessories that do not satisfy these requirements may not comply with FCC RF exposure requirements, and should be avoided.
Operations in the 5150-5250MHz band are restricted to indoor usage only.
Any emission is maintained within the band of operation under all conditions of normal operation. The max. frequency stability is less than 20ppm.
11.5 IC Warning
This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada's licence-exempt RSS(s). Operation is subject to the following two conditions:
1. This device may not cause interference.
2. This device must accept any interference, including interference that may cause undesired operation of the device.
Le présentappareilestconforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitationestautorisée aux deux conditions suivantes:
1. l'appareil ne doit pas produire de brouillage, et
2. l'utilisateur de l'appareildoit accepter tout brouillageradioélectriquesubi, mêmesi le brouillageest susceptible d'encompromettre le fonctionnement.
This radio transmitter (identify the device by certification number or model number if Category II) has been approved by Industry Canada to operate with the antenna types listed below with the maximum permissible gain indicated. Antenna types not included in this list, having a gain greater than the maximum gain indicated for that type, are strictly prohibited for use with this device.
Le présent émetteur radio (identifier le dispositif par son numéro de certification ou son numéro de modèle s'il fait partie du matériel de catégorie II) a été approuvé par Industrie Canada pour fonctionner avec les types d'antenne énumérés ci-dessous et ayant un gain admissible maximal. Les types d'antenne non inclus dans cette liste, et dont le gain est supérieur au gain maximal indiqué, sont strictement interdits pour l'exploitation de l'émetteur.

Ant. Type
internal permanent antenna

Operation Frequencies (MHz) /Max. Ant. Gain (dBi)

2412~2462

5150~5250

5725~5850

1.93

2.25

1.81

Note: 5.2G band is restricted to indoor use only. La bande de 5.2G est limitée à l'usage d'intérieur seulement.

Operator's Manual

11 - 11

11 Guidance and Manufacturer's Declaration
The device has been evaluated to meet general RF exposure requirement. The SAR limit of Canada (ISED) is 1.6 W/kg averaged over one gram of tissue. Device types Veterinary Digital Detector and Imaging System with VetiPad M1 and VetiPad M1 Plus (IC: 29502-AP6398S) has also been tested against this SAR limit. The highest reported SAR values for body-worn are 1.191W/kg (VetiPad M1, HVIN: AMM3543RP) and 1.404W/kg (VetiPad M1 Plus, HVIN: AMM4343RP). This device was tested for typical body-worn operations with the back of the device kept 0mm from the body. The use of accessories that do not satisfy these requirements may not comply with ISED RF exposure requirements, and should be avoided.
L'appareil a été évalué pour répondre aux exigences générales en matière d'exposition aux RF. La limite du das du Canada (ise) est de 1,6 W/kg en moyenne sur un gramme de tissu. Types d'appareils sans fil FPD Portable avec VetiPad M1 et VetiPad M1 Plus (IC: 29502- AP6398S) a également été testé contre cette limite de das. Les valeurs les Plus élevées du das pour le port corporel sont de 1.404W /kg(VetiPad M1, HVIN: AMM3543RP) et de 1.191W /kg(VetiPad M1 Plus, HVIN: AMM4343RP). Cet appareil a été testé pour des opérations typiques d'usure du corps avec le dos de l'appareil à 0mm du corps. L'utilisation d'accessoires qui ne satisfont pas à ces exigences peut ne pas être conforme aux exigences en matière d'exposition aux RF, et doit être évitée.

Operator's Manual

11 - 12

12 Declaration of Conformity

12 Declaration of Conformity

Hereby, Shenzhen Mindray Animal Medical Technology Co., Ltd. declares that the radio equipment type VetiPad M1/VetiPad M1 Plus series Veterinary Digital Detector and Imaging System is in compliance with Directive 2014/53/EU and S.I. 2017 No. 1206.
The full text of the declaration of conformity is available at the following internet address:

Site Location
https://ims.mindrayanimal.com/pub/ detail.aspx?tid=14&rid=21602

QR Code

Operator's Manual

12 - 1

P/N: 046-00465A-00 (2.0)

