LynxPro
Bluetooth® Data Logger Series

The LynxPro Series of data loggers is a Bluetooth-enabled solution capable of monitoring temperature and humidity for a variety of applications from -100°C to +200°C and 0% RH to 95% RH

Bluetooth® Connectivity

Visual Alarms

Store Over 100,000 Readings

Data Where You Need It

Connect the LynxPro to the MadgeTech App with the click of a button and gain instant access to readings, statistics and more
· Complete Control of Device Functions · Graphical & Tabular Data Representation · Easy Data Export & Sharing · Customizable Notifications & Alarms
Compatible with MadgeTech 4 Software
· Advanced Statistics & Analysis · Automatic MKT Calculations · Data Annotations · Multiple Graph Overlays
ISO 17025 Calibration Certi ied

Be Up & Running In Minutes

Find Quality Products Online at: www.GlobalTestSupply.com

[email protected]

SPECIFICATIONS

General
Start Modes

Immediate start Delay start up to 18 months Pushbutton

Stop Modes

Pushbutton Timed (specific date and time)

Multiple Start/Stop Mode

Start and stop the device multiple times without having to download data or communicate with a PC

Real Time Recording

May be used with MadgeTech App to monitor and record data in real time

Memory Reading Rate Alarm LCD Screen

131,071 total readings 21,845 readings in multiple start/stop mode 1 reading every second Up to 1 reading every 12 hours Programmable high and low limits; alarm is activated when measurement reaches or exceeds set limits 1.7"x.975" (40x22mm)

Calibration

Digital calibration through software

Calibration Date

Automatically recorded within device

Battery Type

User replacable CR2450 coin cell battery

Battery Life

6-months typical at a 15 minute reading rate

Data Format

Date and time stamped °C, °F, or %RH

Time Accuracy

±4.1 minute/month at 25 °C (77 °F)

Computer Interface

USB (interface cable required)

Operating System Compatibility

Software Compatibility MT Link | MT4 Operating Environment -20 °C to +60 °C (-4 °F to +140 °F) 0 %RH
to 95 %RH non-condensing

Dimensions

3.75 in x 2.12 in x 0.70 in (95 mm x 55 mm x 17 mm)

Weight

1.8 oz (51 g)

Material

Polycarbonate

Approvals

US (FCC), CA (IC), CE, South Korea (KCC), China (CMIIT), Taiwan (NCC), Japan (LCIE)

BATTERY WARNING
BATTERY MAY LEAK, FLAME OR EXPLODE IF DISASSEMBLED, SHORTED, CHARGED, CONNECTED TOGETHER, MIXED WITH USED OR OTHER BATTERIES, EXPOSED TO FIRE OR HIGH TEMPERATURE. DISCARD USED BATTERY

Specifications are subject to change without notice. Specific warranty remedy limitations apply.

LynxPro

Measurement Range

-20 °C to +60 °C (-4 °F to +140 °F)

Temperature Resolution 0.1 °C (0.18 °F)

Calibrated Accuracy

±0.5 °C (±0.9 °F) from -5 °C to +55 °C (+23 °F to +131 °F)

LynxPro-RTD

Measurement Range

-100 °C to +200 °C (-148 °F to +392 °F)

Temperature Resolution 0.1 °C (0.18 °F)

Calibrated Accuracy

±0.5 °C (±0.9 °F) from -100 °C to 80 °C ±1.0 °C (±1.8 °F) from 80 °C to 200 °C

LynxPro-RH

Ambient Temperature Measurement Range

-20 °C to +60 °C (-4 °F to +140 °F)

Temperature Resolution 0.1 °C (0.18 °F)

Calibrated Accuracy Ambient Humidity

±0.5 °C (±0.9 °F) from -5 °C to +55 °C (+23 °F to +131 °F)

Measurement Range

0 %RH to +95 %RH

Humidity Resolution Humidity Accuracy

0.1 %RH
±3 % RH over the range 25 % RH to 75 % RH, 20 °C to 40 °C, Hysteresis Error 1 % Typical 3 % Max

Ordering Information
LynxPro 902373-00 Bluetooth Temperature Data Logger
LynxPro-RTD 902375-00 Bluetooth Remote Temperature Data Logger
LynxPro-RH 902374-00 Bluetooth Temperature and Humidity Data Logger

Find Quality Products Online at: www.GlobalTestSupply.com