Measurement Report

Kent

Test reports

Kunwei Medical Electronics Co., Ltd KD-17B Behind-The-Ear Hearing Aid 2A9S9-KD-17B 2A9S9KD17B kd 17b

PDF Viewing Options

Not Your Device? Search For Manuals or Datasheets below:


File Info : application/pdf, 47 Pages, 3.83MB

Document DEVICE REPORTGetApplicationAttachment.html?id=6307263
FCC Test Report
Report No: FCS202211204W01 Issued for

Applicant: Address: Product Name:

Kunwei Medical Electronics Co., Ltd
Room 1402,Block B,Builiding 2,Longgang Tianan Cyber Park,Huangge Road,Longgang District Shenzhen Guangdong,CN518172
Behind-The-Ear Hearing Aid

Brand Name:

N/A

Model Name:

KD-17B

Series Model:

KD-X78,KG-7XB,KV-70X

FCC ID:

2A9S9-KD-17B

Issued By: Flux Compliance Service Laboratory
Add: Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan
Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 2 of 47

Report No.: FCS202211204W01

TEST RESULT CERTIFICATION

Applicant's Name...................... :
Address...................................... :
Manufacture's Name................ :
Address...................................... :
Product Description Product Name........................... : Brand Name...............................: Model Name.............................. :

Kunwei Medical Electronics Co., Ltd
Room 1402,Block B,Builiding 2,Longgang Tianan Cyber Park, Huangge Road,Longgang District Shenzhen Guangdong, CN518172 Kunwei Medical Electronics Co., Ltd Room 1402,Block B,Builiding 2,Longgang Tianan Cyber Park, Huangge Road,Longgang District Shenzhen Guangdong, CN518172
Behind-The-Ear Hearing Aid N/A KD-17B

Series Model..............................: KD-X78,KG-7XB,KV-70X

Test Standards.......................... : FCC Rules and Regulations Part 15 Subpart C, Section 247
Test Procedure.......................... : ANSI C63.10:2013
This device described above has been tested by Flux Compliance Service Laboratory, the test results show that the equipment under test (EUT) is in compliance with the FCC requirements. And it is applicable only to the tested sample identified in the report. This report shall not be reproduced except in full, without the written approval of Flux Compliance Service Laboratory, this document may be altered or revised by Flux Compliance Service Laboratory, personal only, and shall be noted in the revision of the document.. Date of Test.................................. :
Date (s) of performance of tests. : Nov 29, 2022~Dec 21, 2022
Date of Issue.................................: Dec 21, 2022
Test Result.....................................: Pass

Tested by

:

Reviewed by

:

Approved by

:

(Scott Shen) (Duke Qian) (Jack Wang)

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 3 of 47

Report No.: FCS202211204W01

Table of Contents

Page

1. SUMMARY OF TEST RESULTS .............................................................................................. 6 1.1 TEST FACTORY ................................................................................................................ 7 1.2 MEASUREMENT UNCERTAINTY .....................................................................................7
2. GENERAL INFORMATION .......................................................................................................8 2.1 GENERAL DESCRIPTION OF THE EUT ..........................................................................8 2.2 BLOCK DIGRAM SHOWING THE CONFIGURATION OF SYSTEM TESTED ............. 10 2.3 DESCRIPTION OF NECESSARY ACCESSORIES AND SUPPORT UNITS .................11 2.4 EQUIPMENTS LIST .........................................................................................................12
3 CONDUCTED EMISSION MEASUREMENT .......................................................................... 13
4. 20 DB BANDWIDTH ............................................................................................................... 17 4.1 Limit .................................................................................................................................. 17 4.2 Test Procedure ................................................................................................................. 17 4.3 Test setup ......................................................................................................................... 17 4.4 Test results ....................................................................................................................... 17 4.5 Original Test Data .............................................................................................................18
5. CONDUCTED OUTPUT POWER ...........................................................................................20 5.1 LIMIT ................................................................................................................................ 20 5.2 TEST PROCEDURE ........................................................................................................ 20 5.3 TEST SETUP ................................................................................................................... 20 5.4 TEST RESULTS ...............................................................................................................20
6 NUMBER OF HOPPING CHANNEL ....................................................................................... 23 6.1 LIMIT ................................................................................................................................ 23 6.2 TEST PROCEDURE ........................................................................................................ 23 5.3 TEST SETUP ................................................................................................................... 23 6.4 EUT OPERATION CONDITIONS .................................................................................... 23 6.5 TEST RESULTS ...............................................................................................................24
7. BAND EDGE AND SPURIOUS(CONDUCTED) .................................................................... 25 7.1 LIMIT ................................................................................................................................ 25 7.2 TEST PROCEDURE ........................................................................................................ 25 7.3 TEST SETUP ................................................................................................................... 25 7.4 TEST RESULTS ...............................................................................................................26

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 4 of 47

Report No.: FCS202211204W01

Table of Contents

Page

7.5 Original test data .............................................................................................................. 26 7.6 For Hopping Band edge ...................................................................................................29
8. RADIATED EMISSION MEASUREMENT .............................................................................. 30 8.1 RADIATED EMISSION LIMITS ........................................................................................30 8.2 TEST PROCEDURE ........................................................................................................ 31 8.3 TESTSETUP .................................................................................................................... 32 8.4. TEST RESULTS ..............................................................................................................34 8.5 (30MHZ-1000MHZ) .......................................................................................................... 35
9. AVERAGE TIME OF OCCUPANCY ....................................................................................... 41 9.1 LIMIT ................................................................................................................................ 41 9.2 TEST PROCEDURE ........................................................................................................ 41 9.3 TEST SETUP ................................................................................................................... 41 9.4 TEST RESULTS ...............................................................................................................42 9.5 ORIGINAL TEST DATA ....................................................................................................42
10. HOPPING CHANNEL SEPARATION MEASUREMEN ....................................................... 44 10.1 LIMIT .............................................................................................................................. 44 10.2 TEST PROCEDURE ...................................................................................................... 44 10.3 TEST SETUP ................................................................................................................. 44 10.4 EUT OPERATION CONDITIONS .................................................................................. 44 10.5 TEST RESULTS .............................................................................................................45
11. ANTENNA REQUIREMENT ..................................................................................................47 11.1 STANDARD REQUIREMENT .........................................................................................47 11.2 RESULT.......................................................................................................................... 47

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 5 of 47

Report No.: FCS202211204W01

Revision History

Rev.

Issue Date

Effect Page

Contents

00

Dec 21, 2022

N/A

N/A

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 6 of 47

Report No.: FCS202211204W01

1. SUMMARY OF TEST RESULTS
Test procedures according to the technical standards: KDB 558074 D01 15.247 Meas Guidance v05r02

FCC Part 15.247,Subpart C

Standard Section
15.207

Test Item Conducted Emission

15.247(a)(1)

Hopping Channel Separation

15.247 (b)(1)

Output Power

15.209 15.247(d) 15.247(a)(1 )(i)

Radiated Spurious Emission
Conducted Spurious & Band Edge Emission
Number of Hopping Frequency

15.247(a)(1)(i)

Dwell Time

15.247(a)(1)

20dB Bandwidth 99% Bandwidth

15.205

Restricted bands of operation

Part 15.247(d)/part 15.209(a)

Band Edge Emission

15.203

Antenna Requirement

NOTE: (1)" N/A" denotes test is not applicable in this Test Report (2) All tests are according to ANSI C63.10-2013

Judgment PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS

Remark ------------

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 7 of 47

Report No.: FCS202211204W01

1.1 TEST FACTORY

Company Name:

Flux Compliance Service Laboratory

Address:

Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan

Telephone:

+86-769-27280901

Fax:

+86-769-27280901

Laboray Accreditations

FCC Test Firm Registration Number: 514908 CNAS Number L15566 Designation numberCN0127 A2LA accreditation number5545.01 ISED Number25801

1.2 MEASUREMENT UNCERTAINTY
The reported uncertainty of measurement y ±U, where expended uncertainty U is based on a standard uncertainty multiplied by a coverage factor of k=2, providing a level of confidence of approximately 95 %.

No.

Item

Uncertainty

1

RF output power, conducted

±0.71dB

2

Unwanted Emissions, conducted

±2.988 dB

3

Conducted Emission (9KHz-150KHz)

±4.13 dB

4

All emissions radiated (9KHz -30MHz)

±3.1 dB

5

Conducted Emission (150KHz-30MHz)

±4.74 dB

6

All emissions,radiated(<1G) 30MHz-1000MHz

±5.2 dB

7

All emissions,radiated 1GHz -18GHz

±4.66 dB

8

All emissions,radiated 18GHz -40GHz

±4.31 dB

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 8 of 47

Report No.: FCS202211204W01

2. GENERAL INFORMATION 2.1 GENERAL DESCRIPTION OF THE EUT

Product Name

Behind-The-Ear Hearing Aid

Trade Name

N/A

Model Name

KD-17B

Series Model Model Difference Channel List

KD-X78,KG-7XB,KV-70X
The above product with same circuit, PCB layout, electrical parts, materials and wiring structures, Appearance shape, the materials of decorative accessories is same, only different color.
Please refer to the Note 2.

Operation frequency

2402MHz-2480MHz

Modulation:

GFSK

Channel number

79 CH

Transmitter rate:

1MHz

Power Supply

DC 3.7V

Battery

DC 3.7V,30mAh

Report number

FCS202211204W01

Hardware version number V1.0

Software version number V1.0

Connecting I/O Port(s)

Please refer to the User's Manual

Note: 1. For a more detailed features description, please refer to the manufacturer's specifications or the User's Manual.

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 9 of 47

Report No.: FCS202211204W01

2.

3. Table for Filed Antenna

Ant.

Brand

1

NA

Model Name
N/A

Antenna Type Connector Gain (dBi) NOTE

Chip Antenna N/A

4.08

Antenna

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 10 of 47

Report No.: FCS202211204W01

2.2 BLOCK DIGRAM SHOWING THE CONFIGURATION OF SYSTEM TESTED During testing channel & power controlling software provided by the customer was used to control the operating channel as well as the output power level. The RF output power selection is for the setting of RF output power expected by the customer and is going to be fixed on the firmware of the final end product.
Block diagram of EUT configuration for test

EUT

Test software: the

The test softeware was used to control EUT work in continuous TX mode, and select test channel, Wireless mode as below table

No.

Test model descrption

1

Low channel GFSK

2

Middle channel GFSK

3

High channel GFSK

4

Hopping GFSK

Note: 1. All the test modes can be supply by battery, only the result of the worst case recorded in the report.
GFSK mode is worst mode. 2. For radiated emission, 3 axis were chosen for testing for each applicable mode. 3. The EUT used fully charge battery when tested. 4. During the test, the dutycycle>98%, the test voltage was tuned from 85% to 115% of the Nominal rate supply votage, and found that the worst case was the nominal rated supply condition, So the report just shows that condition's data

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 11 of 47

Report No.: FCS202211204W01

2.3 DESCRIPTION OF NECESSARY ACCESSORIES AND SUPPORT UNITS
The EUT has been tested as an independent unit together with other necessary accessories or support units. The following support units or accessories were used to form a representative test configuration during the tests.
Necessary accessories

Item Equipment

1

Adapter

Mfr/Brand HW

Model/Type No. 0789SK

Serial No.
N/A

Note
This adapter is for testing only
in report.

Item Equipment

N/A

N/A

Support units

Mfr/Brand

Model/Type No.

N/A

N/A

Serial No.
N/A

Note N/A

Note:
(1) The support equipment was authorized by Declaration of Confirmation. (2) For detachable type I/O cable should be specified the length in cm inLengthcolumn. (3) "YES" is means "shielded" "with core"; "NO" is means "unshielded" "without core".

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 12 of 47

Report No.: FCS202211204W01

2.4 EQUIPMENTS LIST

Radiation Test equipment
Kind of Equipment Manufacturer

Type No.

Company No. Last calibration Calibrated until

EMI Test Receiver

R&S

ESRP 3

FCS-E001 2022.01.28 2023.01.27

Signal Analyzer

R&S

FSV40-N

FCS-E012 2022.01.28 2023.01.27

Active loop Antenna

ZHINAN

ZN30900C

FCS-E013 2022.01.28 2023.01.27

Bilog Antenna

SCHWARZBECK VULB 9168

FCS-E002 2022.01.28 2023.01.27

Horn Antenna

SCHWARZBECK BBHA 9120D

FCS-E003 2022.03.17

SHF-EHF Horn Antenna (18G-40GHz)
Pre-Amplifier(0.1M-3G Hz)
Pre-Amplifier (1G-18GHz)
Pre-Amplifier (18G-40GHz)
Temperature & Humidity

A-INFO EMCI N/A TERA-MW HTC-1

LB-180400-KF EM330N
TSAMP-0518SE TRLA-0400 victor

FCS-E018 FCS-E004 FCS-E014 FCS-E019 FCS-E005

2022.01.28 2022.01.28 2022.01.28 2022.01.28 2022.01.28

Testing Software

EZ-EMC(Ver.STSLAB 03A1 RE)

2023.03.16 2023.01.27 2023.01.27 2023.01.27 2023.01.27 2023.01.27

Conduction Test equipment
Kind of Equipment Manufacturer

EMI Test Receiver

R&S

LISN

R&S

LISN Temperature &
Humidity Testing Software

ETS HTC-1

Type No. ESPI
ENV216 3810/2NM
victor

Company No. Last calibration Calibrated until

FCS-E020

2022.01.28

2023.01.27

FCS-E007

2022.01.28

2023.01.27

FCS-E009

2022.01.28

2023.01.27

FCS-E008

2022.01.28

2023.01.27

EZ-EMC(Ver.EMC-CON 3A1.1)

RF Connected Test
Kind of Equipment
MXA SIGNAL Analyzer
Spectrum Analyzer

Manufacturer Keysight Agilent

Spectrum Analyzer

R&S

Power Sensor

Agilent

Testing Software

Type No. N9020A E4447A FSV-40 UX2021XA

Company No. Last calibration Calibrated until

FCS-E015

2022.01.28

2023.01.27

MY50180039 2022.01.28

2023.01.27

101499

2022.01.28

2023.01.27

FCS-E021

2022.01.28

2023.01.27

EZ-EMC(Ver.STSLAB 03A1 RE)

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 13 of 47

Report No.: FCS202211204W01

3 CONDUCTED EMISSION MEASUREMENT

3.1 LIMIT
Operating frequency band. In case the emission fall within the restricted band specified on Part 207(a) limit in the table below has to be followed.

FREQUENCY (MHz)

Conducted Emissionlimit (dBuV)

Quasi-peak

Average

0.15 -0.5

66 - 56 *

56 - 46 *

0.50 -5.0

56.00

46.00

5.0 -30.0

60.00

50.00

Note: (1) The tighter limit applies at the band edges. (2) The limit of " * " marked band means the limitation decreases linearly with the logarithm of the
frequency in the range.

3.2 TEST PROCEDURE The following table is the setting of the receiver
Receiver Parameters Attenuation
Start Frequency
Stop Frequency
IF Bandwidth

Setting 10 dB 0.15 MHz 30 MHz 9 kHz

a. The EUT was 0.8 meters from the horizontal ground plane and 0.4 meters from the vertical ground plane with EUT being connected to the power mains through a line impedance stabilization network (LISN). All other support equipments powered from additional LISN(s). The LISN provide 50 Ohm/ 50uH of coupling impedance for the measuring instrument.
b. Interconnecting cables that hang closer than 40 cm to the ground plane shall be folded back and forth in the center forming a bundle 30 to 40 cm long.
c. I/O cables that are not connected to a peripheral shall be bundled in the center. The end of the cable may be terminated, if required, using the correct terminating impedance. The overall length shall not exceed 1 m.
d. LISN at least 80 cm from nearest part of EUT chassis. e. For the actual test configuration, please refer to the related Item ­EUT Test Photos.

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

3.3 TEST SETUP

Page 14 of 47

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 15 of 47

Report No.: FCS202211204W01

3.4 TEST RESULTS

Temperature: Test Mode: Result:

25 GFSK(worst mode) L

Relative Humidity: 50%

Test Voltage:

DC 3.7V by Battery

Result:

Pass

No. Frequency Reading

(MHz)

(dBuV)

1 0.1725 2 0.1725 3 0.7125 4 0.7125 5 1.4460 6 1.4460 7 1.9995 8 1.9995 9 3.9750 10 3.9750 11 16.3005 12 16.3005

35.23 16.54 34.88 25.94 22.70 11.31 21.61 11.28 20.64 9.95 25.92 9.07

Correct Factor(dB)
9.78 9.78 9.89 9.89 12.15 12.15 14.93 14.93 18.24 18.24 20.13 20.13

Result (dBuV)
45.01 26.32 44.77 35.83 34.85 23.46 36.54 26.21 38.88 28.19 46.05 29.20

Limit (dBuV)
64.84 54.84 56.00 46.00 56.00 46.00 56.00 46.00 56.00 46.00 60.00 50.00

Margin (dB)
19.83 28.52 11.23 10.17 21.15 22.54 19.46 19.79 17.12 17.81 13.95 20.80

Remark
QP AVG QP AVG QP AVG QP AVG QP AVG QP AVG

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 16 of 47

Report No.: FCS202211204W01

Temperature: Test Mode: Result:

25 GFSK(worst mode) N

Relative Humidity: 50%

Test Voltage:

DC 3.7V by Battery

Result:

Pass

No. Frequency Reading

(MHz)

(dBuV)

Correct Factor(dB)

Result (dBuV)

Limit (dBuV)

1 0.7170 2 0.7170 3 1.2255 4 1.2255 5 2.0175 6 2.0175 7 4.3439 8 4.3439 9 6.5670 10 6.5670 11 15.6300 12 15.6300

37.62 25.83 27.13 14.70 23.93 12.05 21.88 10.16 20.47 7.34 26.71 14.13

9.89 9.89 11.05 11.05 14.96 14.96 18.91 18.91 20.09 20.09 20.21 20.21

47.51 35.72 38.18 25.75 38.89 27.01 40.79 29.07 40.56 27.43 46.92 34.34

56.00 46.00 56.00 46.00 56.00 46.00 56.00 46.00 60.00 50.00 60.00 50.00

Remark: 1. All readings are Quasi-Peak and Average values

Margin (dB)
8.49 10.28 17.82 20.25 17.11 18.99 15.21 16.93 19.44 22.57 13.08 15.66

Remark
QP AVG QP AVG QP AVG QP AVG QP AVG QP AVG

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

4. 20 DB BANDWIDTH 4.1 Limit

Page 17 of 47

Report No.: FCS202211204W01

4.2 Test Procedure (1) Connect EUT's antenna output to spectrum analyzer by RF cable.

(2) Set the spectrum analyzer as follows

RBW:

30kHz

VBW:

100kHz

Detector Mode:

AVG

Sweep time:

auto

Trace mode

Max hold

(3) Allow the trace to stabilize, measure the maximum width of the emission that is constrained by the

frequencies associated with the two outermost amplitude points (upper and lower frequencies) that are

attenuated by 6 dB relative to the maximum level measured in the fundamental emission

4.3 Test setup

4.4 Test results
TestMode Lowest Middle Highest

Channel (MHz)
2402MHz 2441MHz 2480MHz

20dB Bandwidth (KHz)
998.8 964.2 1000.0

Verdict
Pass Pass Pass

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

4.5 Original Test Data

Page 18 of 47
Low

Report No.: FCS202211204W01

Middle

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 19 of 47
High

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 20 of 47

Report No.: FCS202211204W01

5. CONDUCTED OUTPUT POWER 5.1 LIMIT

Section 15.247(b)(3)

FCC Part 15 Subpart C

Test ltem

Limit

Peak output power Power <1W(30dBm)

Frequency Range 2400-2483.5

5.2 TEST PROCEDURE

5.3 TEST SETUP

5.5 TEST RESULTS

TestMode
Lowest Middle Highest

Channel (MHz)
2402MHz 2441MHz 2480MHz

ResultdBm
4.753 4.702 3.811

LimitdBm
30 30 30

Verdict
Pass Pass Pass

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 21 of 47
Low

Report No.: FCS202211204W01

Middle

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 22 of 47
Hight

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 23 of 47

Report No.: FCS202211204W01

6 NUMBER OF HOPPING CHANNEL 6.1 LIMIT

FCC Part 15.247,Subpart C RSS-247 Issue 2

Section

Test Item

15.247

Number of Hopping

(a)(1)(iii) RSS-247

Channel

Limit >15

FrequencyRange (MHz)
2400-2483.5

Result PASS

6.2 TEST PROCEDURE a The EUT was directly connected to the spectrum analyzer and antenna output port as show in the block diagram below. b Spectrum Setting: RBW= 100KHz, VBW=300KHz, Sweep time = Auto

5.3 TEST SETUP

Spectrum Analyzer
6.4 EUT OPERATION CONDITIONS The EUT tested system was configured as the statements of 2.4 Unless otherwise a special operating condition is specified in the follows during the testing.

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 24 of 47

Report No.: FCS202211204W01

6.5 TEST RESULTS

Temperature: Test Mode:

25 °C

Relative Humidity: 60%

Hopping Mode -GFSK Mode Test Voltage:

DC 3.7V

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 25 of 47

Report No.: FCS202211204W01

7. BAND EDGE AND SPURIOUS(CONDUCTED)
7.1 LIMIT In any 100kHz bandwidth outside the frequency bands in which the spread spectrum intentional radiator in operating, the radio frequency power that is produced by the intentional radiator shall be at least 30dB below that in the 100kHz bandwidth within the band that contains the highest level of the desired power.

7.2 TEST PROCEDURE

(1) Connect EUT's antenna output to spectrum analyzer by RF cable.

(2) Establish a reference level by using the following procedure:

Center frequency

DTS Channel center

frequency

RBW:

100kHz

VBW:

300kHz

Span

1.5times the DTS bandwidth

Detector Mode:

Peak

Sweep time:

auto

Trace mode

Max hold

(3) Establish Allow the trace to stabilize, use the peak marker function to determine the maximum peak

power level to establish the reference level.

(4) Set the spectrum analyzer as follows:

RBW:

100kHz

VBW:

300kHz

Span

Encompass frequency range to be

measured

Number of measurement points

span/RBW

Detector Mode:

Peak

Sweep time:

auto

Trace mode

Max hold

(5) Allow the trace to stabilize, use the peak marker function to determine the maximum amplitude of

all unwanted emissions outside of the authorized frequency band

7.3 TEST SETUP

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

7.4 TEST RESULTS Eut set mode GFSK
7.5 Original test data

Page 26 of 47
CH or Frequency CH1 CH79
CH1 2402MHZ

Report No.: FCS202211204W01
Result Pass Pass

CH79 2480MHZ
Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Spurious emissions

Page 27 of 47

Report No.: FCS202211204W01

Low Channel 30MHz-25GHz

Middle Channel 30MHz-25GHz

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 28 of 47

Report No.: FCS202211204W01

High Channel 30MHz-25GHz

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 29 of 47
7.6 For Hopping Band edge

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 30 of 47

Report No.: FCS202211204W01

8. RADIATED EMISSION MEASUREMENT

8.1 RADIATED EMISSION LIMITS
In any 100 kHz bandwidth outside the operating frequency band. In case the emission fall within the Restricted band specified on Part15.205(a)&209(a) limit in the table and according to ANSI C63.10-2013 below has to be followed LIMITS OF RADIATED EMISSION MEASUREMENT (0.009MHz - 1000MHz)

Frequencies

Field Strength

Measurement Distance

(MHz)

(micorvolts/meter)

(meters)

0.009~0.490

2400/F(KHz)

300

0.490~1.705

24000/F(KHz)

30

1.705~30.0

30

30

30~88

100

3

88~216

150

3

216~960

200

3

Above 960

500

3

LIMITS OF RADIATED EMISSION MEASUREMENT (1GHz-25 GHz)

FREQUENCY (MHz) Above 1000

PEAK 74

(dBuV/m) (at 3M) AVERAGE 54

Notes: (1) The limit for radiated test was performed according to FCC PART 15C. (2) The tighter limit applies at the band edges. (3) Emission level (dBuV/m)=20log Emission level (uV/m).

For Radiated Emission Spectrum Parameter Attenuation Detector Start Frequency Stop Frequency
RB / VB (emission in restricted band)

Setting Auto
Peak/AV 1000 MHz(Peak/AV) 10th carrier hamonic(Peak/AV)
PK=1MHz / 1MHz, AV=1 MHz /10 Hz

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 31 of 47

Report No.: FCS202211204W01

For Band edge Spectrum Parameter Detector
Start/Stop Frequency
RB / VB (emission in restricted band)
Receiver Parameter Attenuation
Start ~ Stop Frequency Start ~ Stop Frequency Start ~ Stop Frequency Start ~ Stop Frequency Start ~ Stop Frequency

Setting Peak/AV Lower Band Edge: 2300 to 2403 MHz Upper Band Edge: 2479 to 2500 MHz PK=1MHz / 1MHz, AV=1 MHz / 10 Hz
Setting Auto
9kHz~90kHz / RB 200Hz for PK & AV 90kHz~110kHz / RB 200Hz for QP
110kHz~490kHz / RB 200Hz for PK & AV 490kHz~30MHz / RB 9kHz for QP
30MHz~1000MHz / RB 120kHz for QP

8.2 TEST PROCEDURE
a. The measuring distance of at 3 m shall be used for measurements at frequency 0.009MHz up to 1GHz,and above 1GHz.
b. The EUT was placed on the top of a rotating table 0.8 meters (above 1GHz is 1.5 m) above the ground at a 3 meter anechoic chamber test site. The table was rotated 360 degrees to determine the position of the highest radiation.
c. The height of the equipment shall be 0.8 m(above 1GHz is 1.5 m); the height of the test antenna shall vary between 1 m to 4 m. horizontal and vertical polarizations of the antenna are set to make the measurement.
d. The initial step in collecting conducted emission data is a spectrum analyzer peak detector mode pre-scanning the measurement frequency range. Significant peaks are then marked and then QuasiPeak detector mode re-measured.
e. If the Peak Mode measured value compliance with and lower than Quasi Peak Mode Limit, the EUT shall be deemed to meet QP Limits and then no additional QP Mode measurement performed.
f. For the actual test configuration, please refer to the related Item ­EUT Test Photos.
Note:
Both horizontal and vertical antenna polarities were tested and performed pretest to three orthogonal axis. The worst case emissions were reported

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 32 of 47
8.3 TESTSETUP ARadiated Emission Test-Up Frequency Below 30MHz

Report No.: FCS202211204W01

BRadiated Emission Test-Up Frequency 30MHz~1GHz

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 33 of 47
CRadiated Emission Test-Up Frequency Above 1GHz

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 34 of 47

Report No.: FCS202211204W01

8.4. TEST RESULTS

(9KHz-30MHz)

Temperature:

22.7

Test Voltage:

DC 3.7V

Relative Humidity: Test Mode:

61% GFSK(worst mode)

Freq. (MHz)
---

Reading (dBuV/m)
---

Limit (dBuV/m)
---

Margin (dB) ---

State P/F ---

Test Result
PASS PASS

Note: The amplitude of spurious emissions which are attenuated by more than 20dB below the permissible value has no need to be reported. Distance extrapolation factor =40 log (specific distance/test distance)(dB); Limit line = specific limits (dBuv) + distance extrapolation factor.

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 35 of 47

Report No.: FCS202211204W01

8.5 (30MHZ-1000MHZ)

Temperature:

24.7

Test Voltage:

DC 3.7V

Test Mode:

GFSK(worst mode)

Relative Humidity: Phase:

61% Horizontal

No. Frequency (MHz)

Reading (dBuV)

Correct Factor(dB/
m)

Result (dBuV/m)

Limit (dBuV/m)

1 69.3568 2 105.6414

36.91 31.99

-20.21 -17.98

16.70 14.01

40.00 43.50

3 251.1802 43.20

-17.70

25.50

46.00

4 312.1792 42.33

-15.77

26.56

46.00

5 435.5898 54.31

-14.44

39.87

46.00

6 599.3211 34.68

-11.36

23.32

46.00

Note: 1. Margin = Result (Result =Reading + Factor )­Limit

2. If Peak Result complies with QP limit, QP Result is deemed to comply with QP limit.

3. Test setup: RBW: 120 kHz, VBW: 300 kHz, Sweep time: auto.

Margin (dB)
-23.30 -29.49 -20.50 -19.44 -6.13 -22.68

Remark
QP QP QP QP QP QP

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 36 of 47

Report No.: FCS202211204W01

Temperature: Test Voltage: Test Mode:

22.7 DC 3.7V GFSK(worst mode)

Relative Humidity: Phase:

61% Vertical

No. Frequency (MHz)

Reading (dBuV)

Correct Factor(dB/
m)

Result (dBuV/m)

Limit (dBuV/m)

1 51.3005 2 74.1351 3 109.4116 4 134.5592 5 242.5252 6 435.5898

68.52 59.79 58.68 60.70 66.08 82.33

-40.24 -40.24 -40.24 -40.24 -40.24 -40.24

28.28 19.55 18.44 20.46 25.84 42.09

40.00 40.00 43.50 43.50 46.00 46.00

Note: 1. Margin = Result (Result =Reading + Factor )­Limit 2. If Peak Result complies with QP limit, QP Result is deemed to comply with QP limit. 3. Test setup: RBW: 120 kHz, VBW: 300 kHz, Sweep time: auto.

Margin (dB)
-11.72 -20.45 -25.06 -23.04 -20.16 -3.91

Remark
QP QP QP QP QP QP

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

ABOVE 1GHZ LOW CH(GFSK)
Peak value

Page 37 of 47

Report No.: FCS202211204W01

AV value

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

MIDDLE CH(GFSK) Peak value

Page 38 of 47

Report No.: FCS202211204W01

AV value

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

HIGH CH(GFSK) Peak value

Page 39 of 47

Report No.: FCS202211204W01

AV value

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 40 of 47

Report No.: FCS202211204W01

8.7 RADIATED BAND EDGE DATA Remark: All restriction band have been tested, and only the worst case is shown in report
Low CH (GFSK)

High CH(GFSK)

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 41 of 47

Report No.: FCS202211204W01

9. AVERAGE TIME OF OCCUPANCY 9.1 LIMIT
FCC Parti 5 (15.247), Subpart C

Section 15.247(a)(1)

Test Item Average Time of Occupancy

Limit 0.4 sec

Frequency Range (MHz) 2400-2483.5

9.2 TEST PROCEDURE a. The transmitter output (antenna port) was connected to the spectrum analyzer.
b. Set RBW =1MHz/VBW =3MHz. c. Use a video trigger with the trigger level set to enable triggering only on full pulses. d. Sweep Time is more than once pulse time. e. Set the center frequency on any frequency would be measure and set the frequency span to zero
span. f. Measure the maximum time duration of one single pulse. g. Set the EUT for DH5, DH3 and DH1 packet transmitting. h. Measure the maximum time duration of one single pulse. i. DH5 Packet permit maximum 1600/ 79 / 6 = 3.37 hops per second in each channel (5 time slots RX,
1 time slot TX). So the number of pulses in the observation period of 31.6 seconds is 3.37x31.6 = 106.6. j. DH3 Packet permit maximum 1600 / 79 / 4 = 5.06 hops per second in each channel (3 time slots RX, 1 time slot TX). So the number of pulses in the observation period of 31.6 seconds is 5.06x31.6 = 160. k. DH1 Packet permit maximum 1600 / 79 / 2 = 10.12 hops per second in each channel (1 time slot RX, 1 time slot TX). So the number of pulses in the observation period of 31.6 seconds is 10.12x31.6 = 320.

9.3 TEST SETUP

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 42 of 47

Report No.: FCS202211204W01

9.4 TEST RESULTS

CONDITIO MOD

N

E

NVNT NVNT NVNT

DH1 DH3 DH5

FREQUENCY(M HZ)
2441 2441 2441

PULSE TIME(M
S)
0.385 1.635 2.900

AVERAGE TIME OF
OCCUPANCY( MS)
124.80 262.40 309.33

LIMIT(M S)
400 400 400

RESULT S
PASS PASS PASS

9.5 ORIGINAL TEST DATA

DH1

DH3
Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 43 of 47

Report No.: FCS202211204W01

DH5

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 44 of 47

Report No.: FCS202211204W01

10. HOPPING CHANNEL SEPARATION MEASUREMEN
10.1 LIMIT Frequency hopping systems shall have hopping channel carrier frequencies separated by a minimum of 25 kHz or the 20 dB bandwidth of the hopping channel, whichever is greater. Alternatively, frequency hopping systems operating in the 2400-2483.5 MHz band may have hopping channel carrier frequencies that are separated by 25 kHz or two-thirds of the 20 dB bandwidth of the hopping channel, whichever is greater, provided the systems operate with an output power no greater than 125 mW.

Spectrum Parameter Attenuation
Span Frequency RB VB
Detector Trace
Sweep Time

Setting Auto
> 20 dB Bandwidth or Channel Separation 30 kHz (20dB Bandwidth) / 30 kHz (Channel Separation) 100 kHz (20dB Bandwidth) /100 kHz (Channel Separation)
Peak Max Hold
Auto

10.2 TEST PROCEDURE a. The transmitter output (antenna port) was connected to the spectrum analyser in peak hold mode. b. The resolution bandwidth of 30 kHz and the video bandwidth of 100 kHz were utilised for 20 dB bandwidth measurement. c. The resolution bandwidth of 30 kHz and the video bandwidth of 100 kHz were utilised for channel separation measurement
10.3 TEST SETUP

10.4 EUT OPERATION CONDITIONS The EUT was programmed to be in continuously transmitting mode.

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 45 of 47

Report No.: FCS202211204W01

10.5 TEST RESULTS

Temperature: Test Mode:

25 °C GFSK Mode

Relative Humidity: 60%

Test Voltage:

DC 5V

Modulation GFSK

Frequency (MHz) 2402
2441
2480

Channel Separation
(MHz) 1.002
1.002
1.004

Limit (MHz) 0.666
0.643
0.667

Result
Pass Pass Pass

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 46 of 47

Report No.: FCS202211204W01

Flux Compliance Service Laboratory Room 105 Floor Bao hao Technology Building 1 NO.15 Gong ye West Road Hi-Tech Industrial, Song shan lake Dongguan Tel: 769-27280901 Fax:769-27280901 http://www.FCS-lab.com

Page 47 of 47

Report No.: FCS202211204W01

11. ANTENNA REQUIREMENT
11.1 STANDARD REQUIREMENT For intentional device, according to FCC 47 CFR Section 15.203, an intentional radiator shall be designed to ensure that no antenna other than that furnished by the responsible party shall be used with the device. And according to FCC 47 CFR Section 15.247 (b), if transmitting antennas of directional gain greater than 6dBi are used, the power shall be reduced by the amount in dB that the directional gain of the antenna exceeds 6dBi.
11.2 RESULT The antennas used for this product are Chip Antenna and no antenna other than that furnished by the responsible party shall be used with the device, the maximum peak gain of the transmit antenna is 4.08dBi.

END OF THE REPORT