Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 1 of 35

RF Test Report

Project Number: 5079617 Report Number: 5079617EMC03

Offer Number: SUW-202212003799 Revision Level: 3

Client: Bright Uro Equipment Under Test: Wireless, Catheter-Free, Urodynamic Monitoring System
Product Name: Atusa Male Sensor Model: GUS-M-1000
Referenced Model: Atusa Female Sensor: GUS-F-1000 FCC ID: 2BHMUGUS1000
Applicable Standards: FCC Part 15 Subpart C, § 15.247 RSS-247, Issue 3, August 2023 RSS-GEN, Issue 5, February 2021, Amendment 2 ANSI C63.10:2013
Report issued on: 22 March 2024 Report revision on: 04 March 2025
Test Result: Compliant

FOR THE SCOPE OF ACCREDITATION UNDER CERTIFICATE NUMBER: 3212.01 This report must not be used by the client to claim product certification, approval, or endorsement by A2LA, NIST, or any agency of the Federal
Government.

Prepared by:

Daniel Alvarez, Sr RF/EMC Engineer

Reviewed by:

Martin Taylor, Reviewer

Remarks: This document is issued by the Company subject to its General Conditions of Service printed overleaf, available on request or accessible at http://www.sgs.com//en/Terms-and-Conditions.aspx. And for electronic format documents, subject to Terms and Conditions for Electronic Documents at http://www.sgs.com/en/Terms-and-Conditions/terms-e-document.aspx.
Attention is drawn to the limitation of liability, indemnification and jurisdiction issues defined therein. Any holder of this document is advised that information contained hereon reflects the Company's findings at the time of its intervention only and within the limits of Client's instructions, if any. The Company's sole responsibility is to its Client and this document does not exonerate parties to a transaction from exercising all their rights and obligations under the transaction documents. This document cannot be reproduced except in full, without prior written approval of the Company. Any unauthorized alteration, forgery or falsification of the content or appearance of this document is unlawful, and offenders may be prosecuted to the fullest extent of the law. Unless otherwise stated the results shown in this test report refer only to the sample(s) tested and such sample(s) are retained for a maximum of 30 days only.

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 2 of 35
Table of Contents
1 SUMMARY OF TEST RESULTS ..................................................................................................................................... 3
1.1 MODIFICATIONS REQUIRED FOR COMPLIANCE .............................................................................................................. 3
2 GENERAL INFORMATION............................................................................................................................................. 4
2.1 CLIENT INFORMATION ................................................................................................................................................... 4 2.2 TEST LABORATORY ....................................................................................................................................................... 4 2.3 GENERAL INFORMATION OF EUT .................................................................................................................................. 4 2.4 OPERATING MODES AND CONDITIONS .......................................................................................................................... 4 2.5 EUT CONNECTION BLOCK DIAGRAM ­ CONDUCTED MEASUREMENTS......................................................................... 5 2.6 EUT CONNECTION BLOCK DIAGRAM ­ RADIATED MEASUREMENTS ............................................................................ 5 2.7 SYSTEM CONFIGURATIONS ............................................................................................................................................ 5 2.8 CABLE LIST ................................................................................................................................................................... 5
3 BANDWIDTH ..................................................................................................................................................................... 6
3.1 TEST RESULT................................................................................................................................................................. 6 3.2 TEST METHODS ............................................................................................................................................................. 6 3.3 TEST SITE ...................................................................................................................................................................... 6 3.4 TEST EQUIPMENT .......................................................................................................................................................... 6 3.5 TEST DATA ­ DTS BANDWIDTH (6DB) ......................................................................................................................... 7 3.6 TEST DATA ­ 99% OCCUPIED BANDWIDTH ................................................................................................................... 9
4 OUTPUT POWER ............................................................................................................................................................ 11
4.1 TEST RESULT............................................................................................................................................................... 11 4.2 TEST METHOD ............................................................................................................................................................. 11 4.3 TEST SITE .................................................................................................................................................................... 11 4.4 TEST EQUIPMENT ........................................................................................................................................................ 11 4.5 TEST DATA.................................................................................................................................................................. 12
5 POWER SPECTRAL DENSITY ..................................................................................................................................... 14
5.1 TEST RESULT............................................................................................................................................................... 14 5.2 TEST METHOD ............................................................................................................................................................. 14 5.3 TEST SITE .................................................................................................................................................................... 14 5.4 TEST EQUIPMENT ........................................................................................................................................................ 14 5.5 TEST DATA.................................................................................................................................................................. 14
6 CONDUCTED SPURIOUS EMISSIONS / BAND EDGE............................................................................................. 17
6.1 TEST RESULT............................................................................................................................................................... 17 6.2 TEST METHOD ............................................................................................................................................................. 17 6.3 TEST SITE .................................................................................................................................................................... 17 6.4 TEST EQUIPMENT ........................................................................................................................................................ 17 6.5 TEST DATA ­ DTS BAND EDGE................................................................................................................................... 17 6.6 TEST DATA ­ DTS CONDUCTED SPURIOUS EMISSIONS ............................................................................................... 18
7 EMISSIONS IN RESTRICTED FREQUENCY BANDS .............................................................................................. 21
7.1 TEST RESULT............................................................................................................................................................... 21 7.2 TEST METHOD ............................................................................................................................................................. 21 7.3 TEST SITE .................................................................................................................................................................... 21 7.4 TEST EQUIPMENT ........................................................................................................................................................ 22 7.5 TEST DATA.................................................................................................................................................................. 23
8 INTENTIONAL RADIATOR ANTENNA REQUIREMENT ...................................................................................... 33
8.1 RESULT........................................................................................................................................................................ 33 8.2 REQUIREMENT............................................................................................................................................................. 33

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 3 of 35
8.3 CONCLUSION ............................................................................................................................................................... 33 9 MEASUREMENT UNCERTAINTY............................................................................................................................... 34 10 REVISION HISTORY ...................................................................................................................................................... 35

1 Summary of Test Results

Test Description

Test Specification

Test Result

Bandwidth

15.247(a)(2)

RSS-247 S5.2 (a) RSS-GEN S6.7

Compliant

Output Power

15.247(b)(3)

RSS-247 S5.4 (d)

Compliant

Power Spectral Density

15.247(e)

RSS-247 S5.2 (b)

Compliant

Conducted Spurious Emissions / Band Edge
Emissions in Restricted Frequency Bands
AC Powerline Conducted Emissions
Antenna Requirement

15.247(d) 15.247(d) 15.205, 15.209
15.207
15.203

RSS-247 S5.5 RSS-247 S5.5 RSS-GEN S8.9, S8.10 RSS-GEN S8.8
RSS-GEN S6.8

Compliant Compliant
N/A (2) Compliant (1)

(1) The device uses an internal PCB trace antenna which cannot be replaced by the end user. (2) The device under test is battery powered. Conducted Emissions is not applicable.

Referenced Model: GUS-F-1000 was not tested but was deemed electrically similar to the device that was tested. The difference between the two models is the enclosure as one is used for male patients and the other for female patients.

1.1 Modifications Required for Compliance
None.

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

2 General Information

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 4 of 35

2.1 Client Information
Company Name: Bright Uro, Inc. Address: 3 Goddard
City, State, Zip, Country: Irvine, CA 92618 USA

2.2 Test Laboratory
Name: SGS North America, Inc. Address: 620 Old Peachtree Road NW, Suite 100 City, State, Zip, Country: Suwanee, GA 30024, USA

Accrediting Body: Type of lab:
Certificate Number: CAB ID:

A2LA Testing Laboratory 3212.01 US0186

2.3 General Information of EUT
Equipment Under Test: Wireless, Catheter-Free, Urodynamic Monitoring System Product Name: Atusa Male Sensor Model: GUS-M-1000
Referenced Model: Atusa Female Sensor: GUS-F-1000
Serial Numbers: DV0013 (Conducted Sample) SGS ID#: SUW_SP_20240100251 (Radiated Sample)
Frequency Range: 2402 ­ 2480 MHz Data Modes: Bluetooth Low Energy (BLE1M ­ GFSK) Antenna: Internal PCB Trace (0.5 dBi gain)

Rated Voltage: 3 VDC Test Voltage: 3 VDC

Sample Received Date: 13 November 2023, 15 March 2024 Dates of testing: 14 November 2023 to 15 March 2024

2.4 Operating Modes and Conditions
The EUT was running test mode firmware which allowed the following: 1) selecting low, middle, and high BLE channels. 2) selecting the modulation scheme as used in normal operation, but with 100% duty cycle. 3) Per client statement, the power level was set at 0dBm for testing.

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 5 of 35
2.5 EUT Connection Block Diagram ­ Conducted Measurements

Measurement System

coax 1

A (EUT)

2.6 EUT Connection Block Diagram ­ Radiated Measurements

B (EUT)

Inside Chamber

2.7 System Configurations

Device Reference Manufacturer

Description

A

Bright Uro, Inc.

Atusa Male Sensor (Conducted Sample)

B

Bright Uro, Inc.

Atusa Male Sensor (Radiated Sample)

B

Bright Uro, Inc.

Atusa Male Sensor (Radiated Sample)

B

Bright Uro, Inc.

Atusa Male Sensor (Radiated Sample)

B

Bright Uro, Inc.

Atusa Male Sensor (Radiated Sample)

Model Number
GUS-M1000
GUS-M1000
GUS-M1000
GUS-M1000
GUS-M1000

Serial Number
DV0013 (Conducted Sample)
SGS ID#: SUW_SP_20240100251 (Radiated Sample)
SGS ID#: SUW_SP_20240300967 (Radiated Sample)
SGS ID#: SUW_SP_20240300968 (Radiated Sample)
SGS ID#: SUW_SP_20240300969 (Radiated Sample)

2.8 Cable List

Cable reference

Port Name

1

Conducted RF coax

Start
Antenna Connector

End
Measuring Equipment

Cable Length

Ferrite installed?

Shielded?

<5cm

N

N

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

3 Bandwidth
3.1 Test Result
Test Description
DTS Bandwidth (6dB) Occupied Bandwidth (99%)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 6 of 35

Test Specification

15.247(a)(2)

RSS-247 S5.2 (a)

2.1049

RSS-GEN S6.7

Test Result
Compliant Reported

3.2 Test Methods
The DTS 6dB bandwidth measurements were performed using the procedure from ANSI C63.10 clause 11.8.1, and the 99% occupied bandwidth measurements were performed using the procedure from ANSI C63.10 clause 6.9.3. These procedures are referenced in KDB 558074 D01 15.247 Meas Guidance v05r02.

The BLE 1Mbps data mode was used for these tests.

Limit The minimum 6dB bandwidth shall be at least 500 kHz.

3.3 Test Site
SGS EMC Laboratory, Suwanee, GA

Environmental Conditions Temperature:
Relative Humidity: Atmospheric Pressure:

21.6 °C 22.7 % 95.45 kPa

3.4 Test Equipment

Test End Date:
Equipment
SIGNAL ANALYZER
(TS8997) RF CABLE SMA TO SMA, 0.01-40GHZ TSTPASS SWITCHBOX

23-Dec-2023 Model FSV30
084-0505-059 SB2

Tester: Manufacturer ROHDE & SCHWARZ
TELEDYNE STORM MICROWAVE TSTPASS

DA Asset
B085749
20108 23009

Cal Date 3-Jan-2024
13-Mar-2023 10-Apr-2023

Cal Due Date 3-Jan-2025
13-Mar-2024 10-Apr-2024

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

3.5 Test Data ­ DTS Bandwidth (6dB)

Mode 1M

TX Type
SISO

Frequency (MHz) 2402 2440 2480

ANT
1 1 1

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 7 of 35

6dB Bandwidth (MHz)

Result

Limit

0.724

>=0.5

0.716

>=0.5

0.733

>=0.5

Verdict
Pass Pass Pass

1M_LCH_2402MHz_Ant1_NTNV

1M_MCH_2440MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 8 of 35
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 9 of 35

3.6 Test Data ­ 99% Occupied Bandwidth

Mode 1M

TX Type
SISO

Frequency (MHz) 2402 2440 2480

ANT
1 1 1

99% Occupied Bandwidth (MHz)

Result

Limit

1.058

/

1.059

/

1.056

/

Verdict
Pass Pass Pass

1M_LCH_2402MHz_Ant1_NTNV

1M_MCH_2440MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 10 of 35
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

4 Output Power

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 11 of 35

4.1 Test Result
Test Description Fundamental Emission
Output Power

Test Specification

15.247(b)(3)

RSS-247 S5.4 (d)

Test Result
Compliant

4.2 Test Method
Fundamental maximum peak conducted output power measurements were performed using the method described in ANSI C63.10:2013 clause 11.9.1.1. This procedure is referenced in KDB 558074 D01 15.247 Meas Guidance v05r02.

The BLE 1Mbps data mode was used for this test.

Limit For systems using digital modulation in the 902-928 MHz, 2400-2483.5 MHz, and 5725-5850 MHz bands: 1 Watt. For using antennas with greater than 6dBi of gain, the limit is reduced in dB by the amount the gain exceeds 6dBi (e.g. for a 7.4dBi antenna, the limit is reduced from 30dBm to 28.6dBm). Also, the e.i.r.p. shall not exceed 4 Watts (36 dBm) based on RSS-247 S5.4 (d).

