TR RF HAC App A

Zebra Technologies Corporation TC78B1 Touch Computer UZ7TC78B1 UZ7TC78B1 tc78b1

PDF Viewing Options

Not Your Device? Search For Manuals or Datasheets below:


File Info : application/pdf, 9 Pages, 814.07KB

Document DEVICE REPORTGetApplicationAttachment.html?id=6189237
Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_835

DUT: HAC-Dipole 835 MHz 

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m
3 Ambient Temperature23.5 

DASY5 Configuration

- Probe: ER3DV6 - SN
2526; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2022/3/28

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD835 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 122.6 V/m; Power D
 rift = -0.01 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 107.1 V/m




Average value of Total=(107.1+106.9) / 2 = 107 V/m 


PMF scaled E-field
Grid 1 M4 Grid 2 M4 Grid 3 M4 105.5 V/m 107.1 V/m 104.7 V/m Grid 4 M4 Grid 5 M4 Grid 6 M4 60.50 V/m 61.68 V/m 60.79 V/m Grid 7 M4 Grid 8 M4 Grid 9 M4 105.5 V/m 106.9 V/m 105.2 V/m
Cursor: Total = 107.1 V/m E Category: M4 
 Location: 0.5, -78, 9.7
 mm

0 dB = 107.1 V/m = 40.60 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_835

DUT: HAC-Dipole 835 MHz 

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m
3 Ambient Temperature23.5 

DASY5 Configuration

- Probe: EF3DV3 - SN
4047; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2022/1/24

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD835 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 130.0 V/m; Power D
 rift = -0.04 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 122.2 V/m




Average value of Total=(122.2+105.3) / 2 = 113.75 V/m 


PMF scaled E-field
Grid 1 M4 Grid 2 M4 Grid 3 M4 116.7 V/m 122.2 V/m 121.2 V/m Grid 4 M4 Grid 5 M4 Grid 6 M4 60.60 V/m 63.17 V/m 63.12 V/m Grid 7 M4 Grid 8 M4 Grid 9 M4 101.3 V/m 105.3 V/m 104.6 V/m
Cursor: Total = 122.2 V/m E Category: M4 
 Location: -1.5, -74, 9.7
 mm



0 dB = 122.2 V/m = 41.74 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_1880

DUT: HAC Dipole 1880 MHz 

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: ER3DV6 - SN
2526; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2022/3/28

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 136.8 V/m; Power D
 rift = 0.14 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 88.21 V/m




Average value of Total=(86.96+88.21) / 2 = 87.585 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 85.52 V/m 86.96 V/m 85.30 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 67.66 V/m 68.73 V/m 67.37 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 87.35 V/m 88.21 V/m 85.68 V/m
Cursor: Total = 88.21 V/m E Category: M3 
 Location: 0.5, 31, 9.7
 mm

0 dB = 88.21 V/m = 38.91 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_1880

DUT: HAC Dipole 1880 MHz 

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: EF3DV3 - SN
4047; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2022/1/24

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 172.0 V/m; Power D
 rift = 0.03 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 92.68 V/m




Average value of Total=(89.63+92.68) / 2 = 91.155 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 86.05 V/m 89.63 V/m 89.13 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 64.08 V/m 65.47 V/m 65.45 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 88.95 V/m 92.68 V/m 91.68 V/m
Cursor: Total = 92.68 V/m E Category: M3 
 Location: -1, 33.5, 9.7
 mm

0 dB = 92.68 V/m = 39.34 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_2450

DUT: HAC Dipole 2450 MHz 

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: ER3DV6 - SN
2526; ConvF(1, 1, 1) @ 2450 MHz; Calibrated: 2022/3/28

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD2450 = 10mm & 15mm

2/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,

dy=0.5000 mm

Device Referen
ce Point: 0, 0, -6.3 mm

Reference Value = 78.81 V/m; Power D
 rift = 0.00 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 86.74 V/m




Average value of Total=(82.28+86.74) / 2 = 84.51 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 81.74 V/m 82.28 V/m 79.38 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 80.08 V/m 81.99 V/m 80.70 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 85.34 V/m 86.74 V/m 84.88 V/m
Cursor: Total = 86.74 V/m E Category: M3 
 Location: 0, 22, 9.7 
mm

0 dB = 86.74 V/m = 38.76 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_2450

DUT: HAC Dipole 2450 MHz 

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: EF3DV3 - SN
4047; ConvF(1, 1, 1) @ 2450 MHz; Calibrated: 2022/1/24

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD2450 = 10mm & 15mm

2/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,

dy=0.5000 mm

Device Referen
ce Point: 0, 0, -6.3 mm

Reference Value = 76.37 V/m; Power D
 rift = 0.02 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 81.57 V/m




Average value of Total=(80.07+81.57) / 2 = 80.82 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 76.75 V/m 80.07 V/m 79.61 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 69.41 V/m 71.78 V/m 71.44 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 78.27 V/m 81.57 V/m 80.88 V/m
Cursor: Total = 81.57 V/m E Category: M3 
 Location: -1, 24.5, 9.7
 mm

0 dB = 81.57 V/m = 38.23 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_2600

DUT: HAC Dipole 2600 MHz 

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: ER3DV6 - SN
2526; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2022/3/28

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 64.28 V/m; Power D
 rift = 0.07 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 86.81 V/m




Average value of Total=(83.81+86.81) / 2 = 85.31 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 83.47 V/m 83.81 V/m 80.67 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 83.21 V/m 84.77 V/m 83.42 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 85.55 V/m 86.81 V/m 85.08 V/m
Cursor: Total = 86.81 V/m E Category: M3 
 Location: 0.5, 19.5, 9.
7 mm

0 dB = 86.81 V/m = 38.77 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_2600

DUT: HAC Dipole 2600 MHz 

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: EF3DV3 - SN
4047; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2022/1/24

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 69.98 V/m; Power D
 rift = -0.02 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 87.35 V/m




Average value of Total=(86.16+87.35) / 2 = 86.755 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 83.31 V/m 86.16 V/m 85.60 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 78.66 V/m 80.87 V/m 80.76 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 84.15 V/m 87.35 V/m 86.71 V/m
Cursor: Total = 87.35 V/m E Category: M3 
 Location: -1, 23, 9.7
mm

0 dB = 87.35 V/m = 38.83 dBV/m

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab

Date: 2022/9/20

HAC_E_Dipole_3500

DUT: HAC-Dipole 3500 MHz
Communication System: CW ; Frequency: 3500 MHz;Duty Cycle: 1:1 Medium: Air Medium parameters used:  = 0 S/m, r = 1;  = 0 kg/m3 
 Ambient Temperature23.5 

DASY5 Configuration

- Probe: EF3DV3 - SN
4047; ConvF(1, 1, 1) @ 3500 MHz; Calibrated: 2022/1/24

- Sensor-Surface: (Fix Surface)




- Electronics: DAE4 Sn1512; C
alibrated: 2022/3/29

- Phantom: HAC Test Arch with AMCC; Type: SD H
 AC P01 BA;

- Measurement SW: DASY52, Version 52.10 (4); SEMCAD X Vers
 ion 14.6.14 (7483)

E Scan - measurement distance from the probe sensor center to CD3500 = 10mm &

15mm/Hearing Aid Compatibility Test at 15mm distance (41x121x1): Interpolated grid: dx=0.5000

mm, dy=0.5000 mm

Device Reference Po
int: 0, 0, -6.3 mm

Reference Value = 36.44 V/m; Power D
 rift = 0.04 dB

PMR not calibrated. PMF = 1.000 is applied.




E-field emissions = 88.17 V/m




Average value of Total=(88.17+86.95) / 2 = 87.56 V/m

PMF scaled E-field
Grid 1 M3 Grid 2 M3 Grid 3 M3 85.89 V/m 88.17 V/m 87.72 V/m Grid 4 M3 Grid 5 M3 Grid 6 M3 82.98 V/m 84.45 V/m 84.24 V/m Grid 7 M3 Grid 8 M3 Grid 9 M3 83.92 V/m 86.95 V/m 86.18 V/m
Cursor: Total = 88.17 V/m E Category: M3 
 Location: -0.5, -15, 9.7
 mm

0 dB = 88.17 V/m = 38.91 dBV/m



Related FCC IDs:

Search Any Device: