Date/Time: 1/13/2011 12:26:38 AM, Date/Time: 1/13/2011 12:35:24 AM

John

14282070 App C SAR Highest Test Plots

Apple Inc. E8147A SMARTPHONE BCG-E8147A BCGE8147A e8147a

Apple Inc. E8146A SMARTPHONE BCG-E8146A BCGE8146A e8146a

Apple Inc. E8148A SMARTPHONE BCG-E8148A BCGE8148A e8148a

PDF Viewing Options

Not Your Device? Search For Manuals or Datasheets below:


File Info : application/pdf, 17 Pages, 756.20KB

Document DEVICE REPORTGetApplicationAttachment.html?id=6105383
Test Laboratory: UL Verification Services Inc. SAR Lab C

Date/Time: 7/25/2022 12:34:44 AM

MSS ANT 4

Frequency: 1625.1 MHz; Duty Cycle: 1:2.61096; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 1625.1 MHz;  = 1.354 S/m; r = 41.041;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1545; Calibrated: 2/23/2022 - Probe: EX3DV4 - SN3686; ConvF(7.68, 7.68, 7.68) @ 1625.1 MHz; Calibrated: 1/18/2022 - Sensor-Surface: 2.5mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 AA; Serial: 1948
Edge 2/1-PRB SC-FDMA_Zone 0_ Ch 262466/Area Scan (7x15x1): Measurement grid: dx=15mm,
dy=15mm Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.02 W/kg
Edge 2/1-PRB SC-FDMA_Zone 0_ Ch 262466/Zoom Scan (5x5x7)/Cube 0: Measurement grid:
dx=8mm, dy=8mm, dz=5mm Reference Value = 24.03 V/m; Power Drift = -0.09 dB Peak SAR (extrapolated) = 1.68 W/kg SAR(1 g) = 0.815 W/kg; SAR(10 g) = 0.382 W/kg Smallest distance from peaks to all points 3 dB below = 8 mm Ratio of SAR at M2 to SAR at M1 = 54.1% Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.05 W/kg

0 dB = 1.05 W/kg = 0.21 dBW/kg

Plot No. 1

Test Laboratory: UL Verification Services Inc. SAR Lab 1

Date/Time: 7/26/2022 3:28:35 PM

n41 ANT 2

Frequency: 2593 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2593 MHz;  = 1.991 S/m; r = 38.145;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1359; Calibrated: 1/7/2022 - Probe: EX3DV4 - SN3991; ConvF(7.9, 7.9, 7.9) @ 2593 MHz; Calibrated: 8/20/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Tilt_pi/2 BPSK RB 135,67 Ch 518598/Area Scan (10x17x1): Measurement grid: dx=12mm,
dy=12mm Maximum value of SAR (measured) = 1.37 W/kg
LHS/Tilt_pi/2 BPSK RB 135,67 Ch 518598/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm Reference Value = 25.84 V/m; Power Drift = -0.09 dB Peak SAR (extrapolated) = 1.93 W/kg SAR(1 g) = 0.756 W/kg; SAR(10 g) = 0.276 W/kg Smallest distance from peaks to all points 3 dB below = 6 mm Ratio of SAR at M2 to SAR at M1 = 41.7% Maximum value of SAR (measured) = 1.45 W/kg

0 dB = 1.45 W/kg = 1.61 dBW/kg

Plot No. 2

Test Laboratory: UL Verification Services Inc. SAR Lab 4

Date/Time: 7/26/2022 6:57:14 PM

LTE Band 48 ANT 4

Frequency: 3690 MHz; Duty Cycle: 1:1.59956; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 3690 MHz;  = 2.958 S/m; r = 39.419;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1433; Calibrated: 2/23/2022 - Probe: EX3DV4 - SN3929; ConvF(6.4, 6.4, 6.4) @ 3690 MHz; Calibrated: 3/23/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection), Sensor-Surface: 4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Touch_QPSK RB 50,24 Ch 56640/Area Scan (10x18x1): Measurement grid: dx=12mm,
dy=12mm Maximum value of SAR (measured) = 1.31 W/kg
LHS/Touch_QPSK RB 50,24 Ch 56640/Zoom Scan (7x7x8)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=4mm Reference Value = 15.87 V/m; Power Drift = -0.16 dB Peak SAR (extrapolated) = 2.27 W/kg SAR(1 g) = 0.689 W/kg; SAR(10 g) = 0.238 W/kg Smallest distance from peaks to all points 3 dB below = 6.1 mm Ratio of SAR at M2 to SAR at M1 = 41.1% Maximum value of SAR (measured) = 0.770 W/kg

0 dB = 0.770 W/kg = -1.14 dBW/kg

Plot No. 3

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/22/2022 11:04:53 PM

Wi-Fi 2.4GHz ANT 3 CELL OFF

Frequency: 2437 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2437 MHz;  = 1.836 S/m; r = 41.101;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2437 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 4/802.11b_ch 6/Area Scan (7x17x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 0.930 W/kg
Edge 4/802.11b_ch 6/Zoom Scan (7x9x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 21.18 V/m; Power Drift = -0.14 dB Peak SAR (extrapolated) = 1.74 W/kg SAR(1 g) = 0.778 W/kg; SAR(10 g) = 0.346 W/kg Smallest distance from peaks to all points 3 dB below = 8 mm Ratio of SAR at M2 to SAR at M1 = 45% Maximum value of SAR (measured) = 1.35 W/kg

0 dB = 1.35 W/kg = 1.30 dBW/kg

Plot No. 4

Test Laboratory: UL Verification Services Inc. SAR Lab G

Date/Time: 7/25/2022 5:19:03 PM

Wi-Fi 5.3GHz ANT 5 CELL OFF

Frequency: 5270 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 5270 MHz;  = 4.572 S/m; r = 34.8;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1472; Calibrated: 1/7/2022 - Probe: EX3DV4 - SN7585; ConvF(5.3, 5.3, 5.3) @ 5270 MHz; Calibrated: 4/27/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Rear/802.11n_HT40_Ch 54/Area Scan (11x20x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 1.50 W/kg
Rear/802.11n_HT40_Ch 54/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=2mm Reference Value = 16.35 V/m; Power Drift = -0.19 dB Peak SAR (extrapolated) = 2.96 W/kg SAR(1 g) = 0.717 W/kg; SAR(10 g) = 0.197 W/kg Smallest distance from peaks to all points 3 dB below = 5.8 mm Ratio of SAR at M2 to SAR at M1 = 54.6% Maximum value of SAR (measured) = 1.76 W/kg

0 dB = 1.76 W/kg = 2.46 dBW/kg

Plot No. 5

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/29/2022 11:50:55 AM

Bluetooth (Pstandalone) ANT 3

Frequency: 2441 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2441 MHz;  = 1.864 S/m; r = 38.33;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2441 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 2/GFSK DH5_ch 39/Area Scan (6x18x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.41 W/kg
Edge 2/GFSK DH5_ch 39/Zoom Scan (8x9x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 23.32 V/m; Power Drift = -0.03 dB Peak SAR (extrapolated) = 1.61 W/kg SAR(1 g) = 0.638 W/kg; SAR(10 g) = 0.257 W/kg Smallest distance from peaks to all points 3 dB below = 4.5 mm Ratio of SAR at M2 to SAR at M1 = 47.7% Maximum value of SAR (measured) = 1.19 W/kg

0 dB = 1.19 W/kg = 0.76 dBW/kg

Plot No. 6

Test Laboratory: UL Verification Services Inc. SAR Lab 12

Date/Time: 7/22/2022 8:31:59 AM

MSS ANT 4

Frequency: 1625.1 MHz; Duty Cycle: 1:2.61096; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 1625.1 MHz;  = 1.285 S/m; r = 39.939;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1380; Calibrated: 8/11/2021 - Probe: EX3DV4 - SN3686; ConvF(7.68, 7.68, 7.68); Calibrated: 1/18/2022; - Sensor-Surface: 2.5mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 2/1-PRB SC-FDMA_Zone 0_ Ch 262466/Area Scan (7x15x1): Measurement grid: dx=15mm,
dy=15mm Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 0.713 W/kg
Edge 2/1-PRB SC-FDMA_Zone 0_ Ch 262466/Zoom Scan (5x5x7)/Cube 0: Measurement grid:
dx=8mm, dy=8mm, dz=5mm Reference Value = 23.61 V/m; Power Drift = 0.19 dB Peak SAR (extrapolated) = 1.55 W/kg SAR(1 g) = 0.804 W/kg; SAR(10 g) = 0.397 W/kg Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.12 W/kg

0 dB = 1.12 W/kg = 0.49 dBW/kg

Plot No. 7

Test Laboratory: UL Verification Services Inc. SAR Lab 1

Date/Time: 7/26/2022 10:45:08 AM

n41 ANT 2

Frequency: 2593 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2593 MHz;  = 1.991 S/m; r = 38.145;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1359; Calibrated: 1/7/2022 - Probe: EX3DV4 - SN3991; ConvF(7.9, 7.9, 7.9) @ 2593 MHz; Calibrated: 8/20/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Tilt_pi/2 BPSK RB 135,67 Ch 518598/Area Scan (10x17x1): Measurement grid: dx=12mm,
dy=12mm Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.42 W/kg
LHS/Tilt_pi/2 BPSK RB 135,67 Ch 518598/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm Reference Value = 25.76 V/m; Power Drift = 0.08 dB Peak SAR (extrapolated) = 1.88 W/kg SAR(1 g) = 0.745 W/kg; SAR(10 g) = 0.276 W/kg Smallest distance from peaks to all points 3 dB below = 6 mm Ratio of SAR at M2 to SAR at M1 = 41% Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.44 W/kg

0 dB = 1.44 W/kg = 1.58 dBW/kg

Plot No. 8

Test Laboratory: UL Verification Services Inc. SAR Lab 4

Date/Time: 7/27/2022 12:17:38 AM

LTE Band 48 ANT 4

Frequency: 3690 MHz; Duty Cycle: 1:1.59956; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 3690 MHz;  = 2.958 S/m; r = 39.419;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1433; Calibrated: 2/23/2022 - Probe: EX3DV4 - SN3929; ConvF(6.4, 6.4, 6.4) @ 3690 MHz; Calibrated: 3/23/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection), Sensor-Surface: 4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Touch_QPSK RB 50,24 Ch 56640/Area Scan (10x16x1): Measurement grid: dx=12mm,
dy=12mm Maximum value of SAR (measured) = 1.31 W/kg
LHS/Touch_QPSK RB 50,24 Ch 56640/Zoom Scan (7x7x8)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=4mm Reference Value = 14.56 V/m; Power Drift = -0.17 dB Peak SAR (extrapolated) = 1.82 W/kg SAR(1 g) = 0.555 W/kg; SAR(10 g) = 0.192 W/kg Smallest distance from peaks to all points 3 dB below = 6 mm Ratio of SAR at M2 to SAR at M1 = 42.3% Maximum value of SAR (measured) = 0.619 W/kg

0 dB = 0.619 W/kg = -2.08 dBW/kg

Plot No. 9

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/28/2022 8:02:29 PM

Wi-Fi 2.4GHz ANT 3 CELL OFF

Frequency: 2437 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2437 MHz;  = 1.865 S/m; r = 38.449;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2437 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 4/802.11b_ch 6/Area Scan (6x17x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.04 W/kg
Edge 4/802.11b_ch 6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 26.78 V/m; Power Drift = 0.10 dB Peak SAR (extrapolated) = 1.80 W/kg SAR(1 g) = 0.748 W/kg; SAR(10 g) = 0.285 W/kg Smallest distance from peaks to all points 3 dB below = 3 mm Ratio of SAR at M2 to SAR at M1 = 45% Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.46 W/kg

0 dB = 1.46 W/kg = 1.64 dBW/kg

Plot No. 10

Test Laboratory: UL Verification Services Inc. SAR Lab F

Date/Time: 7/28/2022 9:11:18 PM

Wi-Fi 5.3GHz ANT 5 CELL OFF

Frequency: 5270 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 5270 MHz;  = 4.712 S/m; r = 36.11;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1257; Calibrated: 9/15/2021 - Probe: EX3DV4 - SN3749; ConvF(4.66, 4.66, 4.66) @ 5270 MHz; Calibrated: 1/18/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD; Serial: 1740
Rear/802.11n_HT40_Ch 54/Area Scan (11x20x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 1.70 W/kg
Rear/802.11n_HT40_Ch 54/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=2mm Reference Value = 20.74 V/m; Power Drift = -0.10 dB Peak SAR (extrapolated) = 3.16 W/kg SAR(1 g) = 0.795 W/kg; SAR(10 g) = 0.227 W/kg Smallest distance from peaks to all points 3 dB below = 6.6 mm Ratio of SAR at M2 to SAR at M1 = 55.9% Maximum value of SAR (measured) = 2.00 W/kg

0 dB = 2.00 W/kg = 3.01 dBW/kg

Plot No. 11

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/29/2022 9:17:38 AM

Bluetooth (Pstandalone) ANT 3

Frequency: 2441 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2441 MHz;  = 1.867 S/m; r = 38.446;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2441 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 2.5mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2.5mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 4/GFSK DH5_ch 39/Area Scan (8x18x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 0.763 W/kg
Edge 4/GFSK DH5_ch 39/Zoom Scan (7x9x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 16.41 V/m; Power Drift = -0.19 dB Peak SAR (extrapolated) = 3.47 W/kg SAR(1 g) = 0.745 W/kg; SAR(10 g) = 0.295 W/kg Smallest distance from peaks to all points 3 dB below = 5 mm Ratio of SAR at M2 to SAR at M1 = 9.3%
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.25 W/kg

0 dB = 1.25 W/kg = 0.97 dBW/kg

Plot No. 12

Test Laboratory: UL Verification Services Inc. SAR Lab 1

Date/Time: 7/26/2022 10:03:32 AM

n41 ANT 2

Frequency: 2593 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2593 MHz;  = 1.991 S/m; r = 38.145;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1359; Calibrated: 1/7/2022 - Probe: EX3DV4 - SN3991; ConvF(7.9, 7.9, 7.9) @ 2593 MHz; Calibrated: 8/20/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Tilt_Pi/2 BPSK RB 135,67 Ch 518598/Area Scan (10x17x1): Measurement grid: dx=12mm,
dy=12mm
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.30 W/kg
LHS/Tilt_Pi/2 BPSK RB 135,67 Ch 518598/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm Reference Value = 25.11 V/m; Power Drift = 0.03 dB Peak SAR (extrapolated) = 2.10 W/kg SAR(1 g) = 0.830 W/kg; SAR(10 g) = 0.304 W/kg Smallest distance from peaks to all points 3 dB below = 6 mm Ratio of SAR at M2 to SAR at M1 = 41%
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.60 W/kg

0 dB = 1.60 W/kg = 2.04 dBW/kg

Plot No. 13

Test Laboratory: UL Verification Services Inc. SAR Lab 4

Date/Time: 7/26/2022 5:42:03 PM

LTE Band 48 ANT 4

Frequency: 3690 MHz; Duty Cycle: 1:1.59956; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 3690 MHz;  = 2.958 S/m; r = 39.419;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1433; Calibrated: 2/23/2022 - Probe: EX3DV4 - SN3929; ConvF(6.4, 6.4, 6.4) @ 3690 MHz; Calibrated: 3/23/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection), Sensor-Surface: 4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V8.0 (20deg probe tilt); Type: QD 000 P41 Ax; Serial: xxxx
LHS/Touch_QPSK RB 50,24 Ch 56640/Area Scan (10x18x1): Measurement grid: dx=12mm,
dy=12mm Maximum value of SAR (measured) = 1.39 W/kg
LHS/Touch_QPSK RB 50,24 Ch 56640/Zoom Scan (7x7x8)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=4mm Reference Value = 16.20 V/m; Power Drift = -0.04 dB Peak SAR (extrapolated) = 2.35 W/kg SAR(1 g) = 0.684 W/kg; SAR(10 g) = 0.236 W/kg Smallest distance from peaks to all points 3 dB below = 5.9 mm Ratio of SAR at M2 to SAR at M1 = 39.3% Maximum value of SAR (measured) = 0.780 W/kg

0 dB = 0.780 W/kg = -1.08 dBW/kg

Plot No. 14

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/20/2022 4:53:11 PM

Wi-Fi 2.4GHz ANT 3 CELL OFF

Frequency: 2437 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2437 MHz;  = 1.863 S/m; r = 37.956;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2437 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 4/802.11b_ch 6/Area Scan (7x17x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) = 1.01 W/kg
Edge 4/802.11b_ch 6/Zoom Scan (7x9x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 21.91 V/m; Power Drift = -0.05 dB Peak SAR (extrapolated) = 2.06 W/kg SAR(1 g) = 0.902 W/kg; SAR(10 g) = 0.395 W/kg Smallest distance from peaks to all points 3 dB below = 8 mm Ratio of SAR at M2 to SAR at M1 = 44% Maximum value of SAR (measured) = 1.60 W/kg

0 dB = 1.60 W/kg = 2.04 dBW/kg

Plot No. 15

Test Laboratory: UL Verification Services Inc. SAR Lab F

Date/Time: 7/27/2022 10:20:19 PM

Wi-Fi 5.3GHz ANT 5 CELL OFF

Frequency: 5270 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used: f = 5270 MHz;  = 4.712 S/m; r = 36.11;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1257; Calibrated: 9/15/2021 - Probe: EX3DV4 - SN3749; ConvF(4.66, 4.66, 4.66) @ 5270 MHz; Calibrated: 1/18/2022 - Sensor-Surface: 1.4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD; Serial: 1740
Rear/802.11n_HT40_Ch 54/Area Scan (11x20x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 1.79 W/kg
Rear/802.11n_HT40_Ch 54/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=2mm Reference Value = 21.57 V/m; Power Drift = -0.13 dB Peak SAR (extrapolated) = 2.70 W/kg SAR(1 g) = 0.742 W/kg; SAR(10 g) = 0.209 W/kg Smallest distance from peaks to all points 3 dB below = 4.7 mm Ratio of SAR at M2 to SAR at M1 = 56.6% Maximum value of SAR (measured) = 1.71 W/kg

0 dB = 1.71 W/kg = 2.33 dBW/kg

Plot No. 16

Test Laboratory: UL Verification Services Inc. SAR Lab E

Date/Time: 7/28/2022 2:26:23 PM

Bluetooth(Pstandalone) ANT 4 CELL OFF

Frequency: 2441 MHz; Duty Cycle: 1:1; Room Ambient Temperature: 24.0°C; Liquid Temperature: 23.0°C Medium parameters used (interpolated): f = 2441 MHz;  = 1.867 S/m; r = 38.446;  = 1000 kg/m3 Dasy Configuration: - Area Scan Setting: Find Secondary Maximum within 2.0 dB and with a peak SAR value greater than 0.0012 W/kg - Electronics: DAE4 Sn1377; Calibrated: 9/20/2021 - Probe: EX3DV4 - SN3885; ConvF(7.39, 7.39, 7.39) @ 2441 MHz; Calibrated: 9/23/2021 - Sensor-Surface: 1.4mm (Mechanical Surface Detection) - Phantom: Twin-SAM V5.0 (20deg probe tilt); Type: QD 000 P40 CD; Serial: xxxx
Edge 2/GFSK DH5_ch 39/Area Scan (7x17x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.06 W/kg
Edge 2/GFSK DH5_ch 39/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 21.68 V/m; Power Drift = -0.17 dB Peak SAR (extrapolated) = 1.55 W/kg SAR(1 g) = 0.654 W/kg; SAR(10 g) = 0.237 W/kg Smallest distance from peaks to all points 3 dB below = 2.2 mm Ratio of SAR at M2 to SAR at M1 = 45.6%
Info: Interpolated medium parameters used for SAR evaluation. Maximum value of SAR (measured) = 1.16 W/kg

0 dB = 1.16 W/kg = 0.64 dBW/kg