4.3 Test Site
SGS EMC Laboratory, Suwanee, GA

Environmental Conditions Temperature:
Relative Humidity: Atmospheric Pressure:

21.6 °C 22.7 % 95.46 kPa

4.4 Test Equipment

Test End Date:
Equipment SIGNAL
ANALYZER (TS8997) RF CABLE
SMA TO SMA, 0.01-40GHZ TSTPASS SWITCHBOX

23-Dec-2023 Model FSV30
084-0505-059 SB2

Tester: Manufacturer ROHDE & SCHWARZ
TELEDYNE STORM MICROWAVE TSTPASS

DA Asset
B085749
20108 23009

Cal Date 3-Jan-2024
13-Mar-2023 10-Apr-2023

Cal Due Date 3-Jan-2025
13-Mar-2024 10-Apr-2024

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

4.5 Test Data

Mode

TX Type

Frequency (MHz)

2402

1M

SISO

2440

2480

Note1: Antenna Gain: Ant1: 0.5dBi;

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 12 of 35

Maximum Peak Conducted Output Power

(dBm)

ANT1

Limit

-1.47

<=30

-0.69

<=30

-0.39

<=30

Verdict
Pass Pass Pass

1M_LCH_2402MHz_Ant1_NTNV

1M_MCH_2440MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 13 of 35
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

5 Power Spectral Density

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 14 of 35

5.1 Test Result
Test Description Power Spectral Density

Test Specification

15.247(e)

RSS-247 S5.2 (b)

Test Result
Compliant

5.2 Test Method
Peak power spectral density measurements were performed using the procedures from ANSI C63.10: 2013 clause 11.10.2. These procedures are referenced in KDB 558074 D01 15.247 Meas Guidance v05r02.

The BLE 1Mbps data mode was used for this test.

Limit The maximum limit is 8 dBm / 3 kHz.

5.3 Test Site
SGS EMC Laboratory, Suwanee, GA

Environmental Conditions Temperature:
Relative Humidity: Atmospheric Pressure:

21.6 °C 22.7 % 95.46 kPa

5.4 Test Equipment

Test End Date:
Equipment
SIGNAL ANALYZER
(TS8997) RF CABLE SMA TO SMA, 0.01-40GHZ TSTPASS SWITCHBOX

23-Dec-2023 Model FSV30
084-0505-059 SB2

Tester: Manufacturer ROHDE & SCHWARZ
TELEDYNE STORM MICROWAVE TSTPASS

DA Asset
B085749
20108 23009

Cal Date 3-Jan-2024
13-Mar-2023 10-Apr-2023

Cal Due Date 3-Jan-2025
13-Mar-2024 10-Apr-2024

5.5 Test Data

Mode

TX Type

Frequency (MHz)

2402

1M

SISO

2440

2480

Note1: Antenna Gain: Ant1: 0.5dBi;

Maximum PSD (dBm/3kHz)

ANT1

Limit

-15.18

<=8

-13.63

<=8

-13.49

<=8

Verdict
Pass Pass Pass

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 15 of 35
1M_LCH_2402MHz_Ant1_NTNV
1M_MCH_2440MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 16 of 35
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 17 of 35
6 Conducted Spurious Emissions / Band Edge

6.1 Test Result
Test Description Conducted Spurious and
Band Edge Emissions

Test Specification

15.247(d)

RSS-247 S5.5

Test Result
Compliant

6.2 Test Method
Conducted band edge and spurious band emissions in non-restricted frequency bands were measured using the method defined in ANSI C63.10 clause 11.11. This procedure is referenced in KDB 558074 D01 15.247 Meas Guidance v05r02.

The BLE 1Mbps data mode was used for this test.

Limit Because the maximum peak conducted output power was used to determine compliance with the output power limits, the limit in any 100 kHz band outside of the authorized band is 20 dB below the maximum in-band peak level.

6.3 Test Site
SGS EMC Laboratory, Suwanee, GA

Environmental Conditions Temperature:
Relative Humidity: Atmospheric Pressure:

21.6 °C 22.7 % 95.46 kPa

6.4 Test Equipment

Test End Date:
Equipment
SIGNAL ANALYZER
(TS8997) RF CABLE SMA TO SMA, 0.01-40GHZ TSTPASS SWITCHBOX

23-Dec-2023 Model FSV30
084-0505-059 SB2

Tester: Manufacturer ROHDE & SCHWARZ
TELEDYNE STORM MICROWAVE TSTPASS

DA Asset
B085749
20108 23009

Cal Date 3-Jan-2024
13-Mar-2023 10-Apr-2023

Cal Due Date 3-Jan-2025
13-Mar-2024 10-Apr-2024

6.5 Test Data ­ DTS Band Edge

Mode

TX Type

Frequency (MHz)

ANT

Level of Reference (dBm)

Limit (dBm)

Verdict

2402

1

-0.86

-20.86

Pass

1M

SISO

2440

1

-0.86

-20.86

Pass

2480

1

-0.86

-20.86

Pass

Note1: Refer to RSS-247 section 5.5 and ANSI C63.10-2013, the channel contains the maximum

PSD level was used to establish the reference level.

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 18 of 35
6.6 Test Data ­ DTS Conducted Spurious Emissions
1M_LCH_2402MHz_Ant1_NTNV
1M_LCH_2402MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 19 of 35
1M_MCH_2440MHz_Ant1_NTNV
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 20 of 35
1M_HCH_2480MHz_Ant1_NTNV

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 21 of 35
7 Emissions in Restricted Frequency Bands

7.1 Test Result
Test Description Emissions in Restricted
Frequency Bands

Test Specification

15.247(d) 15.205, 15.209

RSS-247 S5.5 RSS-GEN S8.9, S8.10

Test Result
Compliant

7.2 Test Method
Radiated emissions in restricted frequency bands were measured using methods defined in ANSI C63.10 clause 11.12. These procedures are referenced in KDB 558074 D01 15.247 Meas Guidance v05r02.

The BLE 1Mbps data mode was used for this test. Low, middle, and high channels were investigated ­ the device was commanded to continuously transmit on channels 0, 19 and 39.

Test distances for radiated tests: 9k to 30 MHz ­ Near field prescan to determine if there were any emissions 30 to 1000 MHz - The EUT to measurement antenna distance was 3 meters 1 to 18 GHz - The EUT to measurement antenna distance was 3 meters 18 to 26 GHz - The EUT to measurement antenna distance was 3 meters

Limits within restricted bands of operation:

Frequency

Limits(1)
Microvolts/m

dBuV/m

Peak Limits dBuV/m

30 - 88 MHz

100

40 (2)

--

88 - 216 MHz

150

43.5 (2)

--

216 - 960 MHz

200

46 (2)

--

960 - 1000 MHz

500

54 (2)

--

1 - 40 GHz

500

54 (3)

74

(1) These limits are applicable to emissions outside of the intentional transmit frequency band.

(2) Quasi-peak limit

(3) Average limit

7.3 Test Site
3m Absorber Lined Shielded Enclosure (ALSE), Suwanee, GA

Environmental Conditions Temperature:
Relative Humidity: Atmospheric Pressure:

30-1000MHz 21.81 °C 30.3 % 98.44 kPa

1-26GHz 21.98 °C 51.6 % 98.05 kPa

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

7.4 Test Equipment

30-1000MHz

Test End Date:

15-Mar-2024

Equipment

Model

ANTENNA, BILOG
N to N RF Cable

JB6
EM-B810NM276

RF CABLE

SF106

RF CABLE NM TO NM, 0.01-18GHZ
RF CABLE
LOW NOISE AMPLIFIER EMI TEST RECEIVER

90-195-118 104PE ZKL-2+ ESW44

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 22 of 35

Tester: SGM

Manufacturer

Asset Number

SUNOL

B079690

ECHELON

24000

HUBER & SUHNER

B079713

TELEDYNE STORM MICROWAVE
HUBER & SUHNER MINI-CIRCUITS
ROHDE & SCHWARZ

20125 B079793 B079800
22027

Cal Date
19-Apr2022
15-Jan2024
22-Aug2023

Cal Due Date
19-Apr2024
15-Jan2025
22-Aug2024

8-Aug-2023 8-Aug-2024

7-Aug-2023 14-Sep2023
3-Oct-2023

7-Aug-2024 14-Sep2024
3-Oct-2024

Software: TILE 7 software profile "RE 30-1000MHz 3m Tile7 221014" dated 01/31/2024

1-18GHz

Test End Date:

15-Mar-2024

Equipment

Model

ANTENNA, DRG HORN (MEDIUM) N to N RF
Cable N to N RF
Cable LOW NOISE AMPLIFIER
EMI TEST RECEIVER

3117
EM-B810NM276
90-076-020
TS-PR18
ESW44

Tester: SGM

Manufacturer

Asset Number

ETS LINDGREN

B079691

ECHELON TELEDYNE STORM
MICROWAVE ROHDE & SCHWARZ
ROHDE & SCHWARZ

24000 22037 15003 22027

Cal Date
15-Aug2022
15-Jan2024
26-Dec2023
10-Oct2023
3-Oct-2023

Cal Due Date
15-Aug2024
15-Jan2025
26-Dec2024
10-Oct2024
3-Oct-2024

Software: TILE 7 software profile "RE 1-18GHz 3m Tile7 220307" dated 01/31/2024

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 23 of 35
7.5 Test Data
Between 30-1000MHz, there was no significant deviation with respect to axis or channel. Vertical Radiated Spurious Emissions ­ 30-1000MHz (Low Channel)

Amplitude dBuV/m

80.0 70.0 60.0 50.0 40.0 30.0 20.0 10.0
0 30.00M

Radiated Spurious Emissions Vertical

Limit Vertical Peak Emissions Vertical Final QP measurement RB Avg Limit

100.00M

Frequency

1.00G

Horizontal Radiated Spurious Emissions ­ 30-1000MHz (Low Channel)

Limit

Radiated Spurious Emissions

Horizontal Peak Emissions

Horizontal

Horizontal Final QP measurement

80.0

RB Avg Limit

70.0

60.0

50.0

40.0

30.0

20.0

10.0

0 30.00M

100.00M

Frequency

1.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 24 of 35
Vertical Radiated Spurious Emissions ­ 30-1000MHz (Mid Channel)

Limit

Radiated Spurious Emissions

Vertical Peak Emissions

Vertical

Vertical Final QP measurement

80.0

RB Avg Limit

70.0

60.0

50.0

40.0

30.0

20.0

10.0

0 30.00M

100.00M

Frequency

Horizontal Radiated Spurious Emissions ­ 30-1000MHz (Mid Channel)

1.00G

Limit

Radiated Spurious Emissions

Horizontal Peak Emissions

Horizontal

Horizontal Final QP measurement

80.0

RB Avg Limit

70.0

60.0

50.0

40.0

30.0

20.0

10.0

0 30.00M

100.00M

Frequency

1.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 25 of 35
Vertical Radiated Spurious Emissions ­ 30-1000MHz (High Channel)

Limit

Radiated Spurious Emissions

Vertical Peak Emissions

Vertical

Vertical Final QP measurement

80.0

RB Avg Limit

70.0

60.0

50.0

40.0

30.0

20.0

10.0

0 30.00M

100.00M

Frequency

Horizontal Radiated Spurious Emissions ­ 30-1000MHz (High Channel)

1.00G

Limit

Radiated Spurious Emissions

Horizontal Peak Emissions

Horizontal

Horizontal Final QP measurement

80.0

RB Avg Limit

70.0

60.0

50.0

40.0

30.0

20.0

10.0

0 30.00M

100.00M

Frequency

1.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 26 of 35
Vertical Radiated Spurious Emissions ­ 1-3GHz (Low Channel) ­ X-Axis

PAevagkEEmmisisssioionnss

80.0

Radiated Spurious Emissions Vertical

PAAPLLiieevvmmaageiikkrttMa__MLgRReieeamBBasLi__sutiapmurvkerigemt meennt t

70.0

Amplitude dBuV/m

60.0

50.0

40.0

30.0

20.0 1.00G

Frequency

Horizontal Radiated Spurious Emissions ­ 1-3GHz (Low Channel) ­ X-Axis

3.00G

Radiated Spurious Emissions

PAPAeevvaaggkkEMEMmemeaisasisssusiuroieronemnsmesennt t

80.0

Horizontal

APLLiievmmaeiikrtta__LgRRiemBBLi__tiapmvkigt

70.0

60.0

50.0

40.0

30.0

20.0 1.00G

Frequency

3.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 27 of 35
Vertical Radiated Spurious Emissions ­ 1-3GHz (Mid Channel) ­ X-Axis

Peak Emissions

Radiated Spurious Emissions

Avg Emissions Average Limit

Vertical

Peak Limit Limit_RB_avg

90.0

Limit_RB_pk

80.0

70.0

60.0

50.0

40.0

30.0

20.0 1.00G

Frequency

3.00G

Horizontal Radiated Spurious Emissions ­ 1-3GHz (Mid Channel) ­ X-Axis

Peak Emissions

Radiated Spurious Emissions

Avg Emissions Average Limit

Horizontal

Peak Limit Limit_RB_avg

90.0

Limit_RB_pk

80.0

70.0 60.0 50.0 40.0 30.0 20.0
1.00G

Frequency

3.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 28 of 35
Vertical Radiated Spurious Emissions ­ 1-3GHz (High Channel) ­ X-Axis

PAevagkEEmmisisssioionnss

90.0

Radiated Spurious Emissions Vertical

PAAPLLiieevvmmaageiikkrttMa__MLgRReieeamBBasLi__sutiapmurvkerigemt meennt t

80.0

70.0

60.0

50.0

40.0

30.0

20.0 1.00G

Frequency

3.00G

Amplitude dBuV/m

Horizontal Radiated Spurious Emissions ­ 1-3GHz (High Channel) ­ X-Axis

Peak Emissions

80.0

Radiated Spurious Emissions
Horizontal

APAAPLLiieevvvmmaaggeiikkrttEMa__MLgmRReieeamBBisasLi__ssutiapimurovkernigemtsmeennt t

70.0

60.0

50.0

40.0

30.0

20.0 1.00G

Frequency

3.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 29 of 35
Vertical Radiated Spurious Emissions ­ 3-18GHz (Low Channel) ­ X-Axis

80.0

Radiated Spurious Emissions
Vertical

PAPAAPLLiieeevvvmmaaaggeiikkkrttEMa__EMLgmRReimeeamBBisasiLi__sssutiapsimurovkieronigemtnsmesennt t

70.0

60.0

50.0

40.0

30.0

20.0 3.00G

Frequency

10.00G

18.00G

Horizontal Radiated Spurious Emissions ­ 3-18GHz (Low Channel) ­ X-Axis

80.0

Radiated Spurious Emissions
Horizontal

PAPAAPLLiieeevvvmmaaaggeiikkkrttEMa__EMLgmRReimeeamBBisasiLi__sssutiapsimurovkieronigemtnsmesennt t

70.0

60.0

50.0

40.0

30.0

20.0 3.00G

Frequency

10.00G

18.00G

Amplitude dBuV/m

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 30 of 35
Vertical Radiated Spurious Emissions ­ 3-18GHz (Mid Channel) ­ X-Axis

80.0

Radiated Spurious Emissions
Vertical

PAPAAPLLiieeevvvmmaaaggeiikkkrttEMa__EMLgmRReimeeamBBisasiLi__sssutiapsimurovkieronigemtnsmesennt t

70.0

60.0

50.0

40.0

30.0

20.0 3.00G

Frequency

10.00G

18.00G

Vertical Radiated Spurious Emissions ­ 3-18GHz (Mid Channel) ­ Tabular Data Peak

Frequency Raw Pk Polarity Azimuth Height AF Loss Amp DCCF Final Pk

Limit Margin

MHz

dBuV

(V/H) (degrees) (cm) (dB/m) (dB) (dB) (dB) dBuV/m dBuV/m

dB

7320.18

51.3

V

165.0 192.0 35.5 3.2 42.0 0.9

48.9

74.0

-25.1

16194.14 48.0

V

37.0

118.0 40.8 4.9 40.5 0.9

54.1

74.0

-19.9

17754.06 49.1

V

301.0 138.0 40.8 5.2 40.9 0.9

55.1

74.0

-18.9

17920.50 48.7

V

43.0

177.0 40.9 5.2 40.8 0.9

54.9

74.0

-19.2

Final Pk = Raw Pk + AF + Loss - Amp Margin = Final Pk - Limit

Frequency MHz
7320.18 16194.14 17754.06 17920.50

Raw Avg dBuV 43.9 35.4 37.0 36.7

Polarity (V/H)
V V V V

Azimuth (degrees)
165.0 37.0 301.0 43.0

Average
Height AF Loss (cm) (dB/m) (dB) 192.0 35.5 3.2 118.0 40.8 4.9 138.0 40.8 5.2 177.0 40.9 5.2

Amp (dB) 42.0 40.5 40.9 40.8

DCCF (dB) 0.9 0.9 0.9 0.9

Final Avg dBuV/m
41.5 41.5 43.0 42.9

Limit (dBuV/m)
54.0 54.0 54.0 54.0

Margin (dB) -12.5 -12.5 -11.0 -11.1

Final Avg = Raw Avg + AF + Loss - Amp Margin = Final Avg - Limit

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Amplitude dBuV/m

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 31 of 35
Horizontal Radiated Spurious Emissions ­ 3-18GHz (Mid Channel) ­ X-Axis

80.0

Radiated Spurious Emissions
Horizontal

PAPeevaagkkEEMmmeisaissssiuoironensmsent AAPLLiievvmmageiikrttMa__LgRReieamBBsLi__utiapmrvkeigmt ent

70.0

60.0

50.0

40.0

30.0

20.0 3.00G

Frequency

10.00G

18.00G

Horizontal Radiated Spurious Emissions ­ 3-18GHz (Mid Channel) ­ Tabular Data Peak

Frequency Raw Pk Polarity Azimuth Height AF Loss Amp DCCF

Final Pk Limit Margin

MHz

dBuV

(V/H) (degrees) (cm) (dB/m) (dB) (dB)

(dB)

dBuV/m dBuV/m

dB

7319.94

55.9

H

243.0 102.0 35.5 3.2 41.9

0.9

53.6

74.0

-20.4

17865.98 49.0

H

109.0 100.0 41.0 5.2 40.8

0.9

55.3

74.0

-18.7

Final Pk = Raw Pk + AF + Loss - Amp Margin = Final Pk - Limit

Frequency Raw Avg MHz dBuV
7319.94 52.7 17865.98 37.1

Polarity (V/H)
H H

Azimuth Height (degrees) (cm)
243.0 102.0 109.0 100.0

Average

AF Loss Amp

(dB/m) (dB)

(dB)

35.5 3.2

41.9

41.0 5.2

40.8

DCCF (dB) 0.9 0.9

Avg Value Limit dBuV/m (dBuV/m)

50.4

54.0

43.4

54.0

Margin (dB) -3.6 -10.6

Final Avg = Raw Avg + AF + Loss - Amp Margin = Final Avg - Limit

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 32 of 35
Vertical Radiated Spurious Emissions ­ 3-18GHz (High Channel) ­ X-Axis

80.0

Radiated Spurious Emissions
Vertical

PAPAAPLLiieeevvvmmaaaggeiikkkrttEMa__EMLgmRReimeeamBBisasiLi__sssutiapsimurovkieronigemtnsmesennt t

70.0

Amplitude dBuV/m

60.0

50.0

40.0

30.0

20.0 3.00G
80.0

Frequency

10.00G

Horizontal Radiated Spurious Emissions 3-18GHz (High Channel) ­ X-Axis

18.00G

Radiated Spurious Emissions
Horizontal

PAPeevaagkkEEMmmeisaissssiuoironensmsent AAPLLiievvmmageiikrttMa__LgRReieamBBsLi__utiapmrvkeigmt ent

70.0

Amplitude dBuV/m

60.0

50.0

40.0

30.0

20.0 3.00G

Frequency

10.00G

18.00G

Note: 3-18 GHz Emissions were tested with programmed samples that did not have programming harness attached. (SGS ID#: SUW_SP_20240300967,968,969)

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 33 of 35
8 Intentional Radiator Antenna Requirement

8.1 Result
Test Description Antenna Requirement

Test Specification 47 CFR Part 15, Subpart C 15.203

Test Result
Compliant

8.2 Requirement
An intentional radiator shall be designed to ensure that no antenna other than that furnished by the responsible party shall be used with the device. The use of an antenna that uses a unique coupling to the intentional radiator, the manufacturer may design the unit permanently attached antenna or of so that a broken antenna can be replaced by the user, but the use of a standard antenna jack or electrical connector is prohibited.

8.3 Conclusion
The Antenna on this device is PCB soldered to the circuit board. This meets the FCC antenna requirement.

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 34 of 35
9 Measurement Uncertainty
The measurement uncertainty figures are calculated in accordance with TR 100 028-1 [2] and correspond to an expansion factor (coverage factor) k = 2 (which provides confidence levels of 95.45 % in the case where the distributions characterizing the actual measurement uncertainties are normal (Gaussian)).

Parameter
Radio Frequency total RF power, conducted RF power density, conducted spurious emissions, conducted all emissions, radiated temperature humidity DC and low frequency voltages Conducted disturbance at mains port using AMN

Expanded Uncertainty for Normal k factor equal to 2

Required

Laboratory Actual

±1 x 10-5

±9.8 x 10-8

±1.5 dB

±1.2 dB

±3 dB

±0.7 dB

±3 dB

±2.1 dB

±6 dB

±4.8 dB

±1°C

±0.5°C

±5 %

±3.5%

±3 %

±0.4%

± 3.4 dB

± 2.5 dB

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)

Test Report Number: 5079617EMC03 Rev: 3 Bright Uro / GUS-M-1000; GUS-F-1000 Page: 35 of 35

10 Revision History

Revision Level Draft 0 1
2
3

Description of changes
Draft release Initial Release Updated Client name. Added FCC ID to report cover page. Updated RSS-247 standard to Issue 3. Corrected FCC ID on cover page.

Revision Date
19 March 2024 22 March 2024 22 April 2024
04 March 2025
08 April 2025

SGS North America Inc. Connectivity & Products

620 Old Peachtree Road NW, Suite 100, Suwanee, GA 30024

t (770) 570-1800

www.sgs.com

Member of the SGS Group (SGS SA)



Related FCC IDs: